5 Configuration
780/781 pH/Ion Meter, Manual
47
5.3 Storing measured values
ST ORE
5
CONFIG
:
store meas. value
: store crit immediate, drift, OFF
time time interval
stop time
In the direct measuring modes pH, T, U and Conc(direct) you can use
<STORE>
to store up to 100 measured values with additional infor-
mation in the instrument memory (see Section
7.3). The functions of the
settings to be defined under
CONFIG/store meas. value correspond
to those made under
CONFIG/print meas. value:
store crit: immediate, time, drift, OFF
This storage criterion also only becomes effective when an optional stir-
ring sequence has been concluded.
In contrast to the report output, the combination of the information
stored together with each measured value is fixed.
store crit: immediate
The current measured value is stored with
<STORE>
.
store crit: time
time interval: 0.1...4.0...99999.0 s
stop time: OFF, 1...999999 s
You can automatically record and save the measured values at fixed
time intervals until the stop time has been reached.
store crit: drift
A measured value can also be stored when the drift condition defined
under
PARAMETERS/measuring parameters/drift has been fulfilled.
After 100 measured values have been stored, and each time a further
storage attempt is made, the pH/Ion Meter will produce a correspond-
ing warning:
measured value memory full
You now have the possibility of using
<RECALL>
to view the stored
measured values, of deleting individual measured values or of clearing
the whole measured value memory (see Section
7.3). It is also possible
to output all measured values as a report via the RS232 interface (see
Section 7.2.8).