Manual

7.2 Reports
102
780/781 pH/Ion Meter, Manual
7.2.5 Result report (781 pH/Ion Meter only)
In the Conc mode for standard or sample addition the results can be
produced as a result report. Under
PARAMETER Conc/Standard ad-
dition/Report
you can define whether this result is to be produced
automatically after the end of the measurement. Just as for the calibra-
tion report, you can choose between a short version containing all the
important data and a long version which additionally contains a graph-
ical presentation of the evaluation.
'ca
781 pH/Ion Meter 01104 5.781.0020
date 2004-06-03 09:38:25
user C. Weber
Conc F-StdAdd run number 2
electr.id F-electr.011
addition/subtraction method
meas.type: std.add
temperature 20.1 °C
temp.sensor Pt1000
conc.std 1000.0 ppm
V total 20.0 mL
initial voltage 180.0 mV
factor 25.0
smpl.size 10.0 mL
date 2004-06-03 09:33
dV/ml U/mV dU/mV
std.incr. 1 0.020 115.5 -64.5
std.incr. 2 0.026 95.5 -20.0
std.incr. 3 0.056 76.0 -19.5
std.incr. 4 0.123 55.6 -20.4
variance 0.080
slope -59.0 mV
E(0) 117.7 mV
c F(-1) 4.40 ppm
==============
REPORT
9
+
ADD. DATA