915 KF Ti-Touch Manual 8.915.
Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 915 KF Ti-Touch Program version 5.915.0030 Manual 8.915.
Teachware Metrohm AG CH-9100 Herisau teachware@metrohm.com This documentation is protected by copyright. All rights reserved. Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above. Documentation in additional languages can be found on http://documents.metrohm.com.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents Table of contents 1 Introduction 1 1.1 Instrument description ......................................................... 1 1.1.1 Titration and measuring modes ................................................ 1 1.1.2 Connectors .............................................................................. 2 1.1.3 Intended use ........................................................................... 2 1.2 About the documentation ...................................
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 4.6 Connecting sensors ............................................................ 19 4.6.1 General ................................................................................. 19 4.6.2 Connecting a polarizable electrode ........................................ 19 4.6.3 Connecting a temperature sensor .......................................... 20 4.7 Connecting the Ti-Touch to a network ............................ 21 5 Titrations 5.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 8.7 Titer determination options and data .............................. 63 8.7.1 Titer validity ........................................................................... 63 8.7.2 Properties of the previous titer determinations ....................... 64 9 Reagents 66 9.1 Editing reagent data .......................................................... 66 9.2 Reagent monitoring ........................................................... 67 10 Sensors 70 10.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 12 File manager 113 12.1 Managing files .................................................................. 113 12.1.1 Copying a file ...................................................................... 115 12.1.2 Renaming a file ................................................................... 115 12.1.3 File properties ...................................................................... 116 12.2 External storage medium ....................................
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 16.2 Saving a method ............................................................... 155 16.3 Loading a method ............................................................ 156 16.4 Editing a method .............................................................. 157 16.4.1 Inserting a command ........................................................... 158 16.5 Method options ................................................................ 159 16.5.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 23 Results and more determination data 198 23.1 More determination data ................................................ 199 23.1.1 Details ................................................................................. 200 23.2 Messages ........................................................................... 202 23.3 Local common variables .................................................. 202 23.4 Determination properties .............................
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 27.5 Remote .............................................................................. 248 27.6 USB Sample Processor ..................................................... 249 27.6.1 Moving the lift ..................................................................... 251 27.6.2 Moving to a rack position .................................................... 253 27.6.3 External positions ................................................................
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 28.8 Automation ....................................................................... 319 28.8.1 Rotating sample rack (MOVE) .............................................. 319 28.8.2 Moving the lift (LIFT) ............................................................ 321 28.8.3 Controlling pumps (PUMP) ................................................... 322 28.8.4 Resetting the rack (RACK) .................................................... 323 28.8.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents 31.7.2 31.7.3 31.7.4 31.7.5 31.7.6 31.7.7 31.8 Formatting an external storage medium ............................... 347 Removing an external storage medium ................................ 347 Adjusting the touch screen .................................................. 347 Testing the touch screen ...................................................... 348 Software update (loading program versions and language files) ..............................................
■■■■■■■■■■■■■■■■■■■■■■ Table of figures Table of figures Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Figure 7 Figure 8 Figure 9 Figure 10 Figure 11 Figure 12 Figure 13 Figure 14 Figure 15 Figure 16 Figure 17 Figure 18 XII ■■■■■■■■ Front 915 KF Ti-Touch ....................................................................... 8 Rear 915 KF Ti-Touch ........................................................................ 9 Connecting the power supply unit ..........................................
■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction 1 Introduction 1.1 Instrument description The 915 KF Ti-Touch is a compact titration system for volumetric Karl Fischer titration. This newly designed titrator combines in a single device the touch-sensitive color monitor for convenient and efficient operation, the titration unit, the magnetic stirrer and the integrated membrane pump for adding and aspirating solvents. The upper side of the housing offers space for the titrant and the titration cell.
■■■■■■■■■■■■■■■■■■■■■■ 1.1 Instrument description ■ 1.1.2 MEAS Measuring modes: – Ipol (voltametric measurement with selectable polarization current) – Upol (amperometric measurement with selectable polarization voltage) – T (temperature measurement) Connectors The 915 KF Ti-Touch is equipped with the following connectors: ■ ■ ■ ■ ■ ■ 1.1.3 Mains connection For connecting to the mains supply with the aid of the power supply unit provided.
■■■■■■■■■■■■■■■■■■■■■■ 1.2 1 Introduction About the documentation CAUTION Please read through this documentation carefully before putting the instrument into operation. The documentation contains information and warnings which the user must follow in order to ensure safe operation of the instrument. 1.2.
■■■■■■■■■■■■■■■■■■■■■■ 1.2 About the documentation NOTE This symbol highlights additional information and tips.
■■■■■■■■■■■■■■■■■■■■■■ 2 Safety instructions 2 Safety instructions 2.1 General notes on safety WARNING This instrument may only be operated in accordance with the specifications in this documentation. This instrument has left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully. 2.
■■■■■■■■■■■■■■■■■■■■■■ 2.3 Tubing and capillary connections Protection against electrostatic charges WARNING Electronic components are sensitive to electrostatic charges and can be destroyed by discharges. Do not fail to pull the mains cable out of the mains connection socket before you set up or disconnect electrical plug connections at the rear of the instrument. 2.3 Tubing and capillary connections CAUTION Leaks in tubing and capillary connections are a safety risk.
■■■■■■■■■■■■■■■■■■■■■■ 2.5 2 Safety instructions Recycling and disposal This product is covered by European Directive 2002/96/EC, WEEE – Waste from Electrical and Electronic Equipment. The correct disposal of your old equipment will help to prevent negative effects on the environment and public health. More details about the disposal of your old equipment can be obtained from your local authorities, from waste disposal companies or from your local dealer.
■■■■■■■■■■■■■■■■■■■■■■ 3.1 Front of the instrument 3 Overview of the instrument 3.1 Front of the instrument 12 13 11 1 8 9 10 2 Figure 1 3 4 5 6 7 Front 915 KF Ti-Touch 1 Display Touch-sensitive screen. 2 Fixed key [Home] Opens the main dialog. 3 Fixed key [Back] Saves the entry and opens the next-higher dialog page. 4 Fixed key [Help] Opens the online help for the dialog displayed. 5 Fixed key [Print] Opens the print dialog.
■■■■■■■■■■■■■■■■■■■■■■ 3 Overview of the instrument 11 Bottle holder With holding clips, for reagent bottle. 13 Titration stand With built-in magnetic stirrer and membrane pump for placement of the titration cell. 3.2 12 Support rod (lower part) For mounting the support rod (upper part). Rear of the instrument 8 9 10 1 2 Figure 2 3 4 5 6 7 Rear 915 KF Ti-Touch 1 USB connector (type A) For connecting a printer, USB flash drive, USB hub, USB Sample Processor etc.
■■■■■■■■■■■■■■■■■■■■■■ 4.1 Setting up the instrument 4 Installation 4.1 Setting up the instrument 4.1.1 Packaging The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument. 4.1.2 Checks Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note. 4.1.
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation Figure 3 Connecting the power supply unit Proceed as follows: 1 Connect the plug of the external power supply unit with the mains connection of the Ti-Touch (see Figure 3, page 11). NOTE The plug of the power supply unit is protected against accidental disconnection of the cable by means of a pull-out protection feature. If you wish to pull out the plug, you will first need to pull back the outer plug sleeve marked with arrows.
■■■■■■■■■■■■■■■■■■■■■■ 4.3 Connecting MSB devices 4.3 Connecting MSB devices In order to connect MSB devices, e.g. dosing device or Remote-Box, the Ti-Touch has two connectors at what is referred to as the Metrohm Serial Bus (MSB). Various peripheral devices can be connected in sequence (Daisy Chain) at a single MSB connector (8-pin Mini DIN socket) and be controlled simultaneously by the Ti-Touch.
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation Figure 5 MSB connector CAUTION Switch off the Ti-Touch before you plug in MSB devices. When it is switched on, the Ti-Touch automatically recognizes which device is connected to which MSB connector. The connected MSB devices are entered automatically in the device manager. MSB connections can be extended with the 6.2151.010 cable. The maximum connection length permitted is 6 m. 4.3.1 Connecting dosing devices You can connect two dosing devices to the Ti-Touch.
■■■■■■■■■■■■■■■■■■■■■■ 4.3 Connecting MSB devices 4.3.2 Connecting an additional stirrer or titration stand In addition to the built-in magnetic stirrer, you can also use the magnetic stirrers 801 Stirrer, 803 Ti Stand or the 804 Ti Stand with the propeller stirrer 802 Stirrer. Proceed as follows: 1 Switch off the Ti-Touch. 2 Connect the connection cable of the magnetic stirrer or of the titration stand to MSB 2 (2-5) on the rear of the Ti-Touch.
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation You will find precise information concerning the pin assignment of the interface on the Remote Box in Appendix (see Chapter 31.5, page 341). 4.4 Connecting USB devices 4.4.1 General The 915 KF Ti-Touch has a USB connector (Type A socket) for peripheral devices with USB interface and for USB Sample Processors. If you wish to connect more than one device to the USB, you can also use an additional commercially available USB hub.
■■■■■■■■■■■■■■■■■■■■■■ 4.4 Connecting USB devices 4.4.4 Connecting a balance If you wish to connect a balance to the Ti-Touch, you will require a USB/ RS-232 adapter (6.2148.050). The following table offers an overview of the balances that you can use together with the Ti-Touch and of which cable you will need for connection to the RS-232 interface: Balance Cable AND ER, FR, FX with RS-232 interface (OP-03) 6.2125.020 + 6.2125.
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation Connect the balance as follows: 1 Connect the USB plug of the USB/RS-232 adapter with the USB connector of the Ti-Touch (Type A). The USB/RS-232 adapter will be recognized automatically and entered in the device manager of the Ti-Touch. 2 Connect the RS-232 interface of the USB/RS-232 adapter with the RS-232 interface of the balance (see table for cable). 3 Switch on the balance. 4 If necessary, switch on the RS-232 interface of the balance.
■■■■■■■■■■■■■■■■■■■■■■ 4.4 Connecting USB devices 2 Enter and configure the barcode reader in the device manager of the (see Chapter 11.11, page 110). Settings on the barcode reader: Program the barcode reader as follows (also see manual for the barcode reader): 1 Switch the barcode reader to programming mode. 2 Specify the desired layout for the keyboard (USA, Germany, France, Spain, German-speaking Switzerland). This setting must match the setting in the device manager.
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation The USB Sample Processor will be recognized automatically and entered in the device manager of the Ti-Touch. NOTE The plug on the controller cable is protected against accidental disconnection by means of a pull-out protection feature. If you wish to pull out the plug, you will first need to pull back the outer plug sleeve marked with arrows. 3 Configure the USB Sample Processor in the device manager (see Chapter 11.11, page 110). 4.
■■■■■■■■■■■■■■■■■■■■■■ 4.6 Connecting sensors Figure 6 Connecting a polarizable electrode NOTE The electrode cable is protected against accidental disconnection of the cable by means of a pull-out protection. If you wish to remove the plug, you will first need to pull back the outer plug sleeve. 4.6.3 Connecting a temperature sensor A temperature sensor of the Pt1000 or NTC type can be connected to the Temp. connector.
■■■■■■■■■■■■■■■■■■■■■■ 4 Installation Figure 7 Connecting a temperature sensor NOTE The red plug must always be plugged into the red socket for the purpose of shielding against disruptions. 4.7 Connecting the Ti-Touch to a network The 915 KF Ti-Touch has a network connection (Ethernet). This can be used to integrate your Ti-Touch in your network. You can, for example, store data on a PC within the network or print reports on a network printer.
4.
■■■■■■■■■■■■■■■■■■■■■■ 5 Titrations 5 Titrations 5.1 Water determination according to Karl Fischer (KFT) Karl Fischer Titration is a method for volumetric water determination. Conditioning is carried out automatically both before and after the actual titration. The reagent dosing is controlled in such a way that a predefined endpoint is reached as quickly and as accurately as possible.
■■■■■■■■■■■■■■■■■■■■■■ 6.1 Switching the instrument on and off 6 Operation 6.1 Switching the instrument on and off Switching on the instrument CAUTION Peripheral devices (e.g. printers) must be switched on before you switch on the 915 KF Ti-Touch. Proceed as follows: 1 ■ ■ ■ Press the mains switch on the left-hand side of the rear panel of the 915 KF Ti-Touch. The 915 KF Ti-Touch is initialized. A system test is performed. This process takes some time.
■■■■■■■■■■■■■■■■■■■■■■ 6 Operation Switching off the instrument CAUTION The 915 KF Ti-Touch must be switched off by pressing the mains switch on the rear of the instrument before the electricity supply is interrupted. If this is not done, then there is a danger of data loss. Proceed as follows: 1 ■ Press the mains switch on the left-hand side of the rear panel of the 915 KF Ti-Touch. The current data is saved and the system is shut down. This process takes just a short time.
■■■■■■■■■■■■■■■■■■■■■■ 6.2 Fundamentals of operation 6.2 Fundamentals of operation 6.2.1 Touch-sensitive screen The entire Ti-Touch user interface is touch-sensitive. Simply touch a few of the buttons on the interface to learn how a touch-sensitive screen reacts. You can always return to the main dialog by touching [ ].
■■■■■■■■■■■■■■■■■■■■■■ 6 Operation [Back] saves the entry and opens the next-higher dialog page. [Help] opens the online help feature for the dialog displayed. [Print] opens the printing dialog. [Manual] opens the manual control. [Stop] cancels the ongoing determination. [Start] starts a determination. The method loaded, the time and the system status are displayed in the main dialog in the Title bar.
■■■■■■■■■■■■■■■■■■■■■■ 6.2 Fundamentals of operation The working medium is being conditioned. Conditioning has been paused. The working medium is conditioned. A method has been started. A method has been paused. An action has been started in manual control. 6.2.4 Entering text and numbers In the editing dialog for text or numerical input, enter the individual characters by tapping in the input field.
■■■■■■■■■■■■■■■■■■■■■■ 6 Operation Editing function Description [OK] The modification is applied and the editing dialog is exited. [Cancel] The editing dialog is exited without applying the modification. [Delete entry] The content of the input field is deleted completely. [⌫] The character in front of the cursor is deleted. [⇦] The cursor within the input field is shifted to the left by one character. [⇨] The cursor within the input field is shifted to the right by one character.
■■■■■■■■■■■■■■■■■■■■■■ 6.2 Fundamentals of operation Editing function Description [Cancel] The editing dialog is exited without applying the modification. [Delete entry] The content of the input field is deleted completely. [off] If not only numbers but also special values (e.g. off) can be entered, then the corresponding buttons will be shown to the right of the numerical keypad.
■■■■■■■■■■■■■■■■■■■■■■ 7 System settings 7 System settings Main dialog: System ▶ System settings This chapter describes the various system settings and configurations. ■ ■ ■ ■ ■ ■ Selecting the dialog language (see Chapter 7.1.1, page 31). Setting the date and time (see Chapter 7.1.2, page 32). Defining system-specific dialog options (see Chapter 7.2, page 33). User administration (see Chapter 7.3, page 37). Defining settings for the measured value display (see Chapter 7.4, page 45).
■■■■■■■■■■■■■■■■■■■■■■ 7.1 General system settings 2 Select the dialog language ■ Tap on the list box Dialog language and select the desired language. 3 Save the settings Tap on the fixed keys [ ] or [ ]. The main dialog is displayed in the respective dialog language. 7.1.2 Setting the date, time and local time The Ti-Touch displays the date and time in accordance with ISO standard 8601.
■■■■■■■■■■■■■■■■■■■■■■ 7 System settings 4 Enter the local time ■ Tap on the list box Local time - UTC and select the difference from the UTC (Coordinated Universal Time). The selection off means that the time is saved with no difference from the UTC. 5 Save the settings Tap on the fixed keys [ ] or [ ]. The time settings are saved. 7.
■■■■■■■■■■■■■■■■■■■■■■ 7.2 System-specific dialog options [Command list] Block unneeded method commands (see "Blocking unneeded commands and fixed keys", page 34). [Fixed keys] Block unneeded fixed keys (see "Blocking unneeded commands and fixed keys", page 34). [Routine dialog] Configure functions for the routine dialog (see "Configuring the routine dialog", page 35).
