Manual

2.3 Settings on KF Oven
768 KF Oven
7
2.3 Settings on KF Oven
2.3.1 Keypad
Keys for individual manual functions
<HEATER> Switches heating off/on. The "HEATER" lamp lights up or
flashes when the heating is switched on. (The lamp is on
continuously when full heating power is used; it flashes with
reduced heating power.)
Note:
If the heating is switched off and if the valve is on
"TRANSFER", the solvent may be sucked back from the ti-
tration vessel.
<PUMP> Switches pump off/on. The "PUMP" lamp lights up when the
pump is switched on.
<VALVE> Switches valve. The "VALVE, PURGE" and "VALVE,
TRANSFER" lamps show the position of the valve. For
safety reasons, the valve is always set to "PURGE" at the
end of an automatic determination as well as on switching
off the Oven (suck-back of solution from the titration vessel!)
<BOAT IN> Inserts the sample boat in the hot part of the oven. The boat
continues to move until the key is pressed again or until the
boat reaches the inner stop.
<BOAT OUT> Withdraws the sample boat from the hot part of the oven.
The boat continues to move until the key is pressed again or
the boat reaches the outer stop.
<PRINT> Triggers report printout
Key sequence:
<PRINT><SELECT><ENTER>; press <SELECT> repeat-
edly until the desired report appears in the display
or
<PRINT><Key X><ENTER>; Key X = PARAM or CONFIG
7
HEATER
8 9
PUMP
4 5
VALVE
6
PUMP
1
BOAT IN
2 3
BOAT OUT
0
PRINT
.
QUIT ENTER PARAM START
SELECT CLEAR CONFIG STOP
TRANSFER
PURGE
HEATER
PUMP
VALVE
READY