Owner's manual

2.2 Installation of the 797 VA Computrace Stand
797 VA Computrace / Hardware-Manual 8.797.8001EN
15
3 Insertion of 4-way microtip (option)
For automatic solution addition with Dosinos or Dosimats, the 6.1824.000 4-
way microtip has to be installed. Proceed as follows:
Remove stopper from nipple
27 and insert 4-way microtip 30 into nipple as
far as it will go.
Tighten nipple using a 6.2739.010 Wrench until the 4-way microtip can no
longer move.
Pull the 4 lengths of PTFE tubing of the 4-way microtip in succession from
above through the opening
68 (connection of Dosinos or Dosimats see sec-
tion
2.8).
4 Install stirrer or RDE
in operation with MME:
Insert stirrer (6.1204.200) in opening
63 as far as it will go.
Stretch drive belt
32 (6.1244.020) between drive wheel 35 and drive shaft
of the stirrer
.
in operation with RDE:
Screw electrode tip
98 (6.1204.XXX) to drive shaft 99 (6.1204.210) (see
also section
2.4).
Insert RDE in opening
63 as far as it will go.
Stretch drive belt
32 (6.1244.020) between drive wheel 35 and drive shaft
99 of the RDE.
Attach electrode cable
20 (WE) to the RDE: push cable lug under the screw
and then tighten screw firmly.
Fig. 6: Measuring head arm
23
Measuring head arm
24
Measuring head
54
Opening
for auxiliary electrode
43 (6.0343.000 Pt
- auxiliary electr. or optional GC electr.
comprising 6.1241.020 Electrode holder
and 6.1247.000 GC tip)
59 60 61 62 65 66 67 68 69 70
6463
54
24
57
58
23
55
56