■■■■■■■■■■■■■■■■■■■■■■ 7 System settings Blocking commands Proceed as follows to block unneeded commands: 1 Display the command list Tap on the button [Command list]. The list of all command groups is displayed. 2 Deactivate command groups Deactivate those command groups which are not permitted to be used. 3 Save the settings Tap on the fixed keys [ ] or [ ]. All deactivated commands appear grayed out in the method editor and cannot be used for creating methods.
■■■■■■■■■■■■■■■■■■■■■■ 7.2 System-specific dialog options NOTE The configuration of the routine dialog applies for all routine users. You also have the option of defining user-specific routine settings. To do this, you must work with the login function activated and create an identification profile for each user (see Chapter 7.3.2, page 40). Proceed as follows to modify the configuration for the routine dialog: 1 Open the dialog Tap on the button [Routine dialog].
■■■■■■■■■■■■■■■■■■■■■■ 7 System settings All deactivated functions will appear grayed-out, i.e. they are inactive. 7.3 User administration Main dialog: System ▶ System settings ▶ User admin. All of the functions of the user administration are described in this chapter. No user administration is mandatory for the simple operation of a titration system. NOTE If you work with the login function activated, then the user administration is accessible only for users with administrator rights.
■■■■■■■■■■■■■■■■■■■■■■ 7.3 User administration ■ Operation without login function: If you work without the login function, the users whose names are entered in the list can be selected in the main dialog or a user name can be entered. This makes it possible to document who has operated the titration system. [Login options] Define the settings for the login (see Chapter 7.3.3, page 41). [Create ID profile] Create an identification profile for the selected user on a storage medium (see Chapter 7.3.
■■■■■■■■■■■■■■■■■■■■■■ 7 System settings User The designation of the user is used for unambiguous identification, e.g. the company internal shorthand symbol or the personal number. The user name is printed out in all reports containing determination data and stored in the determination file. Each file always contains the name of the user who created it and the name of the last user to edit it. Entry Default value 24 characters maximum empty Full name Complete name of the user.
■■■■■■■■■■■■■■■■■■■■■■ 7.3 User administration NOTE The last user with administrator rights cannot be deleted anymore. [Favorites] Save methods and sample tables as user specific favorites (see Chapter 18, page 174). 7.3.2 Creating an identification profile User list: User ▶ Create ID profile If you plan to carry out the login with an identification profile (see Chapter 7.3.3, page 41), then you must first create an identification profile on a storage medium for each user.
■■■■■■■■■■■■■■■■■■■■■■ 7.3.3 7 System settings Defining login options Main dialog: System ▶ System settings ▶ User admin. ▶ Login options There are a number of different ways to log onto the system: ■ ■ ■ ■ ■ Without login Login via user name Login via user name and password Login via identification profile Login via identification profile and password NOTE If you work with the login function activated, then the user administration is accessible only for users with administrator rights.
■■■■■■■■■■■■■■■■■■■■■■ 7.3 User administration Login via user name on | off (Default value: off) If this option is activated, then the user must log in with his or her unambiguous identification. Login via identification profile on | off (Default value: off) If this option is activated, then the login will take place via USB storage medium with the identification profile stored on it.
■■■■■■■■■■■■■■■■■■■■■■ 7 System settings Minimum password length Minimum number of characters of the passwords. Input range Default value 1 - 10 1 No. of entry attempts If the user has logged in incorrectly this many times, then it will automatically be deactivated. It can only be reactivated by a user with administrator rights.
■■■■■■■■■■■■■■■■■■■■■■ 7.3 User administration Forgotten password CAUTION If a user has forgotten his password, a new user name must be defined. The same user name can only be used again after a re-installation and re-creating the user list. Proceed as follows: 1 Create a backup Create a backup (see Chapter 12.3, page 119). 2 Initialize the system Switch off the Ti-Touch and carry out a system initialization (see Chapter 29.1, page 331).
■■■■■■■■■■■■■■■■■■■■■■ 7.4 7 System settings Measured value display Main dialog: System ▶ System settings ▶ Meas. value display The number of decimal places can be defined for pH values and voltages. This setting refers exclusively to the display of the measured values in the live display and in the manual control. The values will, however, always be stored with their full accuracy. 7.
■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants Main dialog: System ▶ Titrants This chapter describes how you can create a list of titrants used in the system. Titrants can be used in intelligent buret units or in non-intelligent buret units. Intelligent buret units have a built-in data chip on which the data for the titrant is stored. This data is automatically read out during attachment and entered in the titrant list. The titrant list can contain a maximum of 30 titrants.
■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants ■ ■ Data on exchange/dosing unit: – Parameters for the function PREP – Length and diameter of the tubings – Port assignment of the dosing unit – Cylinder volume – Serial number – etc. etc. NOTE If data is read out from the data chip, then a check is made whether the titrant list already contains a titrant of the same type with the identical serial number.
■■■■■■■■■■■■■■■■■■■■■■ 8.2 Editing titrant data 8.2 Editing titrant data Titrant list: Titrants ▶ New / Edit All of the data for the selected titrant is displayed in the dialog Titrants / Edit. Titrant The designation of the titrant is used for unambiguous identification. Entry Selection 24 characters maximum Selection of frequently used titrants Concentration Concentration of the titrant. Input range Default value –999999999 - 9999999999 1.000 Unit of the concentration.
■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants Input range Default value –999999999 - 9999999999 1.000 Unit of the titer.
■■■■■■■■■■■■■■■■■■■■■■ 8.3 Monitoring the working life 8.3 Monitoring the working life Titrant: Edit ▶ Working life In the dialog Edit titrant / Working life, you can define the time interval after which the titrant must be replaced. This is particularly important if your titrant has a limited working life. If you do not wish to monitor the working life, then you can enter only the date of manufacture for documentation purposes.
■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants Display message A message is displayed. You can select whether you want to continue with the determination or cancel the run. Document message In the determination data it will be documented, that the time interval has been expired. Cancel determination The determination is stopped. 8.4 Dosing unit Titrant: Edit ▶ Dosing unit You can edit data for the dosing unit of the titrant in this dialog. Name Designation of the exchange or dosing unit.
■■■■■■■■■■■■■■■■■■■■■■ 8.4 Dosing unit Cylinder volume Cylinder volume of the dosing unit. It will be read out automatically on dosing units with integrated data chips. Selection Default value 2 | 5 | 10 | 20 | 50 20 Serial no. cyl. Serial number of the dosing cylinder. It will be read out automatically on exchange or dosing units with integrated data chips. The number can be changed manually at any time, e.g. when a cylinder is replaced.
■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants Selection Default value maximum maximum maximum The maximum dosing rate depends on the cylinder volume (see Chapter 31.1, page 338). Dosing rate Dos. port 2 Rate used for the aspiration and ejection of the reagent via dosing port 2. Input range Selection Default value 0.01 - 166.00 mL/min maximum maximum maximum The maximum dosing rate depends on the cylinder volume (see Chapter 31.1, page 338).
■■■■■■■■■■■■■■■■■■■■■■ 8.4 Dosing unit NOTE These parameters are important for the correct execution of the Preparing (PREP command) and Emptying (EMPTY command) functions, because the volumes of the tubing connections are taken into account. Dosing port 1 Port Port to be used as dosing port 1 for the PREP and EMPTY (see Figure 10, page 56) functions. Selection Default value Port 1 | Port 2 | Port 3 | Port 4 Port 1 Length Length of the tubing. Input range Default value 0.0 - 999.9 cm 40.
■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants Diameter Diameter of the tubing. Input range Default value 0.0 - 9.9 mm 2.0 mm Fill port Port Port to be used as fill port for the PREP and EMPTY functions (see Figure 10, page 56). Selection Default value Port 1 | Port 2 | Port 3 | Port 4 Port 2 Length Length of the tubing. Input range Default value 0.0 - 999.9 cm 25.0 cm The setting 0.0 means that this tubing will neither be rinsed nor emptied. Diameter Diameter of the tubing. Input range Default value 0.0 - 9.
■■■■■■■■■■■■■■■■■■■■■■ 8.4 Dosing unit Input range Default value 0.0 - 9.9 mm 2.0 mm 4 5 1 2 3 Figure 10 Dosing unit – port assignment 1 VENT This port is set up for the deaeration of the reagent bottle. An adsorber tube (filled with desiccant) is usually mounted here. 2 Port 1 The default definition of this port is dosing port 1. 3 Port 2 The default definition of this port is the filling port. A riser tube is usually mounted to it.
■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants ascending The ports are moved to in ascending order. automatic The ports are moved to by the shortest path. not over Define a protected port. Not over port This parameter can only be edited with Direction = not over. Define a protected port if the valve disc is not to be rotated over this port (useful with pipetting functions). The protected port can, however, be moved to directly. Selection Default value 8.
■■■■■■■■■■■■■■■■■■■■■■ 8.5 Exchange unit Entry Default value 24 characters maximum empty Serial number Serial number of the exchange or dosing unit. It will be read out automatically on units with integrated data chips. Entry 8 digits maximum Cylinder volume Cylinder volume of the exchange unit. It will be read out automatically on exchange units with integrated data chips. Selection Default value 1 | 5 | 10 | 20 | 50 20 Serial no. cyl. Serial number of the dosing cylinder.
■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants Cylinder volume The entire cylinder volume is being dosed. Cycles Number of rinsing cycles. We recommend carrying out at least two rinsing cycles in order to remove all air bubbles. Selection Default value 1|2|3|4|5|6|7|8|9 2 Dosing rate Rate at which it is dosed. Input range Selection Default value 0.01 - 166.00 mL/min maximum maximum maximum The maximum dosing rate depends on the cylinder volume (see Chapter 31, page 338).
■■■■■■■■■■■■■■■■■■■■■■ 8.5 Exchange unit Diameter Diameter of the tubing. Input range Default value 0.0 - 9.9 mm 2.0 mm Dosing cylinder Tubing to the dosing cylinder (11-3). Length Length of the tubing. Input range Default value 0.0 - 999.9 cm 13.0 cm Diameter Diameter of the tubing. Input range Default value 0.0 - 9.9 mm 2.0 mm Reagent bottle Tubing to the reagent bottle (11-1). Length Length of the tubing. Input range Default value 0.0 - 999.9 cm 25.0 cm Diameter Diameter of the tubing.
■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants 1 2 3 Figure 11 1 Tubing to the reagent bottle 3 Tubing to the dosing cylinder 8.6 Exchange unit – tubing connections 2 Tubing to the dosing tip GLP test for exchange unit and dosing unit Titrant: Edit ▶ Dosing unit / Exchange unit ▶ GLP test In the dialog Exchange unit / GLP test or Dosing unit / GLP test, respectively, you can define the time interval after which a GLP test must be carried out again for the exchange unit or dosing unit.
■■■■■■■■■■■■■■■■■■■■■■ 8.6 GLP test for exchange unit and dosing unit GLP test interval If you define a time interval for the GLP test, then the date in Next GLP test will be tracked automatically. Input range Default value 1 - 999 days 999 days Next GLP test If you define a date for the next GLP test, then the GLP test interval will be tracked automatically. Format: YYYY:MM:DD Action Selection of the action which is carried out when the time interval has expired.
■■■■■■■■■■■■■■■■■■■■■■ 8.7 8 Titrants Titer determination options and data Titrant: Edit ▶ Titer options Detailed information concerning the titer determination is displayed in the dialog Edit titrant / Titer options: ■ ■ ■ Titer method Method by which the titer was determined. If the titer was entered manually, then manual will be displayed. User User who carried out the titer determination.
■■■■■■■■■■■■■■■■■■■■■■ 8.7 Titer determination options and data Titer validity If you define a time interval for the validity of the titer, then the date in Next titer determ. will be tracked automatically. Input range Default value 1 - 999 days 999 days Next titer determ. If you define a date for the next titer determination, then the time interval for the Titer validity will be tracked automatically.
■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants NOTE If multiple determinations have been carried out in order to determine the titer, then only one entry will be made in the history. [Delete History] Delete the entire history. [Graph] Open the diagram of the titer values, see following chapter. Dialog "History / Graph" Titrant: Edit ▶ Titer options ▶ History ▶ Graph In this diagram, titer values are plotted against the date of the titer determination.
■■■■■■■■■■■■■■■■■■■■■■ 9.1 Editing reagent data 9 Reagents Main dialog: System ▶ Reagents This chapter describes how you can create a list of the reagents used in the system. The list of reagents gives the designation and type of each reagent. [New] Add a new reagent to the list, see following chapter. [Delete] Delete the selected reagent from the list. [Edit] Edit the data of the selected reagent, see following chapter. 9.
■■■■■■■■■■■■■■■■■■■■■■ 9 Reagents Reagent The designation of the reagent is used for unambiguous identification. Entry 24 characters maximum Entry 24 characters maximum Comment [Reagent monitoring] Set the parameters for the reagent monitoring, see following chapter. 9.2 Reagent monitoring The conditions for the monitoring of the reagent are defined in the dialog Edit reagent / Reagent monitoring. If one of the following values is reached, then the reagent must be replaced.
■■■■■■■■■■■■■■■■■■■■■■ 9.2 Reagent monitoring ■ at the end of the determination. Number of determ. The number of determinations to be carried out with a certain amount of reagent depends on the type of sample and its amount. Input range Selection Default value 1 - 999 off off Working life Working life of the reagent. Input range Selection Default value 1 - 999 days off off Volume Volume of titrant dosed. Input range Selection Default value 1.0 - 999.
■■■■■■■■■■■■■■■■■■■■■■ 9 Reagents auto If a monitored parameter has reached the limit set, the method defined below is started automatically. Memory This parameter can only be edited with Reagent replacement = auto. Memory location the method is loaded from. All memory locations are selectable, even if they are currently not accessible. Selection Default value Internal memory | External memory 1 | External memory 2 | Shared memory Internal memory Shared memory Shared directory in the network.
■■■■■■■■■■■■■■■■■■■■■■ 10 Sensors Main dialog: System ▶ Sensors This chapter describes how you can create a list of the sensors used in the system. Two standard sensors are defined in the sensor list: Metal electrode and Temperature sensor. These sensors cannot be deleted or renamed. A maximum of 25 additional sensors can be added to these sensors. For each sensor, the following data is displayed in the sensor list: ■ ■ Designation Type – Metal: Metal electrode – Temp.
■■■■■■■■■■■■■■■■■■■■■■ 10.1 10 Sensors Adding a new sensor Before you can use a sensor, you must add it to the sensor list. To do this, use the button [New]. ■ 10.2 The properties dialog is opened after the sensor type has been selected, see following chapter.
■■■■■■■■■■■■■■■■■■■■■■ 10.3 Monitoring the working life Entry 8 digits maximum Entry 24 characters maximum Comment [Working life] Define the working life of the sensor (see Chapter 10.3, page 72). 10.3 Monitoring the working life Sensor: Edit ▶ Working life In the dialog Edit sensor / Working life, you can define the time interval after which the sensor must be replaced. If you do not wish to monitor the working life, then you can enter only the date of manufacture for documentation purposes.
■■■■■■■■■■■■■■■■■■■■■■ 10 Sensors Selection Default value Display message | Document message | Cancel determination Display message For all three options it is documented in the determination data (see dialog More determination data / Messages), that the time interval has been expired. Display message A message is displayed. You can select whether you want to continue with the determination or cancel the run.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager Main dialog: System ▶ Device manager This chapter describes how you can configure the Ti-Touch, the USB Sample Processor and the connected peripheral devices. A detailed description of the hardware installation of the USB Sample Processor can be found in the corresponding installation manual. The instrument list gives the name and type of every instrument.
■■■■■■■■■■■■■■■■■■■■■■ 11.1 11 Device manager Adding a new device Of the following device types, you can enter one device each in the device list, even if it is not yet connected: ■ ■ ■ ■ ■ USB Sample Processor Balance Barcode reader USB/RS-232 adapter PC keyboard Proceed as follows to do this: 1 Displaying the device selection Tap on [New]. 2 Selecting the instrument Tap on the button for the desired device. The new device is entered in the list. 11.
