860 KF Thermoprep Manual 8.860.
Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 860 KF Thermoprep Manual 8.860.
Teachware Metrohm AG CH-9101 Herisau teachware@metrohm.com This documentation is protected by copyright. All rights reserved. Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents Table of contents 1 Introduction 1 1.1 Instrument description ......................................................... 1 1.1.1 Instrument components .......................................................... 1 1.1.2 Intended use ........................................................................... 2 1.2 About the documentation ................................................... 2 1.2.1 Symbols and conventions ...............................................
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 4.4 Modifying numerical values and settings ......................... 25 4.4.1 Selecting from a selection list ................................................. 25 4.4.2 Modifying numerical values ................................................... 25 4.5 Parameters .......................................................................... 26 4.6 Configuration ......................................................................
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 10.3 11 Accessories Index 860 KF Thermoprep Quality Management Principles ........................................ 46 48 11.1 Scope of delivery 2.860.0010 ............................................ 48 11.2 Optional accessories ...........................................................
■■■■■■■■■■■■■■■■■■■■■■ Table of figures Table of figures Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Figure 7 Figure 8 Figure 9 Figure 10 Figure 11 Figure 12 Figure 13 Figure 14 Figure 15 Figure 16 Figure 17 Figure 18 Figure 19 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 VI ■■■■■■■■ Front 860 KF Thermoprep ................................................................. 6 Rear 860 KF Thermoprep ...................................................................
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction 1 Introduction 1.1 Instrument description The 860 KF Thermoprep is used whenever the heating up of a sample and/or the thermal expulsion of moisture in solid substances or liquids is required. In combination with a coulometric or volumetric KF titrator, the 860 KF Thermoprep is the ideal analysis system for water determination in samples that contain disruptive components or from which moisture can be removed only with difficulty.
■■■■■■■■■■■■■■■■■■■■■■ 1.2 About the documentation 1.1.2 Intended use The 860 KF Thermoprep is designed for usage as an auxiliary device for sample preparation in analytical laboratories. Its main area of application is moisture determination according to Karl Fischer (coulometric or volumetric). The 860 KF Thermoprep enables the application of thermal gas extraction technology. The present instrument is suitable for processing chemicals and flammable samples.
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction Warning This symbol draws attention to a possible hazard due to heat or hot instrument parts. Warning This symbol draws attention to a possible biological hazard. Caution This symbol draws attention to a possible damage of instruments or instrument parts. Note This symbol marks additional information and tips. 1.3 Safety instructions 1.3.
■■■■■■■■■■■■■■■■■■■■■■ 1.3 Safety instructions Warning Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched. There are no parts inside the housing which can be serviced or replaced by the user. Mains voltage Warning An incorrect mains voltage can damage the instrument. Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction Warning There is a considerable risk of injury connected with the needles. Grasp the handle of the needle adapter with both hands when you guide the needle into the sample vessel. Take care to ensure that no other person reaches into the danger zone during this procedure. 1.3.4 Flammable samples and solvents Warning All relevant safety measures are to be observed when working with flammable samples or when adding solvents. ■ ■ ■ ■ ■ 1.3.
■■■■■■■■■■■■■■■■■■■■■■ 2 Overview of the instrument 1 8 2 3 4 5 9 6 10 11 7 12 13 Figure 1 Front 860 KF Thermoprep 1 Guide head 2 Needle holder (6.2049.040) With needle adapter. 3 Spring sleeve With vessel stripper. 4 Guidance rod 5 Sample vessel (6.2419.007) With 6 ml content. 6 Sample insert (6.2063.010) 7 Oven module With sheet metal housing. 8 Fixing lever 9 Operating unit With display and keyboard. 10 Drying flasks (6.1608.050) With 6.1602.145 drying flask insert.
■■■■■■■■■■■■■■■■■■■■■■ 2 Overview of the instrument 8 1 2 3 7 4 Figure 2 5 6 Rear 860 KF Thermoprep 1 Air pump inlet With 6.2724.010 dust filter. 2 Mains switch 3 Mains connection socket 4 Type plate Contains specifications concerning mains voltage and serial number. 5 Outlet heating connection For the heating tube. 6 Air/nitrogen connector With M6 interior thread. Inlet for external gassing. 7 Fan For cooling the oven module. 8 Cable holder (6.2060.
