Manual

5.2 Printing out measured values
44
780/781 pH/Ion Meter, Manual
5.2 Printing out measured values
MEAS/
PRI NT
CONFIG
:
print meas.value
: print crit immediate date & time
report header
calibration report
time date & time
report header
calibration report
time interval
stop time
drift date & time
report header
calibration report
plot time interval
time scale
time scale label
stop time
OFF
In the configuration menu print meas.values the output form of
measured values, triggered by
<MEAS/PRINT>,
is defined. This is only
possible for the direct measuring modes pH, T, U and Conc(direct).
print crit: immediate, time, drift, plot, OFF
Depending upon the print criterion a measured value report with be
transmitted via the RS232 interface to a printer or PC either
immediate-
ly
, at defined
time
intervals or only when a
drift
condition has been ful-
filled. If a stirring sequence is intended before the measurement (
PA-
RAMETERS/measuring parameters/stirrer: control,
see Sec-
tion
6), then these print criteria will only be effective when the stirring
sequence has been concluded.
A further way of documenting the measured values is to output a
graphical
plot
of the measure value signals.
print crit.: immediate
With this setting a measured value report will be produced when
<MEAS/PRINT>
is activated from the normal instrument display. The
following options can be used:
date & time: ON, OFF
Here you can choose whether the date and time at which the
measured value was recorded are to be included with each indi-
vidual measured value. This can take place irrespective of wheth-
er a report header is used.
report header: once, always, OFF