■■■■■■■■■■■■■■■■■■■■■■ 11.3 Ti-Touch 11.3 Ti-Touch Instrument list: 915 KF Ti-Touch ▶ Edit Device name This designation is used for identification purposes when selecting control devices (command, manual control). Entry 24 characters maximum Entry 24 characters maximum Comment Switch off display If this time interval has expired without the Ti-Touch having been operated, then the display will be switched off. Touching the display will switch it back on again at any time.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager The "Control Remote Box" is the interface via which the system can be started and stopped externally. If multiple Remote Boxes are connected, then the one that is recognized first when the program starts will be used as the "Control Remote Box." Selection 11.3.1 Name of the control device / Number of the MSB connector | Not available E-mail 915 KF Ti-Touch: Edit ▶ E-mail The system allows you to send displayed messages as e-mails.
■■■■■■■■■■■■■■■■■■■■■■ 11.3 Ti-Touch Send only during running determination on | off (Default value: on) If this parameter is activated, then messages will be sent as e-mails only if a determination is running. Deactivate this parameter if messages are also to be sent in normal status. Mail server Address of the mail server for outgoing mail, e.g. mail.metrohm.ch. You can find the address of the mail server either in your e-mail program settings or obtain it from your IT department.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager created the first time a PC/LIMS report is generated. All three memory locations are listed as possible selections, even if they cannot be accessed at the moment. Selection Default value off | External memory 1 | External memory 2 | Shared memory off off The report will not be saved as a TXT file. Shared memory The report will be saved in a shared directory on the network. The shared directory is selected in the Edit device / Shared memory dialog (see Chapter 11.
■■■■■■■■■■■■■■■■■■■■■■ 11.3 Ti-Touch CAUTION If the computer on which you share a memory location does not have the same subnet mask as the Ti-Touch, then a WINS server must be present. The computer must be entered in this server. Computer Host name of the computer on which a memory location is to be shared. If you are working with Windows, then you will find the host name of the computer as follows: In the Windows command prompt window, enter the command ipconfig -all.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager Entry Default value max. 32 characters empty [Connect] Establish the network connection. If the connection has been set up correctly, then all of the input fields will become inactive and the label switches to [Disconnect]. The network connection can be disconnected with this. 11.3.4 TCP/IP settings 915 KF Ti-Touch: Edit ▶ TCP/IP settings If you have your Ti-Touch connected to your network, then you have to define the network-relevant settings in this dialog.
■■■■■■■■■■■■■■■■■■■■■■ 11.4 Metrohm control devices Entry Default value x.x.x.x empty MAC address Unambiguous address of the network adaptor for identification within a network. This parameter cannot be edited. NetBIOS name Unambiguous designation of the device within the network. This parameter cannot be edited. 11.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager ■ 11.4.1 Properties of the peripheral devices on the MSB connector (see Chapter 11.4.3, page 85) Properties – Measuring input List of devices: Control device ▶ Edit ▶ Measuring input 1 ▶ Properties In this dialog, you define the type of temperature sensor you have connected to the selected measuring input. ADC type Type of analog-digital converter. Serial number Serial number of the measuring interface. Thick film type Type of the thick film.
■■■■■■■■■■■■■■■■■■■■■■ 11.4 Metrohm control devices R (25 °C) This parameter can only be edited with Type = NTC. Nominal resistance of the NTC sensor at 25 °C. Input range Default value 1000 - 99999 ohm 30000 ohm The default value applies to Metrohm sensors with an NTC sensor. B value This parameter can only be edited with Type = NTC. Material constant of the NTC sensor. B values of NTC sensors are frequently based on different reference temperatures (usually 25 °C and 50…100 °C).
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager Switch on on | off (Default value: on) If this parameter is activated, then you will be requested to prepare the dosing device when the Ti-Touch is switched on. Attach an exchange or dosing unit on | off (Default value: on) If this parameter is activated, then you will be requested to prepare the dosing device when the exchange/dosing unit is attached.
■■■■■■■■■■■■■■■■■■■■■■ 11.5 USB Sample Processor The dialog shows the USB Sample Processor with its tower (or towers), the MSB connectors and connected peripheral devices (dosing unit, stirrer, Remote Box). If the control device is not connected, then the properties dialog of the control device will be displayed directly (see Chapter 11.5.1, page 86). The settings of the MSB connectors and the connected peripheral devices are described in Chapter "Properties – MSB connector", page 84.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager Comment Entry 24 characters maximum Program version Program version of the instrument software. Serial number Shows the serial number of the device. Rack name Name of the currently attached sample rack. [Adjustment data] Display the internal adjustment data (EEPROM data) of the USB Sample Processor. CAUTION Do not modify and save these settings under any circumstances. This data is required by the service technician in case of positioning problems. 11.5.
■■■■■■■■■■■■■■■■■■■■■■ 11.5 USB Sample Processor NOTE In the case of a USB Sample Processor with 2 towers, Tower 1 is the tower on the right and Tower 2 is the tower on the left, when viewed from the front. Maximum stroke path Lowest permissible lift position for automatic and manual operation. This is a safety setting. A correct specification of this value can prevent the glass of an electrode from breaking, because a titration head can not be moved below the position specified.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager for 814 USB Sample Processor XL: Input range 100.0 - 300.0 mm Default value 166.0 mm for 815 Robotic USB Sample Processor XL: Input range 100.0 - 300.0 mm Default value 196.0 mm 11.5.3 Properties – Swing Head List of devices: Sample Processor ▶ Edit ▶ Swing Head ▶ Properties The dialog Edit device / Properties shows a list of all external positions with the assigned swing angle and the specific work position for each. [Ext. pos.
■■■■■■■■■■■■■■■■■■■■■■ 11.5 USB Sample Processor Properties which apply to all four external positions can be defined in the dialog Properties / Edit external positions 1-4. Shift position Lift position at which the robotic arm rotates to the external positions. Input range Default value 0 - 'Maximum stroke path' mm 0 mm A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager Properties – Robotic arm Swing Head: Properties ▶ Robotic arm Properties of the robotic arm can be defined in the dialog Properties / Robotic arm. Swing offset The swing offset is a physical angle offset of a specific robotic arm model. The required values can be found in the leaflet for the robotic arm. Input range Default value –270.00 - 270.00 ° 0.00 ° Maximum swing angle Useable swing range of the robotic arm.
■■■■■■■■■■■■■■■■■■■■■■ 11.5 USB Sample Processor Rotation offset The rotation offset is the offset from the center of the tower to the center of the robotic arm. This value does not usually need to be changed. If a Swing Head is to be mounted on the tower with a lateral offset, then this value can be determined by the service technician when the rack is adjusted. Input range Default value –270.00 - 270.00 ° 0.00 ° Swing direction The swing direction of the robotic arm depends on its type.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager Input range Default value (Offset) - (Offset + max. swing range) ° 60.00 ° The offset is made up of a design-dependent angle (approx. 8…9°) together with the robotic arm offset from the robotic arm properties. The maximum swing range is also defined under the robotic arm properties (see "Properties – Robotic arm", page 91). Work position Work position for the selected external position. Input range 11.
■■■■■■■■■■■■■■■■■■■■■■ 11.6 Sample racks ■ ■ Positions Number of positions on the rack. Code The rack code corresponds to the alignment of magnets on the underside of the rack and is read out by the Sample Processor for rack recognition. [Load] Load a new sample rack (see "Loading the sample rack", page 94). [Copy] Create a new sample rack by copying an existing rack (see "Creating a new sample rack", page 95). [Delete] Delete the selected sample rack from the list.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager ■ Tap on [Show files]. The list with the saved sample rack files is opened. 3 Load the sample rack file ■ ■ Select the desired file. Tap on [Load]. The new sample rack is now loaded and appears in the list of available racks. Creating a new sample rack Own sample racks can be simply and conveniently created by copying an existing sample rack.
■■■■■■■■■■■■■■■■■■■■■■ 11.6 Sample racks You can edit the data of the selected rack in the dialog Sample rack / Edit rack data. Beaker radius samples Actual radius of the sample vessels at the general sample positions of the rack. This beaker radius may not be less than the minimum beaker radius defined in the tower properties (see "Min. beaker radius," page 88). If the lift is to be moved to the work position, then these two values will be compared with one another.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager [Special beakers] Define rack positions as special beakers (see "Editing special beakers", page 98). Lift positions Tower 1/2 Sample rack list: Rack ▶ Edit ▶ Lift pos. tower 1 / Lift pos. tower 2 You can define rack-specific lift positions in the dialog Edit rack data / Lift positions tower 1/2. These then apply for all rack positions except those that have been defined as special beakers.
■■■■■■■■■■■■■■■■■■■■■■ 11.6 Sample racks Input range Default value 0 - 'Maximum stroke path' mm 0 mm A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position. The maximum stroke path is defined in the properties of the tower (see "Maximum stroke path", page 88). Shift position The lift is raised to this position with each rotational movement of the rack whenever it is located at a lower lift position.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager A maximum of 16 rack positions can be defined as special beakers for each sample rack. The dialog Edit rack data / Special beakers shows a list of all special beakers with their assigned rack positions. [Edit] Edit the data of the selected special beaker, see the following. You can edit the data of the selected special beaker in the dialog Special beaker / Edit. Rack position Number of rack position for selected special beaker.
■■■■■■■■■■■■■■■■■■■■■■ 11.6 Sample racks lift is to be moved to the work position, then these two values will be compared with one another. Input range Selection Default value 1.0 - 100.0 mm off off off No check takes place. Beaker sensor Each time this special beaker is to be moved to with the MOVE command, the beaker sensor checks whether a vessel is present. In the MOVE command, you define the action that takes place if the beaker sensor does not detect a vessel at the position being moved to.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager 2 Carry out the adjustment ■ ■ ■ ■ Tap on the button [Start] (NOT on the fixed key [ ]). The rack is being initialized. The rack then moves to position 1 and the lift is lowered to the work position. If necessary, the lift position can be corrected with the arrow keys [⇧] and [⇩]. Now use the arrow keys [⇦] and [⇨] to rotate the rack so that the robotic arm or the center of the titration head is positioned exactly above the center of rack position 1.
■■■■■■■■■■■■■■■■■■■■■■ 11.7 Printer The list of devices always includes a printer, even if no corresponding device is connected. In this dialog, you also define when you would like to have a report generated as a PDF file. Device name Here you can enter a device name of your choice. Entry 24 characters maximum Entry 24 characters maximum Comment Printer Selection of the printer type.
■■■■■■■■■■■■■■■■■■■■■■ 11.7.1 11 Device manager PDF settings Printer: Edit ▶ PDF settings The settings for saving a report as a PDF file are defined in this dialog. Memory Memory location where the PDF file will be saved. The report will be saved in the directory PDF_Report. This directory is created when a PDF file is generated. The file name is appended with the date and time when the determination was started and the date and time when the PDF was generated (e.g.
■■■■■■■■■■■■■■■■■■■■■■ 11.7 Printer Add or change comments allowed on | off (Default value: on) If this option is activated, then comments can be added to the PDF file. 11.7.2 Network printer Printer: Edit ▶ Network printer If you have connected your Ti-Touch to your network, then you can specify a network printer for your reports in this dialog. NOTE If you have shared a memory location on a PC within your network and if you have configured the Ti-Touch accordingly (see Chapter 11.3.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager Domain Network domain in which the above-specified computer is located. If you are working with Windows, then you will find the domain name as follows: In the Windows command window, enter the command ipconfig -all. The domain is listed together with other parameters of the computer. Leave the field empty if the computer is not located within a domain. Entry Default value max.
■■■■■■■■■■■■■■■■■■■■■■ 11.8 Balance 11.8 Balance List of devices: Balance ▶ Edit You define the type of balance and its connector in the dialog Device manager / Edit. Device name Here you can enter a device name of your choice. Entry 24 characters maximum Entry 24 characters maximum Comment Balance type If you have connected a balance, then you have to specify the balance type here.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager The following table indicates the balance type that needs to be selected for the balance model: 11.
■■■■■■■■■■■■■■■■■■■■■■ 11.9 USB/RS-232 adapter Baud rate Transfer rate in characters per second. Selection Default value 1200 | 2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200 9600 Data bits Number of data bits. Selection Default value 7|8 8 Parity Type of parity testing. Selection Default value even | odd | none none Stop bits Number of stop bits. Selection Default value 1|2 1 Handshake Type of data transfer protocol.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager Selection Default value none | Software (XON/XOFF) | Hardware (DTR/ CTS) Hardware (DTR/CTS) Software (XON/XOFF) Use the software handshake when you send a PC/LIMS report via an RS-232 interface. 11.10 PC keyboard List of devices: PC keyboard ▶ Edit A commercially available USB keyboard can be connected to make it easier to enter text and numbers. It will be recognized automatically and entered in the list of devices with default settings.
■■■■■■■■■■■■■■■■■■■■■■ 11.11 Barcode reader Table 3 Keyboard assignment Function in the editing dialog PC keyboard key [Cancel] [Esc] Enter respective character Numbers, letters and special characters Enter respective character Numbers, letters and special characters + Shift [OK] [⇄] (Tabulator) or [Enter] on the numerical keypad [⌫] [←] (backspace) [Delete entry] [Delete] Line break (for multi-line text input) [↲] (enter key) The cursor is moved up or down by one character.
■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager A barcode reader can be connected to read in sample data or other texts. A connected barcode reader will be recognized automatically and entered in the list of devices with default settings. You will hear an acoustic signal as confirmation that a character string has been transmitted by the barcode reader and accepted. NOTE The data will be read only if the system is in normal status, i.e. when no determination is running.
■■■■■■■■■■■■■■■■■■■■■■ 11.11 Barcode reader Identification 1 The character string is entered in the input field Identification 1. Identification 2 The character string is entered in the input field Identification 2. Sample size The character string is entered in the input field Sample size. Character strings containing characters other than numbers and decimal separators will be ignored. Keyboard layout Specify the country-specific keyboard layout for the emulation of the PC keyboard.
■■■■■■■■■■■■■■■■■■■■■■ 12 File manager 12 File manager Main dialog: System ▶ File manager The saved methods, determinations, sample tables, etc. are managed in the file manager. You can also create a backup of your system (all data and settings). Similarly, an existing backup can be reloaded. The file memory is organized as follows: ■ ■ 12.
■■■■■■■■■■■■■■■■■■■■■■ 12.1 Managing files NOTE If you use an external storage medium with the FAT or FAT32 file system, then you can save a maximum of 999 files per group. If you find it necessary to store more than 999 files in a single group, then you must reformat the storage medium with the file system ExFAT (see Chapter 31.7.2, page 347). [Show all] Display the list of all the files in the selected memory location. [Show files] Display the list of all the files in the selected group.
■■■■■■■■■■■■■■■■■■■■■■ 12 File manager [Copy] Copy the selected file to a different memory location (see Chapter 12.1.1, page 115). [Delete] Delete the selected file. [Load] Load the selected file. 12.1.1 Copying a file Proceed as follows to copy a file: 1 Select the file ■ ■ Select the desired file. Tap on [Copy]. The selection of memory locations is displayed. 2 Copy the file ■ Tap on the desired memory location. Only memory locations currently being accessible are active. The file is copied.
■■■■■■■■■■■■■■■■■■■■■■ 12.1 Managing files 12.1.3 File properties File manager: Internal memory / External memory 1 / External memory 2 ▶ Show all / Show files ▶ Properties Detailed information concerning the file is displayed in the dialog Show files / Properties. They can be edited to a certain extent. ■ ■ ■ ■ ■ ■ ■ ■ ■ 116 ■■■■■■■■ File name Name of the file. File group Entry or selection of the group in which the file will be saved.