■■■■■■■■■■■■■■■■■■■■■■ 3.1 Setting up the instrument 3 Installation 3.1 Setting up the instrument 3.1.1 Packaging The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument. 3.1.2 Checks Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note. 3.1.
■■■■■■■■■■■■■■■■■■■■■■ 3.3 3 Installation Mounting the guidance rod The guidance rod with the guide head is enclosed separately with the 860 KF Thermoprep and must first be mounted. Proceed as follows: 1 Move the guide head of the guidance rod a bit downwards by pressing the fixing lever. 3 1 Figure 4 2 Loosen the adapter 2 Loosen the adapter on the head of the guidance rod with a hexagon key. 3 Remove the adapter. 4 Insert the tubing jutting out of the housing from below into the guidance rod.
■■■■■■■■■■■■■■■■■■■■■■ 3.3 Mounting the guidance rod 5 4 Figure 5 Mounting the guidance rod 5 Insert the guidance rod into the opening intended for this purpose from above and push it downwards until it hits the stop. Observe the orientation of the guidance rod. 6 Pull out the tubing inside the guidance rod as far as possible. Eventually use a piece of sand paper in order to be able to better grasp the tubing. 7 6 Figure 6 Mounting tubing 7 Fasten the tubing on the olive of the 4.860.4260 adapter.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation 8 9 10 Figure 7 Fix the adapter 9 Fix the adapter in the guidance rod. Fasten the screw in the guidance rod with the aid of a hexagon key. 10 Carefully push the guidance rod downwards. 11 Figure 8 Guidance rod mounted 11 Fix the guidance rod with the V.0434.008 hexagon screw according to the drawing. The screw and the associated hexagon key are packaged separately with the instrument.
■■■■■■■■■■■■■■■■■■■■■■ 3.4 Mounting the sample insert 3.4 Mounting the sample insert 6.2063.010 6.2627.000 Figure 9 Mounting the sample insert Note The sample insert may not be inserted or removed unless it has been cooled down. The dimensions of the supplied sample insert are optimized for the usage of 6 mL sample vials 6.2419.007 from Metrohm. This ensures the best possible transfer of heat between oven and sample. 1 Place the 6.2627.000 oven insert into the oven from above. 2 Place the 6.2063.
■■■■■■■■■■■■■■■■■■■■■■ 3.5 3 Installation Mounting the needles The length of the needle holder defines how deeply the injection needle penetrates the sample vessel. The 6.2049.040 needle holder, which is 58 mm in length, ensures that the needle penetrates the liquid or powdery sample. The carrier gas can flow through the sample and effect an efficient expulsion of the moisture it contains. If there is a danger that the heated sample could clog the needle, then use the 6.2049.
■■■■■■■■■■■■■■■■■■■■■■ 3.6 Mounting the heating tubing 2 Screw the 6.2816.080 outlet needle onto the Luer connector of the needle holder. 3 Carefully guide the 6.2816.070 injection needle into the opening of the distributor from above and allow it to drop down. Note Take care to ensure that the white PTFE seal is positioned securely on the needle. 4 Screw the 6.1805.100 FEP tubing by hand onto the opening of the distributor.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation 2 Use the accompanying knurled screw to screw the 6.2060.010 cable holder tightly onto the guide head from below. 3 Guide the heating tubing cable through the eyelet of the cable holder. Note The cable holder prevents the cable of the heating tubing from coming into contact with parts of the housing that are hot. 4 Connect the heating tubing cable to the Outlet heater connector on the rear panel of the instrument.
■■■■■■■■■■■■■■■■■■■■■■ 3.7 Mounting the dust filter 3.7 Mounting the dust filter The built-in air pump must be protected against dust. A 6.2724.010 dust filter must be placed on the air inlet (Inlet filter) on the right-hand side of the housing for this reason. 6.2724.010 Figure 13 Mounting the dust filter The dust filter should be replaced once a year. 3.8 Assembling the drying flask Two drying flasks with desiccant are integrated into the gas flow in order to dry the gas that is conveyed. Dust (e.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation 2 Screw one 6.1821.040 filter tube into each of the 6.1602.145 drying flask covers from below. Tighten the filter tubes well by hand. 3 Screw the two drying flask covers with the filter tubes onto the drying flasks. Tighten the covers well by hand. Note If drying flask covers or filter tubes are not sufficiently tightly screwed on, then this will prevent a precise, regular flow of gas.
