843 Pump Station with membrane pumps Manual 8.843.
Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 843 Pump Station with membrane pumps Manual 8.843.8001EN 03.
Teachware Metrohm AG CH-9101 Herisau teachware@metrohm.com This documentation is protected by copyright. All rights reserved. Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents Table of contents 1 Introduction 1 1.1 Instrument description ......................................................... 1 1.2 Instrument versions .............................................................. 1 1.3 About the documentation ................................................... 1 1.3.1 Symbols and conventions ........................................................ 1 1.4 Safety instructions .........................................................
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 7 Appendix 7.1 19 Functioning of the membrane pump ................................ 19 7.2 Remote interface ................................................................ 20 7.2.1 Pin allocation of the remote interface .................................... 20 8 Technical data 22 8.1 Membrane pump ................................................................ 22 8.2 Mains connection ............................................................... 22 8.
■■■■■■■■■■■■■■■■■■■■■■ Table of figures Table of figures Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Figure 7 Figure 8 Figure 9 Figure 10 Figure 11 Front of 843 Pump Station with membrane pumps ........................... 5 Rear 843 Pump Station with membrane pumps ................................. 6 Membrane pump with inlet and outlet tubing ................................... 7 Remote connection 869 — 843 — 8xx Titrino plus .........................
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction 1 Introduction 1.1 Instrument description The 843 Pump Station has the following characteristics: ■ Two membrane pumps. ■ Two remote connectors: – Remote 1 for connecting instruments which control the pumps. – Remote 2 for connecting other instruments. An integrated power supply unit (100 ... 240 V), which operates the membrane pumps. ■ 1.2 Instrument versions The 843 Pump Station is available in three different instrument versions: 1.3 2.843.
■■■■■■■■■■■■■■■■■■■■■■ 1.4 Safety instructions Cross-reference to figure legend The first number refers to the figure number, the second to the instrument part in the figure. Instruction step Carry out these steps in the sequence shown. Warning This symbol draws attention to a possible life hazard or risk of injury. Warning This symbol draws attention to a possible hazard due to electrical current. Warning This symbol draws attention to a possible hazard due to heat or hot instrument parts.
■■■■■■■■■■■■■■■■■■■■■■ 1.4.2 1 Introduction Electrical safety The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010. Warning Only personnel qualified by Metrohm are authorized to carry out service work on electronic components. Warning Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched.
■■■■■■■■■■■■■■■■■■■■■■ 1.4 Safety instructions 1.4.3 Working with liquids Caution Periodically check all system connections for leaks. Observe the relevant regulations in respect to working with flammable and/or toxic fluids and their disposal. 1.4.4 Chemical resistance Caution The instrument is not permitted to be utilized for conveying organic solvents and aqueous solutions that contain solids.
■■■■■■■■■■■■■■■■■■■■■■ 2 Overview of the instrument 2 Overview of the instrument 2.1 Front of the instrument Figure 1 Front of 843 Pump Station with membrane pumps 1 Pump 1 For switching membrane pump 1 on and off. 2 Pump 2 For switching membrane pump 2 on and off. 3 Mains switch/Emergency stop switch For switching the instrument on and off. 4 Tubing guide For fixing the tubings in place. 5 Membrane pump 1 For rinsing. 6 Membrane pump 2 For aspiration.
■■■■■■■■■■■■■■■■■■■■■■ 2.2 Rear panel of the instrument 2.2 Rear panel of the instrument 1 2 3 4 5 6 Figure 2 Rear 843 Pump Station with membrane pumps 1 Tubing guide For fixing the tubings in place. 2 Bore hole Additional hole for screwing in the tubing guide in situations where Pump Station is to be operated in a lying position. 3 Remote 1 For connecting an instrument that controls the pump. 4 Remote 2 For connecting other system instruments.
■■■■■■■■■■■■■■■■■■■■■■ 2.3 2 Overview of the instrument Membrane pump 1 2 6 3 7 4 8 6 5 Figure 3 Membrane pump with inlet and outlet tubing 1 Outlet tubing (e. g. 6.1812.000) 2 Union nut 3 Valve opening with protective cap 4 Specification of flow direction 5 Inlet tubing (e. g. 6.1812.
■■■■■■■■■■■■■■■■■■■■■■ 3.1 Setting up the instrument 3 Installation 3.1 Setting up the instrument 3.1.1 Packaging The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument. 3.1.2 Checks Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note. 3.1.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation Mounting tubing Mount the tubing as follows: 1 Cut the inlet and outlet tubing to appropriate lengths. 2 Unscrew the union nuts at both tube connections of the membrane pump. 3 Pull the union nut over the end of the tubing that is to be fastened.