■■■■■■■■■■■■■■■■■■■■■■ 12.2 12 File manager External storage medium [Backup] Create a backup of all data and settings on this storage medium (see Chapter 12.3, page 119). [Restore] Load the backup. This function is active only if a backup is available (see Chapter 12.3, page 119). [Memory info] Display information on the storage medium, see following chapter.
■■■■■■■■■■■■■■■■■■■■■■ 12.2 External storage medium Directory structure The directory structure appears as follows on the external storage medium: 118 ■■■■■■■■ Backup All of the files of the backup are stored in this directory. The directory will be created the first time a backup is created. Files This directory contains all of the groups as well as the files contained in them (methods, determinations, etc.). pc_lims_report PC/LIMS reports are stored in this directory as TXT files.
■■■■■■■■■■■■■■■■■■■■■■ 12.3 12 File manager Creating backups / Restoring data File manager: External memory 1 / External memory 2 You can use the [Backup] function to easily create a backup containing all the data and settings of your system. You should make a backup at regular intervals in order to avoid data loss. NOTE Only one backup can be created on a storage medium. If a backup has already been stored on the medium, then it will be overwritten when this function is carried out again. 12.3.
12.3 Creating backups / Restoring data ■■■■■■■■■■■■■■■■■■■■■■ Restorable data blocks The following data blocks can be loaded individually: ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ 120 ■■■■■■■■ Methods All of the methods stored in the internal memory. Sample table Current sample table. Result table Current result table. Current determination data All the data for the current determination (including the method with which the determination was carried out). Control Settings in the control dialog.
■■■■■■■■■■■■■■■■■■■■■■ 13 GLP manager 13 GLP manager Main dialog: System ▶ GLP manager In the GLP manager you can document data for various GLP tests. The results of the automatic system tests carried out after switching on are also documented. The following tests can be documented: ■ ■ ■ GLP test "Measurement" (see Chapter 13.3, page 123) GLP test "Titration" (see Chapter 13.3, page 123) System validation (see Chapter 13.
■■■■■■■■■■■■■■■■■■■■■■ 13.1 Automatic system test [GLP tests HW/SW] Document the GLP tests "Measurement" and "Titration" (see Chapter 13.3, page 123). [System validation] Document the system validation (see Chapter 13.4, page 127). [Monitoring] Define time intervals for system maintenance and backups (see Chapter 13.5, page 131). [View test data] Display the data of the selected test. 13.1 Automatic system test The system test is carried out automatically when the Ti-Touch is switched on.
■■■■■■■■■■■■■■■■■■■■■■ 13 GLP manager [Delete] Delete the selected test tool from the list. [Edit] Change the designation of the selected test tool. 13.3 GLP tests for measurement and titration Main dialog: System ▶ GLP manager ▶ GLP tests HW/SW ▶ GLP test "Measurement" / GLP test "Titration" You can document the GLP tests for measurements and titrations in the dialog GLP manager / GLP tests Hardware/Software and its subdialogs.
■■■■■■■■■■■■■■■■■■■■■■ 13.3 GLP tests for measurement and titration ■ ■ Select the hardware used for the test (see "Dialogs "GLP test "Measurement" / Hardware" and "GLP test "Titration" / Hardware"", page 126). Tap on the fixed key [ ]. The dialog GLP tests HW/SW / GLP test "Measurement" appears again. 4 Define the test interval ■ ■ ■ 13.3.1 Tap on [GLP test interval].
■■■■■■■■■■■■■■■■■■■■■■ 13 GLP manager [GLP test interval] Define the time interval for the GLP test (see "Dialogs "GLP test "Measurement" / Test interval" and "GLP test "Titration" / Test interval"", page 125). [Hardware] Document the hardware with which the GLP test has been carried out (see "Dialogs "GLP test "Measurement" / Hardware" and "GLP test "Titration" / Hardware"", page 126).
■■■■■■■■■■■■■■■■■■■■■■ 13.3 GLP tests for measurement and titration Display message A message is displayed. You can select whether you want to continue with the determination or cancel the run. Document message In the determination data it will be documented, that the time interval has been expired. Cancel determination The determination is stopped. Dialogs "GLP test "Measurement" / Hardware" and "GLP test "Titration" / Hardware" Test tool Selection of the test tool.
■■■■■■■■■■■■■■■■■■■■■■ 13.4 13 GLP manager System validation Main dialog: System ▶ GLP manager ▶ System validation In the dialog GLP manager / System validation and its subdialogs, you can document the system validation results and define the time interval after which validation must be carried out again. Proceed as follows: 1 Open the properties dialog ■ In the dialog System / GLP manager, tap on the button [System validation]. 2 Edit data ■ Define the method, the result, etc.
■■■■■■■■■■■■■■■■■■■■■■ 13.4 System validation ■ Tap on the fixed key [ ]. The dialog GLP manager / System validation appears again. 5 Insert a note ■ ■ ■ 13.4.1 Tap on [Note (SOP)]. Enter a brief description, e.g. a summary of the SOP (standard operating procedure) according to which the system validation was carried out (see "Dialog "System validation / Note (SOP)"", page 129). Tap on the fixed key [ ].
■■■■■■■■■■■■■■■■■■■■■■ 13 GLP manager [Validation interval] Define the time interval for the system validation (see "Dialog "System validation / Validation interval"", page 129). [Test statistics] Document the statistical data of the system validation (see "Dialog "System validation / Test statistics"", page 130). Dialog "System validation / Note (SOP)" In this dialog, you can enter a brief text, e.g.
■■■■■■■■■■■■■■■■■■■■■■ 13.4 System validation Selection Default value Display message | Document message | Cancel determination Display message For all three options it is documented in the determination data (see dialog More determination data / Messages), that the time interval has been expired. Display message A message is displayed. You can select whether you want to continue with the determination or cancel the run.
■■■■■■■■■■■■■■■■■■■■■■ 13 GLP manager Input range Default value 0.00000 - 100.00000 % empty % a sys Systematic error. Input range Default value 13.5 System monitoring 13.5.1 Service interval 0.00000 - 9999999999.00000 empty Main dialog: System ▶ GLP manager ▶ Monitoring ▶ Service interval In the dialog GLP manager / Service interval, you can define the time interval for the maintenance of the system by the Metrohm Service department. The service interval is checked each time the system is started.
■■■■■■■■■■■■■■■■■■■■■■ 13.5 System monitoring 13.5.2 Backup interval Main dialog: System ▶ GLP manager ▶ Monitoring ▶ Backup interval In the dialog GLP manager / Backup interval, you can define the time interval for backups. The backup interval is checked each time the system is started. Last backup Date on which the last backup was created.
■■■■■■■■■■■■■■■■■■■■■■ 14 Common variables 14 Common variables Main dialog: System ▶ Common variables You can save 25 method-independent variables, or common variables. These variables can be used in future calculations (as variables CV01…CV25). Common variables are useful, e.g. for the following applications: ■ ■ Determination of a blank value which will be taken into account during the content determination of the sample.
■■■■■■■■■■■■■■■■■■■■■■ 14.1 Editing common variables 14.1 Editing common variables List of common variables: Common Variable ▶ Edit The common variables can be modified as follows: ■ ■ Edit manually, see the following. Automatic assignment from the determination run. A calculation result must be configured accordingly for this purpose (see Chapter 14.4, page 137). Name Designation of the common variable. Entry Default value 24 characters maximum empty Value Value of the common variable.
■■■■■■■■■■■■■■■■■■■■■■ 14 Common variables Next assignment This parameter is displayed only if a validity has been defined for the common variable. Date on which the validity of the common variable expires. Format: YYYY:MM:DD [Validity] Define the time interval for the validity of the common variable (see Chapter 14.3, page 136). [Properties] Display the properties of the common variable, see following chapter. 14.
■■■■■■■■■■■■■■■■■■■■■■ 14.3 Monitoring validity 14.3 Monitoring validity Common variable: Edit ▶ Validity In the dialog Edit common variables / Validity, you can define the time interval after which a new value must be assigned to the common variable. Last assignment Date on which the common variable was last assigned a value.
■■■■■■■■■■■■■■■■■■■■■■ 14 Common variables Document message In the determination data it will be documented, that the time interval has been expired. Cancel determination The determination is stopped. 14.4 Assigning a result automatically to a common variable NOTE This instruction is based on the assumption that the method contains a calculation command with a calculation.
■■■■■■■■■■■■■■■■■■■■■■ 14.4 Assigning a result automatically to a common variable 3 Save the settings Tap on the fixed keys [ ] or [ ]. In the future, the calculated result will be assigned to the selected common variable (result name, value and unit).
■■■■■■■■■■■■■■■■■■■■■■ 15 Templates 15 Templates Main dialog: System ▶ Templates You have the option of defining system-specific templates. You can use these templates when editing the respective data. You can create the following templates: ■ ■ ■ ■ ■ 15.1 Sample data (see Chapter 15.1, page 139) Create sample identifications and sample assignments. Custom result templates (see Chapter 15.2, page 143) Define formulas for result calculations. Input lines (see Chapter 15.
■■■■■■■■■■■■■■■■■■■■■■ 15.1 Sample data Sample assignments (see Chapter 15.1.2, page 141) You can assign a particular method to a sample identification. ■ Use sample identification list on | off (Default value: off) Activate this parameter so that the sample identifications defined in the list will be displayed as selections. Use sample assignment table on | off (Default value: off) Activating/deactivating the use of the sample assignment table.
■■■■■■■■■■■■■■■■■■■■■■ 15 Templates You can create a system-specific list with sample identifications in the dialog Sample data / Sample identification list. This list makes it easier for you to enter frequently used sample identifications, i.e. this list is available to you for entering them (main dialog, command REQUEST, etc.). It may be a good idea to define the unchanging part of the identification as a template and to add the variable part during sample data input.
■■■■■■■■■■■■■■■■■■■■■■ 15.1 Sample data In the sample assignment table, a particular method is assigned to a sample identification. In this way, you ensure that your samples will be processed with the correct method; mix-ups are not longer possible. When you start a determination, you need only enter the sample identification; the method is loaded automatically. [New] Add a new sample assignment to the list, see following chapter. [Delete] Delete the selected sample assignment.
■■■■■■■■■■■■■■■■■■■■■■ 15 Templates 32 characters maximum Selection of methods stored in the selected memory Entry Selection 15.2 Custom result templates Main dialog: System ▶ Templates ▶ Custom result templates In the dialog Templates / Custom result templates, you can define the result calculations which can be loaded in the calculation command CALC. With the exception of the following points, creating a result template is identical to editing a calculation in the command CALC (see Chapter 28.4.1.
■■■■■■■■■■■■■■■■■■■■■■ 15.2 Custom result templates 15.2.1 Editing result templates List of result templates: Result template ▶ New / Edit Result name The result name is the text which will be shown in the results display and in the report. Entry Default value max. 24 characters R Calculation formula Display of the calculation formula. A special editor is opened for the definition (see Chapter 28.4.3, page 293). Entry Default value max.
■■■■■■■■■■■■■■■■■■■■■■ 15 Templates Variable for mean value If the statistic calculations have been activated (see method options), the mean value of the single results will be saved as variable SMN1 to SMN9. As default value, always the first free variable is being displayed. SMN1 … SMN9 | off Selection off For the result not statistic calculations are carried out. Save as titer The result can be saved as titer for the selected titrant.
■■■■■■■■■■■■■■■■■■■■■■ 15.2 Custom result templates Display result on | off (Default value: on) If you deactivate this parameter, the result is neither displayed in the result dialog nor printed in the result report. This can be advisable for intermediate results. Save result in result table on | off (Default value: off) The calculated result can be saved in the result table. This may be advisable if e.g. the results of all determinations carried out on a particular day are to be displayed clearly.
■■■■■■■■■■■■■■■■■■■■■■ 15.3 15 Templates Input lines Main dialog: System ▶ Templates ▶ Input lines In the dialog Templates / Input lines, you can define the input signals at the remote interface as a template. You can select these templates in the command SCAN. The list can contain a maximum of 20 templates. Table 4 List of predefined input signals Signal name Input signal Function Cond OK ******1* queries "Cond OK" condition of device.
■■■■■■■■■■■■■■■■■■■■■■ 15.3 Input lines Signal name Input signal Function Sample ready ***1**** waits for a stepping pulse, of, for instance, a connected Sample Processor, as soon as it is ready. [New] Add a new input signal to the list, see following chapter. [Delete] Delete the selected input signal from the list. [Edit] Edit the data of the selected input signal, see following chapter. 15.3.
■■■■■■■■■■■■■■■■■■■■■■ 15.4 15 Templates Output lines Main dialog: System ▶ Templates ▶ Output lines In the dialog Templates / Output lines, you can define the output signals at the remote interface as a template. You can select these templates in the command CTRL. The list can contain a maximum of 20 templates. Table 5 List of predefined output signals Signal name Output signal Function Continue *********p**** sends a stepping pulse to the connected Sample Processor.
■■■■■■■■■■■■■■■■■■■■■■ 15.4 Output lines Signal name Output signal Function Meter Mode T *********0010* switches the 780/781 pH/Ion meter to temperature measurement and starts the measurement. Meter Mode U *********0011* switches the 780/781 pH/Ion meter to voltage measurement and starts the measurement. Start device* ********p****p starts Devices 1 and 2 (e.g. Titrino, Titrando, etc.) *). The signal is transmitted as a short pulse of 200 ms. Start device1 *************p starts Device 1 (e.
■■■■■■■■■■■■■■■■■■■■■■ 15.4.1 15 Templates Editing the output signal List of input signals: Output signal ▶ New / Edit Signal name Designation of the template. 24 characters maximum Entry Output signal Entering the desired bit pattern: Entering the bit pattern: ■ ■ ■ ■ 0 = line inactive 1 = line active * = retain line status p = set pulse The output lines are always numbered from right to left, i.e. with the signal *************1 line 0 is set.
■■■■■■■■■■■■■■■■■■■■■■ 15.5 Report header 15.5 Report header Main dialog: System ▶ Templates ▶ Report header In the dialog Templates / Report header, you can create a system-specific report header that contains information concerning the laboratory, for example. The report header can contain a maximum of four lines of 46 characters each and will always be printed out in front of the standard report header.
■■■■■■■■■■■■■■■■■■■■■■ 15 Templates 2 Copying the file to an external storage medium ■ ■ Copy the file "CustomImage.jpg" to the directory "915". If this structure is not maintained, the file will not be found because the software directly accesses the "915" directory. Plug in the external storage medium at the Ti-Touch. 3 Loading the graphics file ■ ■ In the Templates / Report header dialog, tap on the [Load] button. The file is copied to the internal memory.
■■■■■■■■■■■■■■■■■■■■■■ 16.1 Creating a new method 16 Methods 16.1 Creating a new method Proceed as follows to create a new method: 1 Open the method table ■ ■ In the main dialog, tap on [Load method]. Tap on [New method]. The method table with the stored templates opens: 2 Load the method ■ Tap on [Empty method]. or ■ Select the desired template and tap on [Load template]. The method is now loaded. New method is displayed in the main dialog in the title line.
■■■■■■■■■■■■■■■■■■■■■■ 16.2 16 Methods Saving a method If you modify method parameters, then you can save these as your own method. To save a method, proceed as follows: 1 Open the command sequence ■ In the main dialog, tap on [Edit parameters]. The command sequence of the method is displayed: 2 Modify/apply the method name ■ Tap on [Save method]. A method name will be suggested for new methods. If the method has already been saved once, then the method name will be displayed.
■■■■■■■■■■■■■■■■■■■■■■ 16.3 Loading a method ■ Tap on [Save]. The method will be saved and the command sequence is displayed. Enter a new name: ■ ■ ■ Tap on the input field File name. The text editor opens. Enter a new file name (max. 32 characters) and confirm with [OK]. Tap on [Save]. The method will be saved and the command sequence is displayed. 16.3 Loading a method To load a method, proceed as follows: 1 Open the method table ■ ■ ■ ■ In the main dialog, tap on [Load method].
