can also lead to leakage of the MME, i.e. the mercury flow cannot be stopped any
more.
Proceed as follows:
1 Remove Multi-Mode Electrode from 797 VA Computrace Stand
Unscrew FEP tubing 34 and 42 from the MME, disconnect electrode cable
20 from MME.
Take Multi-Mode Electrode
21 out of measuring head 24 while gently tap-
ping the MME to knock off any mercury drops on the glass capillary into
the measuring vessel.
Place Multi-Mode Electrode in a horizontal position in the electrode holder
92 (see Fig. 11-3).
2 Unscrew slotted screw 74
Using a suitable coin, unscrew slotted screw 74 out of screw thread 76 un-
til the contact surface of the black O-ring at the Plexiglas wall (thin, black
stripe) is just visible below the metal thread.
3 Replace sealing needle 75
If problems with leaks arise owing to a worn, deformed or damaged sealing
needle, it must be replaced. Three spare needles are supplied separately in a
protective plastic package. After unpacking a needle, please avoid any contact
whatsoever with the needle tip. The spare needle
75 is installed as follows:
Carefully pull old sealing needle out of PTFE membrane of the slotted screw
74.
Carefully insert new sealing needle without tilting into the hole in the PTFE
membrane of the slotted screw.
NOTE: Do not touch the needle with bare fingers but use the plastic package
4 Dismantle old capillary
Position Multi-Mode Electrode vertically in the electrode holder (see Fig. 11-
1).
Undo retaining nut
86 completely by turning anticlockwise and lift up until
the lower part of the glass capillary
87 with the wide opening is visible.
Gently tap the glass capillary to knock off any residual mercury in the wide
opening into the MME.
Press the retaining nut downward with one hand and with your other hand
take glass capillary completely out of the mount.
5 Dispose of old capillary
Connect filling tubing 93 with the tubing coupling 96 to the two gas wash
bottles
95 and the vacuum pump (see Fig. 10).
Insert glass capillary
87 (capillary end) in the filling cone 94 of the filling
tubing.
Remove mercury from capillary with the vacuum pump.
6 Replenish mercury if necessary
Proceed as described in section 2.3.2.