3.2 Mounting tubing ■■■■■■■■■■■■■■■■■■■■■■ 4 Mount the inlet and outlet tubing on the connection nipples of the tube connections. The tubings should be pushed up against the stops, in order to ensure that the tubing connections are secure. 5 Guide the union nuts to the tube connection and screw them tightly in order to ensure a secure tubing connection. The openings of the PTFE tubing may need to be widened with a sharp object (e.g. with a screwdriver) for fixation purposes.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation 7 Insert the tubings through the tubing guide in order to fix them in place. 3.3 Connecting control devices Various possibilities are shown below for connecting different instruments to the 843 Pump Station. Instruments, which control the pumps of the Pump Station, are connected to Remote 1 , while all other instruments are connected to Remote 2. The figures also contain the order numbers of the required cables.
■■■■■■■■■■■■■■■■■■■■■■ 3.3 Connecting control devices Connection of one Dosimat plus, one Compact Sample Changer (Remote 1) and one Titrino plus (Remote 2). 869 Compact Sample Changer 843 Pump Station Dosimat plus Titrino plus 6.2141.26 0 6.2141.23 6.2141.24 0 0 Figure 5 6.2141.23 0 Remote connection 8xx Dosimat plus — 869 — 843 — 8xx Titrino plus or Connection of one Compact Sample Changer (Remote 1) and one Titrino plus and one Dosimat plus (both to Remote 2).
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation Connection of one Compact Autosampler (Remote 2) and one VA Computrace (Remote1). 863 Compact Autosampler 843 Pump Station 797 VA Computrace 6.2141.23 0 Figure 7 6.2141.28 0 Remote connection 863 — 843 — 797 Example 4 Connection of one Advanced Sample Processor (Remote 2) and one VA Computrace (Remote 1). 838 Advanced Sample Processor 843 Pump Station 797 VA Computrace Ext. Pump 1 Ext.
■■■■■■■■■■■■■■■■■■■■■■ 3.4 Connecting mains cable 3.4 Connecting mains cable Warning This instrument may not be operated except with the mains voltage specified for it (see rear side of the instrument). Protect the connection sockets against moisture.
■■■■■■■■■■■■■■■■■■■■■■ 4 Operation 4 Operation The control of the Pump Station is handled via remote signal by the 869 Compact Sample Changer or the 797 VA Computrace. It can however also be operated manually. Mains switch/ Emergency stop Switching the Pump Station on and off. Press the switch for an emergency stop. Illumination of the LED indicates ready-for-operation status. Pump 1 Hold down the membrane pump 1 switch for manual rinsing.
■■■■■■■■■■■■■■■■■■■■■■ 5.1 General information 5 Handling and maintenance 5.1 General information 5.1.1 Care The 843 Pump Station requires appropriate care. Excess contamination of the instrument may result in functional disruptions and a reduction in the service life of the sturdy mechanics and electronics. Spilled chemicals and solvents must be removed without delay.
■■■■■■■■■■■■■■■■■■■■■■ 5.2 5 Handling and maintenance Tubings The tubing must be inspected regularly to ensure trouble-free operation: ■ ■ ■ 5.3 do not kink PTFE tubing. Take resistance to solvents into account . Examine the tubing for possible damage on a regular basis. Quality Management and validation with Metrohm Quality Management Metrohm offers you comprehensive support in implementing quality management measures for instruments and software.
■■■■■■■■■■■■■■■■■■■■■■ 6.1 Problems 6 Troubleshooting 6.1 Problems Problem Cause Remedy The liquid is not aspirated, but just "bubbles" instead. The inlet and outlet tubing are switched. The outlet tubing is submerged in the solution to be aspirated instead of the inlet tubing. Switch the tubing. Inlet tubing into the solution and outlet tubing into the waste container. A connected instrument does not run. Mains cable is not connected. Connect mains cable.
■■■■■■■■■■■■■■■■■■■■■■ 7 Appendix 7 Appendix 7.1 Functioning of the membrane pump Figure 10 Functioning of the membrane pump 1 Outlet valve 2 Inlet valve 3 Conveyor space 4 PTFE membrane 5 Eccentric Principle The membrane pump is located in the pump head (see Chapter 2.3, page 7). The membrane 4 made of PTFE is moved upward and downward by means of an eccentric 5. These results in periodic volumetric changes in the conveyor space 3.
■■■■■■■■■■■■■■■■■■■■■■ 7.2 Remote interface Note Before utilizing any particularly aggressive media, it is imperative that you determine whether or not the materials mentioned above are resistant to them In addition, the membrane pump has an exceptionally long service life. The only parts subject to wearing are the membrane and the valve plates. The dosing precision of the membrane pump is significantly enhanced by a valve (see Figure 3, page 7) on the pump outlet.