■■■■■■■■■■■■■■■■■■■■■■ 16 Methods 3 Load the method ■ Tap on [Load]. The method is now loaded. The method name is displayed in the main dialog in the title line. NOTE The data of the current determination will be deleted when you load a method. 16.4 Editing a method Main dialog: Edit parameters In the dialog Parameters / Sequence, you can define and edit the method commands which are executed successively during a determination.
■■■■■■■■■■■■■■■■■■■■■■ 16.4 Editing a method [Insert command] Insert a new method command. It is inserted before the selected command. NOTE Not all commands are available for subsequences. Commands which cannot be inserted into subsequences are disabled. [Delete command] Delete the selected method command. [Edit command] Edit the selected method command. 16.4.1 Inserting a command A method is comprised of individual commands.
■■■■■■■■■■■■■■■■■■■■■■ 16 Methods This dialog contains all of the commands which can be inserted into a method run, organized into thematic groups. 2 Select a command group ■ Tap on the desired command group. In the case of [Report], the command is inserted directly into the command list. For all other command groups, the available commands will be displayed (titration / measuring commands, etc.). 3 Insert the command Tap on the desired command.
■■■■■■■■■■■■■■■■■■■■■■ 16.5 Method options Statistics on | off (Default value: off) If this parameter is activated, then statistics calculations will be carried out for all of the defined results. The prerequisite here, however, is that a variable for the mean value is defined in the result options (see "Dialog "Edit calculation / Result options"", page 289). Number of determinations The number of determinations that are carried out for the statistics calculations.
■■■■■■■■■■■■■■■■■■■■■■ 16 Methods Method check at start on | off (Default value: on) If this option is activated, then the following tests are carried out at the method start. ■ ■ ■ ■ Are the required control and peripheral devices connected? Are the titrant and the sensor on hand? Have the monitoring intervals expired? etc. NOTE We recommend that the method check only be deactivated if the titration must begin immediately after the start of the determination (e.g.
■■■■■■■■■■■■■■■■■■■■■■ 16.5 Method options Selection Default value no | Selection of configured sample racks no Increase sample variable automatically on | off (Default value: on) The sample variable describes the current position of the sample on the rack of the Sample Processor. If this option is activated, then the sample variable is automatically increased by 1 at the end of each determination.
■■■■■■■■■■■■■■■■■■■■■■ 16 Methods NOTE A line set active is not being reset automatically, not even at the end of the determination. Entering the bit pattern: ■ ■ ■ ■ 0 = line inactive 1 = line active * = retain line status p = set pulse The output lines are always numbered from right to left, i.e. with the signal *************1 line 0 is set. With a pulse, the length is set to 200 ms. If you wish to set pulses with other lengths, you have to define a corresponding template.
■■■■■■■■■■■■■■■■■■■■■■ 16.5 Method options 16 characters maximum Entry Fixed sample size on | off (Default value: off) Activate this parameter if the same sample amount is always to be used for all determinations. If this is case, then you can define the sample size here. It will be displayed in the main dialog, but it can no longer be changed there. Sample size This parameter can only be edited when Fixed sample size is activated. Input range Default value –999999999 - 9999999999 1.
■■■■■■■■■■■■■■■■■■■■■■ 16 Methods Modifying sample data This instruction is for the purpose of clarifying the settings which are possible in this dialog. The following settings should be made: ■ ■ ■ Change the designation for the input field of the first sample identification. Hide the input field for the second sample identification. Define a fixed sample size. Proceed as follows: 1 Change the designation for the input field ■ ■ ■ Tap on the input field Identification 1. The text editor opens.
■■■■■■■■■■■■■■■■■■■■■■ 16.5 Method options Defining limit values for the sample size In the dialog Sample data / Sample size limits, you can define the limit values for the sample size. These values are not monitored during sample data input, but rather: ■ ■ ■ ■ when the determination is started. when the determination is finished. upon automatic sample data query via the command REQUEST. when the determination is being recalculated.
■■■■■■■■■■■■■■■■■■■■■■ 16.5.4 16 Methods Method properties Main dialog: Edit parameters ▶ Method options ▶ Properties Detailed information concerning the method is displayed in the dialog Method options / Properties. ■ ■ ■ ■ ■ 16.5.5 Method status Current status of the method. – new The method has been newly created and not yet saved. – saved The current method has been saved. – modified The current method has been modified. Created by User who created the method.
■■■■■■■■■■■■■■■■■■■■■■ 16.5 Method options Save determination automatically on | off (Default value: off) If this parameter is activated, then the determination will be saved automatically. Memory Memory location where the determination will be saved. Selection Default value External memory 1 | External memory 2 | Shared memory External memory 1 Shared memory The determination will be saved in a shared directory on the network.
■■■■■■■■■■■■■■■■■■■■■■ 16 Methods ent of the write protection of the operating system on your computer. This write protection feature protects the saved determination data against accidental modifications or modifications by unauthorized persons. Create PC/LIMS report on | off (Default value: off) If this parameter is activated, then a machine-readable report with all of the important data for the determination is created, a so-called PC/LIMS report.
■■■■■■■■■■■■■■■■■■■■■■ 17 Control Main dialog: Control In the dialog Control, the settings for the execution of a single determination or of one sample series are defined. If you work with the login function activated, you can log off the system in this dialog. The login dialog will then be shown immediately. Statistics on | off (Default value: off) With this option, you can enable or disable the statistics calculations for individual determinations (see "Statistics", page 172).
■■■■■■■■■■■■■■■■■■■■■■ 17 Control Autostart on | off (Default value: off) If this option is activated, a new determination is started automatically at the end of a determination. This continues until the number specified has been reached (siehe Number of autostarts). NOTE If you deactivate this option while a determination is running, then that determination will be ended, although the next one will however also not be started. This can be used to cancel a sample series.
■■■■■■■■■■■■■■■■■■■■■■ [Delete statistics] Delete all statistics data. The statistics data should be deleted manually if a sample series has been canceled and a new sample series is to be started. The statistics data is deleted automatically in following cases: ■ ■ ■ when all of the determinations of the determination series have been carried out and a new determination has been started afterwards. when a new method is loaded.
■■■■■■■■■■■■■■■■■■■■■■ 17 Control value of the statistics counter corresponds to that in effect before the interruption. 5 Continue the sample series ■ 915 KF Ti-Touch Carry out the remaining determinations in the sample series.
■■■■■■■■■■■■■■■■■■■■■■ 18 Favorites Main dialog: Control ▶ Favorites Main dialog: System ▶ System settings ▶ User admin. ▶ Edit ▶ Favorites This chapter describes how you can create favorites. Favorites are comparable to the favorites/bookmarks in your Internet browser. In the main dialog, a button is created for each favorite. By pressing a single key, you can trigger an action without having to navigate through different dialogs.
■■■■■■■■■■■■■■■■■■■■■■ 18.1 18 Favorites Creating favorites The list of favorites gives the designation and position of the button on the main dialog of each favorite configured. [New] Add a new favorite to the list, see following chapter. [Delete] Delete the selected favorite from the list. [Edit] Edit the properties of the selected favorite, see following chapter. 18.1.
■■■■■■■■■■■■■■■■■■■■■■ 18.1 Creating favorites Position On the main dialog, a button is created for each favorite. These buttons are ordered in three rows at fixed positions. The position 1 can be found top left. Input range Default value 1 - 14 1 Name The designation of the favorite is used for unambiguous identification. Entry 24 characters maximum Type Definition as to whether the favorite represents a single method or a complete sample table.
■■■■■■■■■■■■■■■■■■■■■■ 18 Favorites Autostart on | off (Default value: off) If this option is activated, a new determination is started automatically at the end of a determination. This continues until the number specified has been reached (siehe Number of autostarts). Number of autostarts This option can only be edited when Autostart is activated. Number of automatic starts.
19.1 Entering sample data in the main dialog ■■■■■■■■■■■■■■■■■■■■■■ 19 Sample data You can enter the sample data (identification, sample size, etc.) in a variety of ways: ■ ■ ■ Directly in the main dialog. Using the sample table. This is particularly useful with sample series. The sample table is a table in which the sample data for up to 999 samples can be entered (see Chapter 20, page 181). Automatic request immediately after the start of the determination (see Chapter 19.2, page 179).
■■■■■■■■■■■■■■■■■■■■■■ 19 Sample data Identification 2 Sample identification. The sample identification can be used in calculations as the variable CI2. If you have defined a sample identification list (see Chapter 15.1.1, page 140) then the entries can be selected here. Entry Default value 24 characters maximum empty Sample size Sample size. The value of the sample size can be used in calculations as the variable C00. Input range Default value –99999999 - 9999999999 1.0 Unit of sample size.
19.2 Requesting sample data at the start of the determination ■■■■■■■■■■■■■■■■■■■■■■ To accomplish this, you must insert and configure the command REQUEST in the method (see Chapter 28.9.3, page 329). If the Hold sequence parameter is activated, then the run will be paused and must be continued with [Continue] after the sample data has been entered.
■■■■■■■■■■■■■■■■■■■■■■ 20 Sample table 20 Sample table 20.1 General Main dialog: Sample table The sample table is a table in which the sample data for up to 999 samples can be entered. If you are processing larger-sized sample series, it makes sense to enter the sample data in this table. Enabling the sample table Proceed as follows to enable the sample table: 1 Open the control dialog ■ In the main dialog, tap on [Control].
■■■■■■■■■■■■■■■■■■■■■■ 20.1 General The display will show the number of determinations already carried out and the total number of sample lines containing data. The sample table is still empty in this example. The sample table contains numbered lines; each line represents one determination. In the standard settings, the first sample identification and the sample size of each sample are displayed. In the properties, you define the columns to be displayed (see Chapter 20.3, page 186).
■■■■■■■■■■■■■■■■■■■■■■ 20 Sample table [Edit] Edit the data of the selected line (see Chapter 20.2, page 184). Saving a sample table NOTE Sample tables can only be saved on external storage media or on a shared file system. Proceed as follows to save a sample table: 1 Open the dialog ■ In the dialog Sample table, tap on the button [Load/ Save]. The dialog Sample table / Load/Save is displayed. If no external storage medium is plugged in or if no file system is shared, then [Save] is disabled.
■■■■■■■■■■■■■■■■■■■■■■ 20.2 Edit the sample data Loading a sample table Proceed as follows to load a sample table: 1 Display the list of saved sample tables ■ ■ ■ ■ ■ In the dialog Sample table, tap on the button [Load/ Save]. The dialog Sample table / Load/Save is displayed. If no external storage medium is plugged in or if no file system is shared, then [Load] is disabled. Tap on [Load]. The selection of memory locations is displayed.
■■■■■■■■■■■■■■■■■■■■■■ 20 Sample table At the very top you will see the line number of the selected line. In this example, the sample data of the first line is displayed. You can scroll between individual data sets with the keys [–] and [+]. Method Method used for processing the sample. If you use the sample assignment table (see Chapter 15.1.2, page 141), then the method can no longer be defined here.
■■■■■■■■■■■■■■■■■■■■■■ 20.3 Properties Balance connected and configured: Value from the connected balance. Fixed sample size defined in method: The value will be entered in the corresponding line when the method is loaded. The existing entry will be overwritten. NOTE The sample size limits defined in the method are not monitored when the sample data is being entered in the sample table. 20.3 Properties Method memory Memory location the method is loaded from.
■■■■■■■■■■■■■■■■■■■■■■ 20 Sample table Method Only the assigned method is displayed for each sample. Identification 1 The first sample identification and the sample size are displayed for each sample. Identification 2 The second sample identification and the sample size are displayed for each sample. Number of lines Maximum number of lines which can be present in the sample table.
■■■■■■■■■■■■■■■■■■■■■■ 20.3 Properties End line with Selection as to with which input the editing dialog for the next sample is automatically displayed. Selection Default value Sample size | Identification 1 | Identification 2 | manual Sample size manual The editing dialog for the next sample can be displayed with the [New sample] button in the editing dialog.
■■■■■■■■■■■■■■■■■■■■■■ 21 Determination run 21 Determination run This chapter provides a description of how to carry out a determination, i.e. how to process a single sample or a sample series. The current sample data is used for calculations. 21.1 Carrying out a single determination When you carry out a determination, you can enter the sample data in various ways (see Chapter 19, page 178). The following instructions describe how to enter the sample data in the main dialog.
■■■■■■■■■■■■■■■■■■■■■■ 21.2 Performing a sample series Sequence of the determination ■ 21.2 After a determination is started, a check is made as to – whether all of the devices required are connected and ready for operation. – whether all titrants and sensors are present. – whether the monitoring intervals for titrants, sensors and common variables are still valid. – if a sample assignment table is activated (see Chapter 15.1.
■■■■■■■■■■■■■■■■■■■■■■ 21.3 21 Determination run Canceling determinations manually You can cancel a determination at any time with the fixed key [ ]. The command being carried out at that moment is canceled and no further commands are executed. If you are working with the autostart function activated (see Chapter 17, page 170), the entire series will be canceled. NOTE If you do not wish to cancel the entire determination, but only a single command, use the function [Cancel command] in the live display.
■■■■■■■■■■■■■■■■■■■■■■ 22.1 Editing the sample data of the running determination 22 Live modifications 22.1 Editing the sample data of the running determination The sample data can be entered or modified in the main dialog while a determination is running. The sample data entered at the end of the determination is always used in calculations. Proceed as follows to edit the sample data: 1 Display the main dialog ■ Tap on the fixed key [ ]. The main dialog is displayed.
■■■■■■■■■■■■■■■■■■■■■■ 22.2 22 Live modifications Editing the sample table while a determination is running You can insert new lines or delete existing ones or edit sample data while a determination is running. NOTE To ensure that no problems arise during the run and that the current data is always available for the calculation, we recommend that the input dialogs always be closed.
■■■■■■■■■■■■■■■■■■■■■■ 22.3 Live display The live dialog is displayed once again. 22.3 Live display Main dialog: Live display The current method and the system status are displayed in the title line. The currently running command is displayed directly underneath the title line. The following functions are available to you while a determination is running: ■ ■ ■ ■ ■ ■ [Hold] Pause the determination. The label changes to [Continue]. This can be used to continue the run.
■■■■■■■■■■■■■■■■■■■■■■ 22 Live modifications View Type of curve display. Selection Default value Curve | Curve & meas. values | Measured values | Curve & sample data Curve & meas. values Curve Only the curve is displayed. Curve & meas. values The measured values are displayed to the right next to the curve. Measured values Only the measured values are displayed. Curve & sample data The sample data is displayed to the right next to the curve. [Meas.
■■■■■■■■■■■■■■■■■■■■■■ 22.4 Live parameters 22.4 Live parameters Live display: Edit parameters Certain method parameters can be edited while a determination is being carried out. The modified parameters are taken into account at once. If you modify, for instance, the start conditions after the start volume has been dosed, then these modifications will not be taken into account until the next determination.
■■■■■■■■■■■■■■■■■■■■■■ 22 Live modifications 2 Edit the method parameters ■ Change the desired parameters accordingly. 3 Display the live dialog ■ ■ Tap on the fixed key [ ]. The main dialog is displayed. Tap on [Live display]. The live dialog is displayed once again.
■■■■■■■■■■■■■■■■■■■■■■ 23 Results and more determination data Main dialog: Results Detailed specifications concerning the current determination are displayed in the Results dialog: ■ ■ ■ ■ ■ Current method Method with which the determination was carried out. Determination Name of the determination. Displayed only if the determination has been saved. Determination time Date and time at which the determination was carried out. Results Calculated results.
■■■■■■■■■■■■■■■■■■■■■■ 23 Results and more determination data [Recalculate] Recalculate the determination. The procedure will be executed immediately. Recalculation cannot be undone (see Chapter 23.8, page 212). [Curve] Display the curve of the current determination (see Chapter 23.7, page 210). [More data] Display detailed specifications for the determination (determination properties, measuring point list, endpoint details, etc.), save determination (see Chapter 23.1, page 199). 23.
■■■■■■■■■■■■■■■■■■■■■■ 23.1 More determination data [Messages] Display list with all of the messages which occurred during the determination (see Chapter 23.2, page 202). [View data] Display all of the details for the selected command. 23.1.1 Details Results dialog: More data ▶ View data All of the variables generated by the determination are listed in the More determination data / View data dialog. These can be used in calculations.