■■■■■■■■■■■■■■■■■■■■■■ 3.9 Air/nitrogen connector 3 Screw the second 6.1805.080 FEP tubing (25 cm length) to the gas inlet (at the rear, labeled 2). Screw the other tubing end on the rear drying flask to the M6 connector with point marking (see right-hand arrow). 4 Screw the 6.1805.010 FEP tubing (13 cm length) tightly onto the remaining M6 connectors of the drying flasks. The marking Drying flask on the right-hand side of the instrument displays the diagram for the tubing.
■■■■■■■■■■■■■■■■■■■■■■ 3.10 3 Installation Inserting the heating tube into the KF titration cell The 860 KF Thermoprep can be used for either coulometric or volumetric water determinations. The mounting of the tip of the heating tube is accomplished in different ways for the respective KF titration cells. Coulometric KF titration cell ■ ■ Remove the protective cover from the tip of the heating tube and the E.3010.032 O-ring. Disassemble the accompanying 6.1446.170 heating tube stopper into three parts.
■■■■■■■■■■■■■■■■■■■■■■ 3.10 Inserting the heating tube into the KF titration cell 5 Shift the tip vertically in such a way that the outlet opening of the tubing is immersed as deeply as possible. The tip must not however be permitted to get in the way of the stirring bar in the KF titration cell. Afterwards give the heating tube stopper its final tightening. Volumetric KF titration cell ■ ■ Remove the protective cover from the tip of the heating tube and the E.3010.032 O-ring.
■■■■■■■■■■■■■■■■■■■■■■ 3 Installation 5 Shift the height of the tip of the heating tube vertically in such a way that the outlet opening of the tubing is immersed as deeply as possible. The tip must not however be permitted to get in the way of the stirring bar in the KF titration cell. Afterwards give the septum stopper its final tightening.
■■■■■■■■■■■■■■■■■■■■■■ 4.1 Switching the instrument on and off 4 Operation 4.1 Switching the instrument on and off The rocker switch for switching on and off is located at the rear of the instrument. Figure 19 Mains switch The oven switches on automatically after the instrument is switched on and is heated up to the temperature that was most recently set. The outlet heating of the heating tube is also switched on at the same time. 4.
■■■■■■■■■■■■■■■■■■■■■■ 4 Operation [◀] Reduces gas flow by 1 mL/min during ongoing operation. Keep the key pressed down briefly for this purpose. [▶] Increases gas flow by 1 mL/min during ongoing operation. Keep the key pressed down briefly for this purpose. [▲] Reduces the brightness of the display. [▼] Increases the brightness of the display. 4.3 The display 4.3.1 Display elements 1 3 4 2 5 Figure 21 Main dialog 1 Temperature display Shows the measured oven temperature.
■■■■■■■■■■■■■■■■■■■■■■ 4.3 The display Symbol Significance An error has occurred. The type of the error will be identified in the status display. 4.3.3 Error messages Table 2 Error messages Error message Cause Remedy Flow rate error The tubing system is leaking or the flow rate is set too high. Screw all of the tubing connections and the screw connectors of the drying flasks tightly. Check the functioning of the pump. Max. temp. reached Adj.
■■■■■■■■■■■■■■■■■■■■■■ 4.4 4 Operation Modifying numerical values and settings The instrument settings can be modified in two different ways: ■ ■ Selection of specified values from a list. Parameters with list boxes are indicated with a colon (:). Modifying numerical values. The different parameters can be selected with the arrow keys [▼] and [▲]. The current line is displayed in inverted fashion. 4.4.1 Selecting from a selection list To modify parameters with list boxes (e.g.
■■■■■■■■■■■■■■■■■■■■■■ 4.5 Parameters 4.5 Parameters [OK] The instrument parameters can be adjusted in a menu that is opened by pressing the [OK] key. Exit the Parameters menu with [QUIT]. Note Modified values are applied after confirmation with the [OK] key. A modification is discarded with [QUIT]. Temp. 1 Set value for temperature 1. Range Default value 50 ... 250 °C (Increment: 1) 100 °C Temp. 2 Set value for temperature 2. Range Default value 50 ...
■■■■■■■■■■■■■■■■■■■■■■ 4 Operation Only when 'gas type' = 'other'. Range 0.001 ... 2.000 (Increment: 0.001) Default value 1.000 Table 3 Measuring factors of different gases Gas type Measuring factor Argon 0.950 Helium 0.300 Oxygen 1.000 Air 1.000 Nitrogen 1.000 Flow source Selecting the flow source. Selection Default value pump | valve pump pump Built-in air pump. valve Inlet valve for inert gas. 4.