■■■■■■■■■■■■■■■■■■■■■■ 7 Appendix Table 1 Functions and connections of the remote interfaces Remote 1 Remote 2 Pin No. Function/Connection Pin No. 1 ——— 8 2 ——— 2 3 Pump 1 - 4 Pump 2 - 5 ——— 9 6 0 Volt (GND) 6 - - 8 ——— 1 9 ——— 5 Inputs (Pin 3 and 4) +5 V approx.
■■■■■■■■■■■■■■■■■■■■■■ 8.1 Membrane pump 8 Technical data 8.1 Membrane pump >450 mL/min (with a pressure height of 2 m) Delivery rate 8.2 Mains connection Voltage 100…240 V Frequency 50…60 Hz Power consumption 45 W Fuse 1.0 ATH 8.3 Safety specifications Design and testing ■ ■ ■ ■ ■ Safety instructions 8.4 EN/IEC 61010-1: 2001 UL 61010-1: 2004 CSA-C22.2 No.
■■■■■■■■■■■■■■■■■■■■■■ 8 Technical data ■ ■ 8.5 EN/IEC 61000-4-14:2004 NAMUR:2004 Ambient temperature Nominal function range +5…+45 °C Storage –40…+70 °C Transport –40…+70 °C 8.6 Reference conditions Ambient temperature +25 °C (± 3 °C) Relative humidity ≤ 60 % 8.7 Dimensions Width 0.16 m Height 0.40 m Depth 0.31 m Weight (without accessories) 5.26 kg Material Polyurethane, painted (upper part) Steel housing, stove-enameled (lower part) 8.
■■■■■■■■■■■■■■■■■■■■■■ 9.1 Declaration of Conformity 9 Conformity and warranty 9.1 Declaration of Conformity This is to certify the conformity to the standard specifications for electrical appliances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company.
■■■■■■■■■■■■■■■■■■■■■■ 9 Conformity and warranty Metrohm Ltd. is holder of the SQS certificate ISO 9001:2000 Quality management system for development, production and sales of instruments and accessories for ion analysis. Herisau, 19 September 2007 D. Strohm Ch. Buchmann Vice President, Head of R&D Vice President, Head of Production Responsible for Quality Assurance 9.2 Quality Management Principles Metrohm Ltd.
■■■■■■■■■■■■■■■■■■■■■■ 9.3 Warranty (guarantee) Manufacture The measures put into practice in the production of our instruments guarantee a constant quality standard. Production planning and manufacturing procedures, maintenance of production means and testing of components, intermediate and finished products are prescribed. Customer support and service Customer support involves all phases of instrument acquisition and use by the customer, i.e.
■■■■■■■■■■■■■■■■■■■■■■ 10 Accessories 10 Accessories 10.1 Scope of delivery Note Subject to change without notice. 10.1.1 843 Pump Station with Membrane Pumps 2.843.0020 Order no. Description 1 1.843.0020 843 Pump Station with membrane pumps 4 6.1801.120 PVC tubing / 4 mm / 6 mm / 2 m Qty. Material: Outer diameter (mm): Inner diameter (mm): Length (m): 2 6.1812.000 PTFE tubing 4/6 mm, 4m For rinsing on Sample Changers Material: Outer diameter (mm): Inner diameter (mm): 2 6.1828.000 6.
■■■■■■■■■■■■■■■■■■■■■■ 10.1 Scope of delivery Qty. Order no. Description Cable plug according to customer requirements. Switzerland: Type SEV 12 6.2122.020 Germany, …: Type CEE(7), VII 6.2122.040 USA, …: Type NEMA/ASA 6.2122.070 1 10.1.2 8.843.8001EN 843 Pump Station with membrane pumps Manual 843 Pump Station with Membrane Pumps (Rinsing and aspiration equipment) 2.843.0030 Order no. Description 1 1.843.0020 843 Pump Station with membrane pumps 1 6.1543.120 Aspiration tip / M8 thread Qty.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 2 10 Accessories Order no. Description 6.1621.100 Cap for 10-L container Cap for 10-L container. For easy and safe transportation of filled containers. Material: PE Material remark: Cap for 10-L canister Height (mm): 19 Outer diameter (mm): 26.7 4 6.1801.120 PVC tubing / 4 mm / 6 mm / 2 m Material: Outer diameter (mm): Inner diameter (mm): Length (m): 1 6.1805.200 PTFE tubing / M8 / 0.5 m Material: Outer diameter (mm): Inner diameter (mm): Length (mm): 3 6.1805.