■■■■■■■■■■■■■■■■■■■■■■ 23 Results and more determination data The individual measured values are listed in the View data / Measuring point list dialog. You can configure which columns are displayed ([View] button). These settings are saved separately for each mode (KFT, MEAS, etc.). Column 1 / Column 2 / Column 3 Quantity displayed in the columns 1, 2 and 3. Both the selection of the quantities and the default setting depend on the mode.
■■■■■■■■■■■■■■■■■■■■■■ 23.2 Messages NOTE Recalculation: The titer of the titrant can be modified retroactively. The values displayed here are used for the recalculation of a loaded determination. The values saved in the dialog System / Titrants are not changed as a result. 23.2 Messages Results dialog: More data ▶ Messages If messages occurred during a determination, they will be displayed in the More determination data / Messages dialog.
■■■■■■■■■■■■■■■■■■■■■■ 23.4 23 Results and more determination data Determination properties Results dialog: More data ▶ Properties Detailed information concerning the determination is displayed in the dialog More determination data / Properties: ■ ■ ■ ■ ■ ■ ■ 915 KF Ti-Touch User User who carried out the determination. Sample number Number which was entered in the dialog Control at the start of the determination.
■■■■■■■■■■■■■■■■■■■■■■ 23.5 Loading a determination 23.5 ■ Status of deter. – original The determination has not been recalculated. – recalculated The determination has been recalculated. – loaded The determination has been loaded. – saved The determination has been saved. The version number indicates how many times the file has been saved. ■ Used devices Hardware that was used for the determination. – Ti-Touch with program version of the software and the serial number.
■■■■■■■■■■■■■■■■■■■■■■ 23 Results and more determination data 2 Selecting a determination ■ Select the desired determination. 3 Loading the determination ■ Tap on [Load]. The determination is now loaded. The properties are displayed in the results dialog. NOTE When you load a determination, the method with which the determination was carried out will be loaded automatically.
■■■■■■■■■■■■■■■■■■■■■■ 23.5 Loading a determination In the Show files / View dialog you can depict the individual determinations in greater detail. The sorting criterion and the number of saved determinations are displayed at the very top. The list contains numbered lines; each line represents one determination. In the default settings, the first sample identification and the first calculated result of each determination are displayed.
■■■■■■■■■■■■■■■■■■■■■■ 23 Results and more determination data In the View / Properties dialog, you can define which data is displayed in the columns and the criteria according to which the determinations are to be sorted. Column 1 Selection of what is to be displayed in the first column. Selection Default value Identification 1 | Identification 2 | Determ. time | Result name | Method | Sample size | User Identification 1 Column 2 Selection of what is to be displayed in the second column.
■■■■■■■■■■■■■■■■■■■■■■ 23.5 Loading a determination Result Sort according to the numerical value of the result. Method Sort alphabetically according to the method name. Sample size Sort according to the numerical value of the sample size. User Sort alphabetically according to the user who carried out the determination. Sort sequence Selection Default value ascending | descending descending [Filter] Specify the filter criteria for the list.
■■■■■■■■■■■■■■■■■■■■■■ 23 Results and more determination data no filter The list itself will not be filtered. Identification 1 In the input field is equal to, select or enter the first sample identification according to which filtering is to be carried out. Identification 2 In the input field is equal to, select or enter the second sample identification according to which filtering is to be carried out.
■■■■■■■■■■■■■■■■■■■■■■ 23.6 Saving a determination 23.6 Saving a determination Results dialog: More data ▶ Load/ Save ▶ Save Determinations can be saved as follows: ■ ■ Manually in this dialog. The suggested default setting for file names is the first 16 characters of Identification 1, followed by the determination time in the format YYYYMMDD-hhmmss. Extending the file name with the determination time has the advantage that the file names of all determinations are unique.
■■■■■■■■■■■■■■■■■■■■■■ 23 Results and more determination data [Endpoint details] Display details for each endpoint found (see "Endpoint details", page 201). [Measuring point list] Display measuring point list of titrations and measurements (see "Measuring point list", page 200). [Curve options] Define the settings for the curve display (see "Curve options", page 211).
■■■■■■■■■■■■■■■■■■■■■■ 23.8 Recalculation and reevaluation y2 axis Quantity which is shown on the y2 axis of the graph. Selection Default value none | Volume | Measured value | Time | Temperature | Volume drift | Signal drift none Color Selection of a color for the curves. Grid on | off (Default value: on) If this parameter is activated, grid lines are shown. Display measuring points on | off (Default value: off) If this parameter is activated, the curve is shown with the single measured values. 23.
■■■■■■■■■■■■■■■■■■■■■■ 23 Results and more determination data ■ ■ EVAL commands: All evaluations are carried out again. You can also insert new evaluations retroactively. CALC command: All calculations are carried out again. You can also define new calculations retroactively. If you subsequently modify the titrant in a method, then the variable TITER of the "new" titrant will be used.
■■■■■■■■■■■■■■■■■■■■■■ 24 Statistics Main dialog: Results ▶ Statistics In the dialog Results, you can use [Statistics] to display the statistical overview of a determination overview. You can statistically evaluate a maximum of nine results calculated in a determination. A statistical series can contain a maximum of 20 determinations. Activating the statistics function Proceed as follows to activate the statistics function: 1 Edit the method options ■ ■ ■ ■ In the main dialog, tap on [Edit parameters].
■■■■■■■■■■■■■■■■■■■■■■ 24 Statistics In this dialog, the method with which the determination was carried out is displayed in the first line. The second line shows how many determinations have already been carried out and the total number of determinations to be carried out. The name of the result and the mean value are shown for each result. [Add determ.] Add a further determination to the determination series (see Chapter 24.3, page 217).
■■■■■■■■■■■■■■■■■■■■■■ 24.1 Displaying details for a result 24.1 Displaying details for a result Results dialog: Statistics ▶ Details In the dialog Statistics / Details, the following details are displayed: ■ ■ ■ ■ ■ Result name Mean value On the right-hand side of the dialog is displayed the number of single results from which the mean value was calculated. In this example, it is 3.
■■■■■■■■■■■■■■■■■■■■■■ 24 Statistics NOTE If the mean value has been assigned to a common variable or to the variable TITER (see command CALC), then the determination must be recalculated manually (see Chapter 23.8, page 212). 24.2 Deleting statistical data In the dialog Control, you can delete all of the statistics data manually with the function [Delete statistics]. The statistics data is deleted automatically in the following cases: ■ ■ 24.
24.3 Adding a determination to a determination series ■■■■■■■■■■■■■■■■■■■■■■ NOTE The number of samples defined in the method options does not, however, change as a result.
■■■■■■■■■■■■■■■■■■■■■■ 25 Result table 25 Result table Main dialog: Results ▶ Result table The result table is suitable for displaying in detail the results of all of the determinations performed on a single day, for instance. You can save the results of up to 200 determinations. A maximum of 9 results from a single determination can be saved in the result table. You define which results from this table are to be saved using the CALC command.
■■■■■■■■■■■■■■■■■■■■■■ 25.1 Properties [Delete] Delete the selected line from the result table. [Details] Display details concerning the selected determination. "Result table / Details" dialog The following determination data is displayed in this dialog: ■ ■ ■ ■ ■ 25.1 User User who carried out the determination. Method Method with which the determination was carried out. Determ. time Date and time at which the determination was carried out. Sample data (Identification 1 and 2 and sample size).
■■■■■■■■■■■■■■■■■■■■■■ 25 Result table Column 2 Selection of what is to be displayed in the second column. Selection Result 1…9 The numbering of the results corresponds to the order in which the results were calculated in the determination run. Sort by Selection according to which criterion the list is to be sorted. Selection Default value Identification 1 | Identification 2 | Determ. time | Result | Method | Sample size | User Determ.
■■■■■■■■■■■■■■■■■■■■■■ 25.1 Properties [Delete automat.] Define the settings for the automatic deletion of the result table (see "Deleting the result table automatically", page 223). Filter criteria Result table: Properties ▶ Filter You can set filters in the Properties / Filter dialog. A filter is a rule defining what is to be displayed in the result table. If a filter has been set, this is indicated with a corresponding symbol in the Results / Result table dialog.
■■■■■■■■■■■■■■■■■■■■■■ 25 Result table Result In the input fields Value from … to, specify the value range according to which filtering is to be carried out. Method In the field is equal to, select or enter the method name according to which filtering is to be carried out. Sample size In the input fields Value from … to, specify the value range according to which filtering is to be carried out. User In the field is equal to, select or enter the user according to whom filtering is to be carried out.
■■■■■■■■■■■■■■■■■■■■■■ 25.2 Saving the result table Save result table on | off (Default value: on) If this option is activated, the result table will be deleted as soon as you have saved it. 25.2 Saving the result table NOTE When you save the result table, the complete determination will not be saved with each entry. If you wish to save the determination data for each determination, use the Save determination automatically function (see Chapter 16.5.6, page 167). 25.
■■■■■■■■■■■■■■■■■■■■■■ 26 Printing 26 Printing Reports can be printed out as follows: ■ ■ Manually with the [ ] fixed key in this dialog. The fixed key is disabled while a determination is running. Automatically in the determination run with the command REPORT (see Chapter 28.5, page 299). How to configure a printer is described in Chapter 11.7, page 101. This screenshot shows the printing dialog associated with the main dialog.
■■■■■■■■■■■■■■■■■■■■■■ More reports Fixed key [ ]: More reports The dialog Print / More reports displays a list of all of the available reports. NOTE If you are using a CUSTOM Neos printer, you can print only the reports marked with a #. [System reports] Opening the list of reports for the system settings. NOTE If you are using a CUSTOM Neos printer, you cannot print system reports. [Method reports] Opening the list of reports for the current method.
■■■■■■■■■■■■■■■■■■■■■■ 26 Printing [Edit] Edit the settings of the selected report. [Print] Print the selected report. 26.1 General report options Fixed key [ ]: Report options The settings for the report printout can be adjusted in the dialog Print / Report options to indicate, for example, whether a report header, a signature line or a frame is to be printed. Report header Defining the output of the report header.
■■■■■■■■■■■■■■■■■■■■■■ 26.2 Settings of the individual reports on each page The signature line will be printed on every page. Frame on | off (Default value: on) If this parameter is activated, a frame is printed as lateral marking. 26.
■■■■■■■■■■■■■■■■■■■■■■ 26.3 26 Printing List of all printable reports The following reports can be printed out with the fixed key [ ]: NOTE If you are using a CUSTOM Neos printer, you can print only the reports that are marked with a # on the display. Report Contents Result report Report with determination properties, sample data, calculated results, etc. If there are multiple determinations, the statistics will also be printed out. Curve Curve report.
■■■■■■■■■■■■■■■■■■■■■■ 26.3 List of all printable reports Report Contents Result table Result table List of all determinations in the result table with results and with the determination data, as saved in the result table. More reports ▶ Method reports Method sequence Method properties and the list of all method commands. Parameters full Method properties and options, all method commands with all parameters. All parameters which no longer have their default settings will be printed in bold.
■■■■■■■■■■■■■■■■■■■■■■ 26 Printing Report Contents Sensor list List of all the sensors configured in the system. All sensor data short The most important sensor data for all sensors. All sensor data full All of the sensor data for all of the sensors (name, working life, etc.). Sensor data short The most important sensor data of the selected sensor, context-sensitive only from the editing dialog.
■■■■■■■■■■■■■■■■■■■■■■ 26.3 List of all printable reports Report Contents Sample rack list List with all of the sample racks configured in the system, together with designation, number of positions and rack code.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control 27 Manual control The following functions are available in the manual control: [Measure] Carrying out manual measurements (see Chapter 27.2, page 235). Measuring modes: ■ ■ ■ Ipol (voltametric measurement with selectable polarization current) Upol (amperometric measurement with selectable polarization voltage) T (temperature measurement) [Dosing] Dosing manually (see Chapter 27.3, page 237).
■■■■■■■■■■■■■■■■■■■■■■ 27.1 Opening and closing the manual control 27.1 Opening and closing the manual control Opening the manual control 1 Open the dialog window ■ Tap on the fixed key [ ]. Manual control opens: 2 Select the function ■ Tap on the button for the desired function. If the function can be carried out by several devices, a dialog to select the device will appear first; otherwise the dialog for the selected function will appear immediately.
■■■■■■■■■■■■■■■■■■■■■■ 27.2 27 Manual control Measuring Manual control ▶ Measure Manual measurements can be carried out with the function [Measure]. Proceed as follows: 1 Select a sensor ■ ■ Select the desired sensor from the sensor list. All of the sensors available in the sensor list are displayed. Sensors are defined under System ▶ Sensors. Select the measuring input at which the sensor is connected. 2 Select a measuring mode ■ Select the desired measuring mode.
■■■■■■■■■■■■■■■■■■■■■■ 27.2 Measuring The ongoing measurement is canceled with [Stop]. The label switches once again to [Start]. 27.2.1 Parameter description Dialog "Manual control / Measure" Sensor Selection of the sensor from the sensor list. Sensors are defined under System ▶ Sensors. Selection Selection of configured sensors Measuring input Selection of the measuring input the sensor is connected to. The selection depends on whether the control device has one or two measuring interfaces.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control Temperature Temperature entered manually. If a temperature sensor is connected then the temperature will be measured continuously. Input range Default value –20.0 - 150.0 °C 25.0 °C [Info sensor] Display information on the sensor. 27.3 Dosing Manual control ▶ Dosing The following dosing functions are available: ■ ■ ■ ■ ■ Preparing the exchange or dosing unit (see Chapter 27.3.3, page 244) Emptying the dosing unit (see Chapter 27.3.
■■■■■■■■■■■■■■■■■■■■■■ 27.3 Dosing ■ ■ ■ ■ Titrants For dosing devices of the type 8XX with an intelligent exchange unit or dosing unit the titrant name is read off the data chip. Cylinder volume For dosing devices of the type 8XX with an exchange unit without data chip the cylinder volume is displayed only. --For exchange or dosing units which are not or not correctly attached. Manual busy When a dosing function is being carried out.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control ■ Type 8XX without integrated data chip: If an exchange unit without integrated data chip is attached, then the titrant can be selected from the titrant list. We recommend always selecting the titrant. This ensures that the parameters defined for the titrant for the preparing are used. Titrants are defined under System ▶ Titrants. [Empty] Empty the cylinder and the tubings of the dosing unit (see Chapter 27.3.4, page 245).
■■■■■■■■■■■■■■■■■■■■■■ 27.3 Dosing With the function [Dosing], dosing is continuous for as long as you hold the button [Dose] pressed down. Proceed as follows: 1 Configure the dosing function NOTE ■ ■ ■ ■ The dosing and filling rates should be decreased for viscous liquids. The maximum dosing and filling rates depend on the cylinder volume (see Chapter 31, page 338). Enter the dosing rate. Enter the filling rate. 2 Start dosing ■ Tap on [Dose]. The volume dosed is displayed.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control NOTE If you exit the dialog with the fixed key [ will be filled automatically. ], the dosing cylinder Parameter description Dosing rate Rate at which it is dosed. Input range Selection Default value 0.01 - 166.00 mL/min maximum | dynamic maximum maximum The maximum dosing rate depends on the cylinder volume (see Chapter 31, page 338). dynamic This setting can only be selected when the dosing cylinder is filled.
■■■■■■■■■■■■■■■■■■■■■■ 27.3 Dosing You can dose a particular volume with the [Dosing fixed vol.] function. Proceed as follows: 1 Configure the dosing function NOTE ■ ■ ■ ■ ■ The dosing and filling rates should be decreased for viscous liquids. The maximum dosing and filling rates depend on the cylinder volume (see Chapter 31, page 338). Enter the desired volume. Enter the dosing rate. Enter the filling rate. 2 Start dosing ■ Tap on [Start]. The volume display is shown.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control Continuous dosing is paused with [Hold]. The label changes to [Continue]. This can be used to continue the run. 3 Fill the cylinder With the default settings (see "Fill automatic.", page 244) the dosing cylinder is filled automatically. Otherwise: ■ Tap on [Fill]. The dosing cylinder is filled. The displayed volume value will be reset to 0.0000 mL. Parameter description Volume Volume which is dosed. Input range Default value 0.00000 - 99999.9 mL 0.