■■■■■■■■■■■■■■■■■■■■■■ 4.6 Configuration in addition to switching the gas flow on and off. Furthermore it is still possible to modify the brightness of the display. Selection Default value off | on off Temp.corr Temperature correction. The target value for the temperature control of the oven can be influenced with this correction value. A temperature differential between the oven module and the sample in the vessel can thus be brought into line as needed. Range Default value -10 ...
■■■■■■■■■■■■■■■■■■■■■■ 5 Performing a determination 5 Performing a determination Moisture determinations according to the heating method require the conditioning of the entire tubing system and of the KF titration cell before a determination can be carried out. Because the sample vials and the septum seals may contain small amounts of moisture, three to five blank value determinations must be carried out with sealed, empty vessels.
■■■■■■■■■■■■■■■■■■■■■■ 5.1 Conditioning the system It is advantageous to have air bubbles stirred into the working medium and forcefully broken up in the process. The air in the KF titration cell must also be dried. Conditioning the working medium 1 Load a conditioning method on the KF titrator and begin the conditioning by pressing the [Start] key. You will find the necessary settings in the KF titrator manual.
■■■■■■■■■■■■■■■■■■■■■■ 5 Performing a determination Figure 22 Insert the sample vessel into the oven Inserting the needle 1 Grip the guide head with both hands and press the fixing lever. Figure 23 Move the guide head down 2 Move the guide head down and insert the needle into the sample vessel.
■■■■■■■■■■■■■■■■■■■■■■ 5.1 Conditioning the system Figure 24 Insert the needle Switching on the gas flow 1 Press the [OK] key and adjust the flow rate (Gas flow), e. g. 50 mL/min. Use the arrow keys [▶] or [◀] to select the number of decimal places, and then [▲] or [▼] to change the value. 2 Confirm the entry by pressing [OK] and select Flow source. 3 Using the arrow keys [▶] or [◀], select pump or valve (if you are using an external carrier gas) and confirm by pressing [OK].
■■■■■■■■■■■■■■■■■■■■■■ 5 Performing a determination Note The signal drift on the KF titrator should be less than 20 µg/min (or µL/min) and remain stable for at least 2 minutes before the blank value determination can be started. 5.2 Determining the blank value Because the surface of the sample vessels and the air enclosed in the vessel could contain moisture, it is imperative that three to five blank value determinations be carried out.
■■■■■■■■■■■■■■■■■■■■■■ 5.3 Determining samples Note Caution! The vessel is hot. 4 Place the blank into the oven using the crimping tongs. 5 Use both hands to move the guide head downward and insert the needle into the blank. Waiting for the determination 1 Wait until the determination is finished. Conditioning will then be resumed automatically. 2 Repeat the blank value determination with the next blank. 5.
■■■■■■■■■■■■■■■■■■■■■■ 5 Performing a determination Starting the sample determination 1 Start the determination by pressing the [Start] key on the KF titrator. 2 Use both hands to move the guide head of the 860 KF Thermoprep upward while pressing the fixing lever. Refer to the illustrations on pages 31 ff. 3 Remove the blank from the oven with the crimping tongs. Note Caution! The vessel is hot. 4 Place the sample vessel into the oven using the crimping tongs.
■■■■■■■■■■■■■■■■■■■■■■ 6.1 General 6 Handling and maintenance 6.1 General The 860 KF Thermoprep requires appropriate care. Excess contamination of the instrument may result in functional disruptions and a reduction in the service life of the sturdy mechanics and electronics of the instrument. Severe contamination can also have an influence on the measured results. Regular cleaning of exposed parts can prevent this to a large extent. Spilled chemicals and solvents must be removed immediately.
■■■■■■■■■■■■■■■■■■■■■■ 6 Handling and maintenance addition, various application bulletins are also available on the subject, which also contain Standard Operating Procedures (SOP = Standard Operating Procedure) for testing analytical measuring instruments for reproducibility and correctness. Maintenance Electronic and mechanical functional groups in Metrohm instruments can and should be checked as part of regular maintenance by specialist personnel from Metrohm.