■■■■■■■■■■■■■■■■■■■■■■ 10.1 Scope of delivery Qty. 1 Order no. Description 6.1808.240 Distributor Rinsing / Aspiration for 869 Compact Sample Changer Distributor with 3 x M6, 1 x M8 and 2 x 1/8 connection for the usage of rinse and / or aspiration equipment on the 869 Compact Sample Changer. 2 6.1812.000 PTFE tubing 4/6 mm, 4m For rinsing on Sample Changers. Material: PTFE Outer diameter (mm): 6 Inner diameter (mm): 4 2 6.1828.000 PVDF connection nipple For 6.1621.
■■■■■■■■■■■■■■■■■■■■■■ Qty. Order no. 10 Accessories Description 6.2122.040 Type NEMA/ASA 6.2122.070 USA, …: 1 10.1.3 8.843.8001EN 843 Pump Station with membrane pumps Manual 843 Pump Station with Membrane Pumps for VA applications 2.843.0040 Order no. Description 1 1.843.0020 843 Pump Station with membrane pumps 4 6.1446.040 Thread stopper / M6 Qty. Material: Height (mm): Outer diameter (mm): 1 6.1456.210 PVDF 21.5 4.
■■■■■■■■■■■■■■■■■■■■■■ 10.1 Scope of delivery Qty. 2 Order no. Description 6.1621.000 Container 10 L As rinsing or waste container in automated systems. Material: PE Width (mm): 265 Height (mm): 400 Volume (mL): 10000 1 6.1805.020 FEP tubing / M6 / 52 cm With light and kink protection Material: Inner diameter (mm): Length (mm): 1 6.1805.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 10 Accessories Order no. Description 6.1805.180 FEP tubing / M6 / 16 cm With light and kink protection Material: Inner diameter (mm): Length (mm): 4 6.1805.530 FEP tubing / M6 / 2 m With light and kink protection Material: Inner diameter (mm): Length (mm): 1 6.1808.
■■■■■■■■■■■■■■■■■■■■■■ 10.1 Scope of delivery Qty. 4 Order no. Description 6.1820.010 Screw connector Connection piece with gas thread 1/8 in. and M6 thread for 6.1805.XXX tubing. For 683 Rinsing Unit. 1 6.1828.020 5-way tubing connector M6 for canister For connecting up to 5 lengths of tubing Material: FEP Material 2: ETFE 1 6.1829.020 FEP aspiration tube M6, 0.5 m For Dosing Unit Material: Inner diameter (mm): Length (mm): 1 6.2141.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 1 10 Accessories Order no. Description 6.2621.000 Adjustable wrench Max. opening: 20 mm. For IC instruments Length (mm): 150 1 6.2122.0x0 Mains cable with C13 line socket IEC-60320-C13 Cable plug according to customer requirements. Switzerland: Type SEV 12 6.2122.020 Germany, …: Type CEE(7), VII 6.2122.040 USA, …: Type NEMA/ASA 6.2122.070 1 8.843.8001EN 843 Pump Station with membrane pumps Manual 10.2 Optional accessories 10.2.
■■■■■■■■■■■■■■■■■■■■■■ 10.2 Optional accessories 10.2.2 843 Pump Station with Membrane Pumps (Rinsing and aspiration equipment) 2.843.0030 Order no. Description 6.2141.260 Remote cable Compact Sample Changer - Dosimat plus and Titrino plus Cable for remote connection between Compact Sample Changer and Dosimat plus and further devices (e.g. Titrino plus or Pump Station). 6.2141.
■■■■■■■■■■■■■■■■■■■■■■ 10 Accessories Order no. Description 6.1801.120 PVC tubing / 4 mm / 6 mm / 2 m Material: Outer diameter (mm): Inner diameter (mm): Length (m): 6.1808.010 PVC (transparent) 6 4 2 Coupling piece 2 X M6 2 M6 outer threads. Part of the screw connections for Exchange Units and 676 Sample Changer Material: PCTFE 6.1812.000 PTFE tubing 4/6 mm, 4m For rinsing on Sample Changers Material: Outer diameter (mm): Inner diameter (mm): 6.1819.
■■■■■■■■■■■■■■■■■■■■■■ 10.2 Optional accessories Order no. Description 6.1820.020 Screw connector M6 Connection piece with M10 thread for 6.1812.XXX tubing and M6 thread for 6.1805.XXX tubing. Material: PVDF 6.1828.000 PVDF connection nipple For 6.1621.
■■■■■■■■■■■■■■■■■■■■■■ Index Index C Chemical resistance ................ 2, 4 Connecting Compact Sample Changer . . 11 Control devices ................... 11 Dosimat plus ...................... 11 Mains cable ........................ 14 Titrino plus ......................... 11 VA Computrace .................. 11 E Electrical safety ........................... 2 Electrostatic charge .................... 3 Emergency stop switch ......... 5, 15 F Front of the instrument .............. 5 G GLP .............