■■■■■■■■■■■■■■■■■■■■■■ 27.3 Dosing Input range Selection Default value 0.01 - 166.00 mL/min maximum maximum maximum The maximum filling rate depends on the cylinder volume (see Chapter 31, page 338). Fill automatic. on | off (Default value: on) If this parameter is activated, then the dosing cylinder is being automatically filled at the end of the dosing. 27.3.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control Figure 13 27.3.4 Live display "Preparing the exchange unit" Emptying Manual control: Dosing ▶ Empty NOTE The function Empty is possible only with dosing units. The cylinders and the tubings of the dosing unit are emptied with this function. If the titrant is selected (see "Selecting the dosing function", page 238), the parameters defined for the titrant for preparing/emptying and the tubing parameters will be used.
■■■■■■■■■■■■■■■■■■■■■■ 27.4 Stirring 27.3.6 Replacing reagent NOTE You can easily change the reagent in a dosing unit without no contact with the chemicals using the functions Prepare and Empty. Proceed as follows: 1 Execute the function [Empty]. 2 Attach the dosing unit to the bottle with a suitable solvent. 3 Execute the function [Prepare]. The last traces of the reagent in the cylinder and tubings are removed with the solvent. 4 Execute the function [Empty] again.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control The MSB connector and type of stirrer is indicated for each connected stirrer. You can control a connected stirrer manually with the function [Stir]. Proceed as follows: 1 Set the stirring rate ■ Tap on the button [–] or [+]. Each time one of the buttons is tapped, the stirring rate is reduced or increased by one level. 2 Switch on the stirrer ■ Tap on [Start]. The stirrer is started. 3 Switch off the stirrer ■ Tap on [Stop]. The stirrer is stopped.
■■■■■■■■■■■■■■■■■■■■■■ 27.5 Remote 27.5 Remote Manual control ▶ Remote With the function [Remote], you can define manual output signals to the remote interface of a connected Remote Box. It is not possible to define input signals; the current status of the input lines is, however, displayed. Inputs Current status of the input lines. Outputs Selection of the signal out of the templates or entering the required bit pattern. Templates are defined under System ▶ Templates ▶ Output lines.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control [Set] Set a defined output signal. 27.6 USB Sample Processor Manual control ▶ Sample Processor With the function [Sample Processor], you can control a connected Sample Processor manually. The following functions are available: ■ ■ ■ ■ ■ ■ Moving a sample rack Moving the lift Defining specific lift positions (work position, rinse position, etc.
■■■■■■■■■■■■■■■■■■■■■■ 27.6 USB Sample Processor NOTE Note that the rack has to be set to a valid position for all lift movements! Otherwise a corresponding error message will be displayed. Move to a valid rack position with [Previous] or [Next]. Alternatively, you can carry out a MOVE command. [Home position] Move the lift to the home position (corresponds to lift position 0 mm). [Work position] Move the lift to the work position of the current rack position or the external position.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control [Rack position] Move the rack to any sample position or to defined positions which are defined as special beakers (see Chapter 27.6.2, page 253). [External position] This button is only enabled when a Swing Head is mounted to the Sample Processor. Swing the robotic arm to an external position outside the sample rack and assign the swing angle and the lift position (see Chapter 27.6.3, page 254). 27.6.
■■■■■■■■■■■■■■■■■■■■■■ 27.6 USB Sample Processor Input range Selection Default value 0 - 'maximum stroke path' mm A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position. The maximum stroke path is defined in the properties of the tower (see "Maximum stroke path", page 88). If a higher value is entered, an error message will be displayed.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control 2 Assign the lift position ■ Tap on [Assign lift pos.]. The dialog Lift / Assign lift position is displayed. ■ Select the desired specific lift position and tap on [Assign]. The current lift position is assigned to the specific lift position. 27.6.2 Moving to a rack position Manual control: Sample Processor ▶ Rack position In the dialog Sample Processor / Rack position, you can move to any position on the attached rack. Current rack pos.
■■■■■■■■■■■■■■■■■■■■■■ 27.6 USB Sample Processor Input range Default value Selection 1 - Number of positions on the rack attached. 1 Special beaker 1…16 Special beaker 1…16 The selection depends on how many rack positions are defined as special beakers (see "Editing special beakers", page 98). Shift rate Rate at which the sample rack is moved. Input range Default value 3 - 20 °/s 20 °/s Shift direction Direction in which the rack is moved.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control NOTE This dialog is accessible only if you have a Swing Head with robotic arm attached to the Sample Processor. In the dialog Sample Processor / External position, you can swivel the robotic arm to an external position outside of the sample rack. As with the rack positions, here you can also define up to four frequently used external positions as specific positions. You can assign specific lift positions to these positions, analogous to the rack positions.
■■■■■■■■■■■■■■■■■■■■■■ 27.6 USB Sample Processor [⇦] Swing the robotic arm in clockwise direction as long as the button is pressed down. [⇨] Swing the robotic arm in counterclockwise direction as long as the button is pressed down. Swing rate Rate at which the robotic arm is swung in the manual control. Input range Default value 10 - 55 °/s 55 °/s Move to ext. pos. Selecting a predefined position or entering any swing angle. Input range Selection (Offset) - (Offset + max.
■■■■■■■■■■■■■■■■■■■■■■ 27 Manual control 2 Assign the external position ■ Tap on [Assign Ext. pos.]. The dialog External position / Assign swing angle is displayed. ■ Select the desired external position and tap on [Assign]. The current angle position of the robotic arm is assigned to the external position. 3 Move to the external position ■ ■ Tap on [ ]. The next higher-level dialog is displayed. Select Move to ext. pos. = External position X (X = 1…4) and tap on [Start].
■■■■■■■■■■■■■■■■■■■■■■ 27.6 USB Sample Processor 5 Assign lift position ■ Tap on [Assign lift pos.]. The dialog Lift / Assign lift position is displayed. ■ Select the desired specific lift position. Tap on [Assign]. ■ The current lift position is assigned to the specific lift position.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters 28 Parameters NOTE For most of the numerical parameters, the result of a calculation can also be utilized in place of a number. A description of this can be found in the appendix (see Chapter 31.6, page 344). 28.1 Volumetric Karl Fischer titrations (KFT) 28.1.1 Start conditions The parameters that are carried out before the start of titration are defined under [Start conditions]. Pause 1 Waiting time, e.g.
■■■■■■■■■■■■■■■■■■■■■■ 28.1 Volumetric Karl Fischer titrations (KFT) Initial measured value The conditions for the entry of the first measured value in the measuring point list are defined under [Initial meas. value]. Signal drift Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time. Measuring mode Ipol: Input range 0.
■■■■■■■■■■■■■■■■■■■■■■ 28.1.2 28 Parameters Control parameters The control parameters for the endpoint are defined under [Control parameters]. Endpoint at Measured value for the endpoint. The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the 32 Technical specifications chapter, page 355. Measuring mode Ipol: Input range –2,000.0 - 2,000.0 mV Default value 250.
■■■■■■■■■■■■■■■■■■■■■■ 28.1 Volumetric Karl Fischer titrations (KFT) controlled. The closer the endpoint, the slower the dosing until the volume increment defined under Min. increment has been reached. The larger the control range, the slower the titration. Outside the control range, dosing is carried out continuously, and the dosing rate is defined under Max. rate. The values that are actually measurable may vary from the values specified here, depending on the hardware.
■■■■■■■■■■■■■■■■■■■■■■ Table 6 28 Parameters Default values of the predefined titration rates for KFT Titration rate slow optimal fast – Ipol 300.0 mV 100.0 mV 30.0 mV – Upol 40.00 µA 10.00 µA 5.00 µA Max. rate 1.00 mL/min maximum maximum Min. increment minimum (= cylinder volume/10,000) minimum (= cylinder volume/10,000) 5.00 µL Dynamics Stop criterion Stop criterion The titration is stopped when the endpoint has been reached and this stop criterion has been fulfilled.
■■■■■■■■■■■■■■■■■■■■■■ 28.1 Volumetric Karl Fischer titrations (KFT) Stop drift This parameter can only be edited with Stop criterion = drift. The titration is stopped when the endpoint and the stop drift have been reached. Input range Default value 1 - 999 µL/min 20 µL/min Delay time This parameter can only be edited with Stop criterion = time. When the endpoint has been reached, the specified time is allowed to elapse after the last dosing and the titration is then stopped.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Extraction time Minimum duration of the titration. The titration will not be stopped during the extraction time, even if the endpoint has already been reached. The titration is however stopped if a stop condition is fulfilled during this time (see Chapter 28.1.4, page 265). The entering of an extraction time is e.g. recommended with samples that release water slowly or if a Karl Fischer oven is used.
■■■■■■■■■■■■■■■■■■■■■■ 28.1 Volumetric Karl Fischer titrations (KFT) Selection Default value off off Filling rate Rate at which the dosing cylinder is filled. Input range Selection Default value 0.01 - 166.00 mL/min maximum maximum maximum The maximum filling rate depends on the cylinder volume (see Chapter 31, page 338). 28.1.5 Conditioning The conditions required for conditioning are defined under [Conditioning].
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters off No drift correction takes place. Drift value This parameter can only be edited with Drift correction = manual. Drift for manual drift correction. Input range Default value 0.0 - 99.9 µL/min 0.0 µL/min Measured value display on | off (Default value: off) If this parameter is activated, the currently measured value is displayed during the conditioning. Conditioning options [Cond. options] Cond.
■■■■■■■■■■■■■■■■■■■■■■ 28.1 Volumetric Karl Fischer titrations (KFT) NOTE This parameter must be activated if commands follow which must not run simultaneously with the conditioning. Example: Emptying the titration cell followed by adding new working medium. 28.1.6 Cell The monitoring of the reagent used is activated or deactivated under [Cell]. Reagent monitoring Selection of the reagent from the list of reagents. The selection depends on the titration mode.
■■■■■■■■■■■■■■■■■■■■■■ 28.1.8 28 Parameters Sensor The parameters for the sensor are edited under [Sensor]. Measuring input Selection of the measuring input the sensor is connected to. The selection is not dependent on whether the control device has one or two measuring interfaces. Selection Default value 1|2 1 Sensor Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors.
■■■■■■■■■■■■■■■■■■■■■■ 28.1 Volumetric Karl Fischer titrations (KFT) Electrode check on | off (Default value: off) For the following electrodes, an electrode check can be carried out: ■ ■ ■ pH electrodes Metal electrodes Ion-selective electrodes A check is made that the electrode is properly connected and that no short-circuit is present. The electrode check is carried out when this command is started. Temp. measurement Type of temperature measurement.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without integrated data chip, the cylinder volume and the type of dosing drive are checked.
■■■■■■■■■■■■■■■■■■■■■■ 28.2 Measurements (MEAS) 28.2 Measurements (MEAS) 28.2.1 Measuring parameters Under [Measuring parameters], the parameters influencing the sequence of the entire measurement are defined. Signal drift Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters off Measured value acceptance will take place after the maximum waiting time has elapsed. Min. waiting time The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements. Input range Default value 0 - 999999 s 0s Max.
■■■■■■■■■■■■■■■■■■■■■■ 28.2 Measurements (MEAS) Temperature This parameter is not available with the command MEAS T (temperature measurement). Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp. measurement = automatic or continuous is defined (see sensor dialog of the command). Input range Default value –20.0 - 150.0 °C 25.0 °C Time interval MP Time interval for entering a measuring point in the measuring point list.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Selection Default value 1|2 1 Sensor Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors. You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.
■■■■■■■■■■■■■■■■■■■■■■ 28.2 Measurements (MEAS) Temp. measurement Type of temperature measurement. Selection Default value continuous | automatic | off automatic continuous A temperature sensor must be connected. The temperature is measured continuously. automatic If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters). off The temperature will not be measured.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Switch off automatically on | off (Default value: on) If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc. 28.3 Evaluations (EVAL) Various additional evaluations (EVAL commands) can be carried out for titrations and measurements. The evaluation always refers to the last titration or measurement prior to the EVAL command.
■■■■■■■■■■■■■■■■■■■■■■ 28.3 Evaluations (EVAL) Fixed quantity Selection of the quantity to which the associated value is interpolated from the measuring point list. Selection Default value Measured value | Time | Volume Measured value Fixed EP1 at The fixed endpoint must lie between the first and the final entry in the measuring point list. The values that are actually measurable may vary from the values specified here, depending on the hardware.
■■■■■■■■■■■■■■■■■■■■■■ 28.3.2 28 Parameters Minimum and maximum evaluation (EVAL MIN/MAX) For the minimum or maximum measured value the associated volume, time and temperature are interpolated from the measuring point list. The evaluation begins as soon as the slope of the curve exceeds a particular threshold value.
■■■■■■■■■■■■■■■■■■■■■■ 28.3 Evaluations (EVAL) Measuring mode T: Input range Default value 28.3.3 0.1 - 20.0 °C/s 1.0 °C/s Break point evaluation (EVAL BREAK) A break point evaluation is used to determine sharp changes of direction in the titration curve. This evaluation is primarily used for photometric and conductivity titrations. The method is based on the search for extremes in the second derivative of the curve.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Set windows This parameter allows to evaluate break points only in a certain range (socalled window) of the curve. Break points outside this window will not be evaluated. Only one window can be defined. Within this window however, all break points are evaluated. Selection Default value off | Measured value | Time | Volume off off The entire titration curve is being evaluated. Measured value The window is defined on the measured value axis.
■■■■■■■■■■■■■■■■■■■■■■ 28.3 Evaluations (EVAL) Upper limit Measured value, time or volume for the upper limit. The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the 32 Technical specifications chapter, page 355. Measured value (measuring mode Ipol): Input range –2,000.0 - 2,000.0 mV Default value 2,000.0 mV Measured value (measuring mode Upol): –200.00 - 200.00 µA Input range Default value 200.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters [Edit] Edit the settings of the selected window. Lower limit Time for the lower limit. Input range Default value 0 - 999999 s 0s Upper limit Time for the upper limit. Input range Default value 28.4 0 - 999999 s 999999 s Calculations The following calculation commands are available: 28.4.1 CALC Defining calculations of a determination.
■■■■■■■■■■■■■■■■■■■■■■ 28.4 Calculations The result name is specified in the list for each calculation. [New] Define a new calculation or load an existing template, see following chapter. [Delete] Delete the selected calculation from the list. [Edit] Edit the data of the selected calculation (see Chapter 28.4.1.2, page 287).
■■■■■■■■■■■■■■■■■■■■■■ 28.4.1.1 28 Parameters Loading a result template The most common calculations are already available, the so-called result templates, in order to facilitate the definition of a new calculation. You also have the option, however, of defining templates of your own (see dialog System / Templates). Proceed as follows to load an existing template: 1 Select a result template ■ In the dialog Sequence / Edit command, tap on the button [New].
■■■■■■■■■■■■■■■■■■■■■■ 28.4 Calculations 2 Confirm the message ■ Tap on [OK]. The Note dialog is displayed. This note is defined in the dialog Edit result template / Note for wizard. 3 Enter the numerical values of the wildcards ■ Tap on [Next]. The list of wildcards is displayed: ■ Enter the numerical values of the wildcards. Tap on [Next].
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters The calculation can now be edited further, e.g. define result limits, see following chapter. 28.4.1.2 Editing a calculation Result name The result name is the text which will be shown in the result display and in the report. The default name corresponds to the result variable. Entry Default value max. 24 characters R1…R9 Calc. formula R1...R9 Shows the calculation formula. A special editor is opened for the definition (see Chapter 28.4.3, page 293).