■■■■■■■■■■■■■■■■■■■■■■ 7.1 Problems and their solutions 7 Troubleshooting 7.1 Problems and their solutions Problem Cause Remedy The drift is very high during conditioning. Molecular sieve of drying flasks and/or the titration cell exhausted. Replace molecular sieve. The titration cell is leaking Check seals. If necessary, replace. The sample is non-homogenous Reduce the size of the sample before weighing in. Not the entirety of the moisture expelled.
■■■■■■■■■■■■■■■■■■■■■■ 8 Appendix 8 Appendix 8.1 Practical instructions Carrier gas selection If the hot sample is sensitive to air or oxygen (decomposition) and releases substances that disrupt the KF reaction, then nitrogen (N2) should be used as the carrier gas. Temperature settings The temperature selected should be as high as the sample allows (high temperature = shorter analysis time). The sample may not however decompose. It must not give off any oxidizable substances, only water.
■■■■■■■■■■■■■■■■■■■■■■ 8.2 Literature 8.2 Literature ■ ■ 40 ■■■■■■■■ E. Scholz, Hydranal®-Manual, Riedel-de Haën, 1996 P. Bruttel, R. Schlink, Water determination by Karl Fischer Titration. Monograph, Metrohm, 2006, Best. Nr. 8.026.
■■■■■■■■■■■■■■■■■■■■■■ 9 Technical data 9 Technical data 9.1 Oven Temperature range 50…250 °C Accuracy ±3 °C Correction range -10…+10 °C Heating cartridge performance 165 W typical Depending on mains voltage Heating rate Typically 15 °C/min (with 80…180 °C, 230 V) Dependent on temperature, mains voltage, sample amount and vessel dimensions Cooling rate Typical. 9 °C/min (at 180…80 °C) Dependent on temperature, sample amount and vessel dimensions 9.
■■■■■■■■■■■■■■■■■■■■■■ 9.4 Mains connection 9.4 Mains connection Voltage 100…120 / 220…240 V Frequency 50 / 60 Hz Power consumption 200 W Fuse 2.0 ATH 9.5 Safety specifications Design and testing According to EN/IEC 61010-1: 2001, UL 61010-1: 2004, CSA-C22.2 No. 61010-1: 2004, EN/IEC 61010-2-010: 2003, protection class Ⅰ Safety instructions This document contains safety instructions which have to be followed by the user in order to ensure safe operation of the instrument. 9.
■■■■■■■■■■■■■■■■■■■■■■ 9.7 9 Technical data Ambient temperature Nominal function range 5…45 °C Humidity < 80 % Storage –20…70 °C Transport –40…70 °C 9.8 Reference conditions Ambient temperature 25 °C (±3 °C) Relative humidity ≤60 % Mains voltage 230 V 9.9 Dimensions Width 0.44 m Height 0.18 m (without support rod) 0.50 m (with support rod) Depth 0.23 m Weight (without accessories) 8.
■■■■■■■■■■■■■■■■■■■■■■ 10.1 Declaration of Conformity 10 Conformity and warranty 10.1 Declaration of Conformity This is to certify the conformity to the standard specifications for electrical appliances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company. Name of commodity 860 Thermoprep Temperature-controlled oven unit for the sample treatment in coulometric or volumetric water determinations.
■■■■■■■■■■■■■■■■■■■■■■ 10 Conformity and warranty EN 61010-2-010 Manufacturer Particular requirements for laboratory equipment for the heating of materials Metrohm Ltd., CH-9101 Herisau/Switzerland Metrohm Ltd. is holder of the SQS-certificate ISO 9001:2000 Quality management system for development, production and sales of instruments and accessories for ion analysis. Herisau, 27 November, 2007 10.2 D. Strohm A.
10.3 Quality Management Principles 10.3 ■■■■■■■■■■■■■■■■■■■■■■ Quality Management Principles Metrohm Ltd. holds the ISO 9001:2000 Certificate, registration number 10872-02, issued by SQS (Swiss Association for Quality and Management Systems). Internal and external audits are carried out periodically to assure that the standards defined by Metrohm’s QM Manual are maintained.
■■■■■■■■■■■■■■■■■■■■■■ 10 Conformity and warranty tional Qualification and Performance Verification of the system components or in carrying out the System Validation for the quantitative determination of a substance in a given matrix.
■■■■■■■■■■■■■■■■■■■■■■ 11.1 Scope of delivery 2.860.0010 11 Accessories Note Subject to change without notice. 11.1 Qty. 1 Scope of delivery 2.860.0010 Order no. Description 1.860.0010 860 KF Thermoprep Instrument for the thermal extraction of humidity from solid and liquid samples. 1 6.1446.170 Heating tube stopper for coulometric KF vessels in connection with heatable tubing connection Material: PP Height (mm): 34 Outer diameter (mm): 13 1 6.1448.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 2 11 Accessories Order no. Description 6.1608.050 Drying bottle / 100 mL / GL 45 Material: Height (mm): Outer diameter (mm): Volume (mL): 1 6.1805.010 FEP tubing / M6 / 13 cm With light and kink protection Material: Inner diameter (mm): Length (mm): 2 6.1805.
■■■■■■■■■■■■■■■■■■■■■■ 11.1 Scope of delivery 2.860.0010 Qty. 1 Order no. Description 6.1805.100 FEP tubing / M6 / 40 cm With light and kink protection Material: Inner diameter (mm): Length (mm): 1 6.1808.040 FEP 2 400 Adapter M6 outer / M8 inner Outer thread M6, inner thread M8. Material: PTCFE 1 6.1808.050 Connector tubing nozzle M8 1 M8 outer thread and 1 tubing olive. E.g. for thermostat jacket of Exchange Units and stability measuring instruments. Material: PVDF Length (mm): 31.
■■■■■■■■■■■■■■■■■■■■■■ Qty. 2 11 Accessories Order no. Description 6.1821.040 Filter tube Filter tube for 6.1608.050 Drying bottle. For Rancimats and Karl Fischer Ovens. Length (mm): 112 1 6.1830.030 Heatable outlet tubing Heatable outlet tubing for 860 KF Thermoprep. 1 6.2048.030 Guidance rod Guidance rod for 860 KF Thermoprep. 1 6.2049.
■■■■■■■■■■■■■■■■■■■■■■ 11.1 Scope of delivery 2.860.0010 Qty. 1 Order no. Description 6.2060.010 Cable support Used with 860 KF Thermoprep. 1 6.2063.010 Sample holder for 6.2419.007 sample vials (6 mL) 1 6.2419.007 Sample vial 6 ml / 100 pieces 6.1448.050 septum caps. Used for KF-determinations with oven. Volume (mL): 6 1 6.2621.110 Crimping tongs Crimping tongs for sealing 6 mL sample vials (6.2419.000, 6.2419.007).
■■■■■■■■■■■■■■■■■■■■■■ Qty. 11 Accessories Order no. Description 1 6.2621.120 Hexagon key 1.5 mm 1 6.2621.130 Hexagon key 2 mm 1 6.2627.000 Oven insert Used with 860 KF Thermoprep. 1 6.2724.
■■■■■■■■■■■■■■■■■■■■■■ 11.1 Scope of delivery 2.860.0010 Qty. 1 Order no. Description 6.2739.000 Wrench For tightening connectors Length (mm): 1 6.2811.000 68 Molecular sieve Molecular sieve. Bottle containing 250 g. Pore size: 0.3 nm. Without moisture indicator. For Rancimats and Karl Fischer instruments. 1 6.2816.070 Injection needle Used with 832 KF Thermoprep. Material: 1 6.2816.080 Stainless steel (AISI 304) Outlet needle Used with KF Thermoprep and Oven Sample Processor.
■■■■■■■■■■■■■■■■■■■■■■ Qty. Order no. 11 Accessories Description Cable plug according to customer requirements. Switzerland: Type SEV 12 6.2122.020 Germany, …: Type CEE(7), VII 6.2122.040 USA, …: Type NEMA/ASA 6.2122.070 1 8.860.8003EN 11.2 860 KF Thermoprep Manual Optional accessories Order no. Description 6.1448.050 Aluminum septum caps / 1000 pieces Including Al sealing. Material: Material 2: Height (mm): Outer diameter (mm): Aluminum Silicone 7.4 20.7 6.2049.
■■■■■■■■■■■■■■■■■■■■■■ Index Index Numbers/Symbols 6.2627.000 oven insert ............ 12 A Arrow key .......................... 23, 25 Assembling Drying flasks ....................... 16 B Basic settings ............................ 27 Blank value ............................... 33 Brightness ................................ 23 C Cable holder ......................... 7, 15 Carrier gas ................................ 39 Colon ....................................... 25 Conditioning ......................