■■■■■■■■■■■■■■■■■■■■■■ 28.4 Calculations [Note] Entering a note on the calculation. [Result variable] Modifying the result variable. [Result limits] Defining the limits for the monitoring of the result. [Result options] Defining additional settings for the calculation. Dialog "Edit calculation / Note" In this dialog you can enter a short text, e.g. to describe the variables used.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Upper limit When this value is exceeded, the action defined in the following is triggered. Input range Default value 0 - 9999999999 9999999999 Action Selection of the action, which is carried out when the result limits are exceeded. Selection Default value Display message | Document message | Cancel determination Display message Display message A message is displayed. You can select whether you want to continue with the determination or cancel the run.
■■■■■■■■■■■■■■■■■■■■■■ 28.4 Calculations NOTE The titer is assigned to the titrant defined in the last titration command before the calculation. Please note that the CALC command containing the titer assignment is inserted after the titration command with which the titer is determined. Selection Default value Single value | Mean value | off off Single value The result of the current determination is saved as titer.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Precision Setting, with which accuracy the result is used in additional calculations. Selection Default value Round | Truncate | Full precision Round Round The result is rounded to the defined number of decimal places (commercial rounding, in accordance with the US Pharmacopeia USP). If the digit at the first dropped decimal place is 1, 2, 3 or 4, then it will be rounded off; if this digit is 5, 6, 7, 8 or 9, then it will be rounded up.
■■■■■■■■■■■■■■■■■■■■■■ 28.4 Calculations Calculation formula Shows the calculation formula. A special editor is opened for the definition (see Chapter 28.4.3, page 293). max. 100 characters empty Entry Default value Decimal places Number of decimal places used to display the result. 0-5 2 Input range Default value Result unit The result unit is displayed and saved together with the result. max.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters 3 Display the live display ■ Tap twice on [ ]. The current result is displayed, 28.4.3 Formula editor The formulas for the calculations are entered with the formula editor. The maximum formula length is 100 characters. The formula editor contains buttons for entering numbers, mathematical operators and variables. The variables can be selected from lists. This is advisable in order to avoid spelling mistakes.
■■■■■■■■■■■■■■■■■■■■■■ 28.4 Calculations Key Description ⇦ The cursor within the input field is shifted to the left by one character. ⇨ The cursor within the input field is shifted to the right by one character. If the method contains more than one data-generating command (titration, measurement etc.) before the calculation command, then you will have to indicate the command identification in front of the variable: Variable Description #M. Titration or measuring command #C.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters ■ Common variables List of common variables. All of the common variables are listed, no matter whether they possess a value or not. You will find a detailed description of all of the variables in chapter 28.4.3.3, page 296. 28.4.3.
■■■■■■■■■■■■■■■■■■■■■■ 28.4 Calculations 28.4.3.3 Variable list The following table contains all of the variables which can be used for calculations. For variables having an index (e.g. 'EP1'), the index must be entered manually. In the following table, this index is characterized with the character "#". Variable Description C00 Sample size CI1, CI2 Sample identifications The sample identifications can be used in calculations only if numerical values are entered.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Variable Description MDC Drift for drift correction DDC Time for drift correction MCV End volume, i.e. total dosed volume at the end of the titration MCQ End quantity, i.e. total amount of removed water or of generated bromine, respectively, at the end of the titration (in µg) MCC Final charge, i.e.
■■■■■■■■■■■■■■■■■■■■■■ 28.
■■■■■■■■■■■■■■■■■■■■■■ 28.5 28 Parameters Reports (REPORT) The reports to be printed out are defined with the command REPORT. A maximum of ten reports can be defined for each command. The following reports can be inserted, among others: ■ ■ ■ ■ ■ Result report Calculation report Parameter report System reports (system settings, sensor list, GLP data etc.) Method reports (method run, method parameters etc.) [Report options] Defining the general settings for the report output.
■■■■■■■■■■■■■■■■■■■■■■ 28.5 Reports (REPORT) NOTE You can also define your own report header, which is printed in addition to this report header (System ▶ Templates ▶ Report header). Selection Default value off | once | on each page on each page off The report header will not be printed. once The report header will only be printed on the first page. on each page The report header will be printed on every page. Signature line Output of a special line for date and signature.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters ■ ■ ■ ■ ■ ■ ■ ■ ■ 28.5.3 Measuring point list Definition of the method command to which the measuring point list is to be printed. Calculations Definition of the accuracy with which the variables used as well as the results calculated are printed. Statistics short Definition of the time at which the report is to be printed (with each determination or only at the end of a sample series).
■■■■■■■■■■■■■■■■■■■■■■ 28.5 Reports (REPORT) Report Contents Statistics short Summary of the statistics calculations. The number of determinations, the mean value, and the absolute and relative standard deviations are printed out for each result. Statistics overview Detailed statistical overview. The sample data and all individual results are printed out for each determination.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Report Contents All titrant data full All of the titrant data of all of the titrants (working life, titer options, data concerning the exchange unit/dosing unit). Sensors Sensor list List of all the sensors configured in the system. All sensor data short The most important sensor data for all sensors. All sensor data full All of the sensor data for all of the sensors (name, working life, etc.).
■■■■■■■■■■■■■■■■■■■■■■ 28.6 Dosing and Liquid Handling 28.6 Dosing and Liquid Handling The following dosing commands are available: 28.6.1 Preparing the exchange unit or dosing unit (PREP) Rinsing the cylinder and the tubings of the exchange unit / dosing unit. Emptying the dosing unit (EMPTY) Emptying the cylinder and the tubings of the dosing unit. Dosing a fixed volume (ADD) Dosing a specified volume. Liquid Handling (LQH) Carrying out complex dosing tasks with a Dosino.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters the type of dosing drive matches. For buret units without integrated data chip, the cylinder volume and the type of dosing drive are checked. For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination. Selection Default value Selection of configured titrants | not defined not defined not defined No check takes place.
■■■■■■■■■■■■■■■■■■■■■■ 28.6 Dosing and Liquid Handling which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list. For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without integrated data chip, the cylinder volume and the type of dosing drive are checked.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Filling rate Rate at which the dosing cylinder is filled. Input range Selection Default value 0.01 - 166.00 mL/min maximum maximum maximum The maximum filling rate depends on the cylinder volume (see Chapter 31, page 338). 28.6.3.2 Control device The control device the determination is carried out with is selected under [Control device]. Control devices are defined under System ▶ Device manager.
■■■■■■■■■■■■■■■■■■■■■■ 28.6 Dosing and Liquid Handling For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without integrated data chip, the cylinder volume and the type of dosing drive are checked.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Selection Default value Selection of configured titrants | not defined not defined not defined No check takes place. Filling rate Rate at which the dosing cylinder is filled. Input range Selection Default value 0.01 - 166.00 mL/min maximum maximum maximum The maximum filling rate depends on the cylinder volume (see Chapter 31, page 338).
■■■■■■■■■■■■■■■■■■■■■■ 28.6 Dosing and Liquid Handling Table 7 Maximum dosing rate for different dosing cylinders Cylinder volume 28.6.3.4 maximum dosing rate Exchange unit Dosing unit 1 mL 2.25 mL/min – 2 mL – 5.00 mL/min 5 mL 11.25 mL/min 12.50 mL/min 10 mL 22.50 mL/min 25.00 mL/min 20 mL 45.00 mL/min 50.00 mL/min 50 mL 112.50 mL/min 124.50 mL/min Stirrer The parameters for the stirrer are edited under [Stirrer].
■■■■■■■■■■■■■■■■■■■■■■ 28.6.4 28 Parameters Liquid Handling (LQH) The command LQH can be used to carry out Liquid Handling tasks with a dosing device of the type Dosino 800. The manifold options of the Dosino can be exploited in full in combination with a USB Sample Processor. The four ports of the dosing units can be used at will as output or input ports. This means that not only simple dosing and filling procedures are possible.
■■■■■■■■■■■■■■■■■■■■■■ 28.6 Dosing and Liquid Handling Compensate Because of the fact that the dosing units are interchangeable, the coupling of the Dosino piston rod (spindle) exhibits a low mechanical tolerance that can be noticed when the piston changes its direction of movement. This tolerance can be compensated with this function. A short piston movement is first made in the same direction as the previous movement, which is then followed by a piston movement in the reverse direction.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Selection Default value maximum maximum maximum The maximum dosing and filling rates depend on the cylinder volume (see Chapter 31, page 338). 28.6.4.2 Control device The control device the determination is carried out with is selected under [Control device]. Control devices are defined under System ▶ Device manager. NOTE This button will only be shown when, in addition to the Ti-Touch, a Sample Processor has been configured as control device.
■■■■■■■■■■■■■■■■■■■■■■ 28.7 Communication the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination. Selection Default value Selection of configured titrants | not defined not defined not defined No check takes place. 28.7 Communication The following communications commands are available: 28.7.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Input signal Selection of the signal out of the templates or entering the required bit pattern. Templates are defined under System ▶ Templates ▶ Input lines. Entering a bit pattern: ■ ■ ■ 0 = line inactive 1 = line active * = retain line status The input lines are always numbered from right to left, i.e. with the signal *******1 line 0 is expected to be active.
■■■■■■■■■■■■■■■■■■■■■■ 28.7 Communication Document message In the determination data it will be documented, that the time interval has been expired. Cancel determination The determination is stopped. 28.7.2 Setting remote lines (CTRL) The command CTRL can be used to define output signals at the remote interface. Control device This parameter will only be shown when, in addition to the Ti-Touch, a Sample Processor has been configured as control device.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters If it is required to set pulses with other lengths, you have to define a corresponding template. NOTE We recommend masking lines that are of no interest or for which no defined condition can be predicted with an asterisk (*). Entry Default value Selection 28.7.3 Bit patterns containing exactly 14 characters or a max.
■■■■■■■■■■■■■■■■■■■■■■ 28.7 Communication Selection Default value Display message | Document message | Cancel determination Display message For all three options it is documented in the determination data (see dialog More determination data / Messages), that the time interval has been expired. Display message A message is displayed. You can select whether you want to continue with the determination or cancel the run.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters 28.8 Automation 28.8.1 Rotating sample rack (MOVE) The command MOVE is used to move to a rack position or an external position. Tower Selection of the tower, with which the command is carried out. Both Tower 1 and Tower 2 can always be selected, even if your Sample Processor has only one tower. Selection Default value 1|2 1 1 View from the front, the right tower. 2 View from the front, the left tower. Destination Selection of the required target position.
■■■■■■■■■■■■■■■■■■■■■■ 28.8 Automation Current sample + Starting from the current sample (defined by the sample variable) the rack is moved forward by the number of rack positions (1…999) entered in the field to the right. Current sample Starting from the current sample (defined by the sample variable) the rack is moved backward by the number of rack positions (1…999) entered in the field to the right. Next position Starting from the current rack position, the rack is moved forward by one position. Prev.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Rotate rack The rack is moved to the next vessel on a sample position. Exception: If the sample variable is newly defined in the method before this command (command SAMPLE), then the option Display message is automatically valid. Display message A message is displayed. You can select whether you want to continue with the determination or cancel the run. Options Shift rate Rate at which the sample rack is moved.
■■■■■■■■■■■■■■■■■■■■■■ 28.8 Automation Selection Default value 1|2 1 1 View from the front, the right tower. 2 View from the front, the left tower. Lift position Selecting a predefined lift position or enter any lift position. The predefined lift positions are defined in the device properties of the Sample Processor. They can however also be assigned in the manual control.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Selection Default value 1|2|1+2 1 1 Pump 1 of the selected tower is switched. 2 Pump 2 of the selected tower is switched. 1+2 Both pumps of the selected tower are switched at the same time. Status/Duration Switching the pump(s) on and off. Input range Selection Default value 0 - 999 s (Increment: 1) The pump is running during this time. on | off on on The pump is switched on. The pump is running until it is explicitly switched off.
■■■■■■■■■■■■■■■■■■■■■■ 28.8 Automation The sample variable can be altered specifically with the command SAMPLE. Sample variable Modifying the sample variable. Selection Default value =|+|+ = The sample variable corresponds to the number entered in the field Value. + The sample variable corresponds to the current value plus the number entered in the field Value. The sample variable corresponds to the current value minus the number entered in the field Value.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters NOTE Not all commands are available for subsequences. Commands which cannot be inserted into subsequences are disabled. [Delete command] Delete the selected method command. [Edit command] Edit the selected method command. 28.8.6.1 Loading / saving a subsequence [Load/ Save] is used to load stored subsequences or to save new subsequences. This means that they are available for all methods and only have to be created once. Loading a subsequence [Own subseq.
■■■■■■■■■■■■■■■■■■■■■■ 28.8 Automation NOTE In order to be able to use the subsequences also for other titration systems, you have to create a backup. You can only restore the subsequences from this backup on another system. File name File name of the subsequence. Entry max. 32 characters [Save] Saving the subsequence to the internal memory. 28.8.6.2 Subsequence options In this dialog, the settings can be defined which correspond to the entire subsequence.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Input range Default value Selection 0 - 999 1 Calibration Calibration For automatic calibrations with a USB Sample Processor (relevant only if the Ti-Touch supports calibrations). [Note] Entering a note on the subsequence. Dialog "Subsequence options / Note" A short text can be entered in this dialog, e.g. for the description of the subsequence commands or for the application purpose. [Display options] Definition when the note is being displayed.
■■■■■■■■■■■■■■■■■■■■■■ 28.9 Miscellaneous commands Input range Selection Default value 0 - 999 s (Increment: 1) During this time the stirrer is running. on | off on on The stirrer is switched on. The stirrer runs until it is explicitly switched off. If the determination is canceled with the key [ ], the stirrer is also switched off. off The stirrer is switched off. Stirring rate Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm.
■■■■■■■■■■■■■■■■■■■■■■ 28 Parameters Message text This parameter can only be edited when Message is activated. Text which is displayed during the waiting time. Entry Default value 28.9.3 max.
■■■■■■■■■■■■■■■■■■■■■■ 28.9 Miscellaneous commands Hold sequence on | off (Default value: on) If this parameter is activated, then the run will be paused during the request. If this parameter is deactivated, then the method continues in the background until the following measurement or titration has been completed. 28.9.4 Defining an acoustic signal (BEEP) The BEEP command can be used to produce an acoustic signal. Duration Approximate playing time in seconds of the permanently stored melody.
■■■■■■■■■■■■■■■■■■■■■■ 29 Operation and maintenance 29 Operation and maintenance 29.1 System initialization In very rare instances, a faulty file system (e.g. because of a program crash) may lead to an impairment of program functioning. The internal file system must be initialized in such cases. CAUTION If you carry out a system initialization, then all user data except the methods will be deleted. We recommend creating a backup of the system at regular intervals in order to avoid data losses.
29.2 Quality management and qualification with Metrohm 29.2 ■■■■■■■■■■■■■■■■■■■■■■ Quality management and qualification with Metrohm Quality management Metrohm offers you comprehensive support in implementing quality management measures for instruments and software. Qualification Please contact your local Metrohm representative for support in qualification of instruments and software.
■■■■■■■■■■■■■■■■■■■■■■ 30 Troubleshooting 30 Troubleshooting 30.1 Editing methods Problem Cause Remedy The method command cannot be inserted. The method command is disabled. In the dialog Dialog options / Command list, activate the desired command (see "Blocking unneeded commands and fixed keys", page 34). No control device which supports the method command is configured in the device manager.
■■■■■■■■■■■■■■■■■■■■■■ 30.4 Printing Problem Cause Remedy No statistics variable has been assigned to the calculation. In the result options (command CALC), define a variable for the mean value (see "Variable for mean value", page 145). Statistics is not activated. In the control dialog, activate the option Statistics (see Chapter 17, page 170). A result has been removed from the statistics, the mean value has not been reassigned to the common variable or to the TITER variable.
■■■■■■■■■■■■■■■■■■■■■■ 30 Troubleshooting Problem Cause Remedy A button in the manual control is disabled. The required hardware is either not connected or incorrectly connected. 1. Switch off the 915 KF Ti-Touch. 2. Connect the hardware correctly. 3. Switch the 915 KF Ti-Touch back on. The required hardware is being used in the ongoing determination. Wait until the determination is finished. 30.
■■■■■■■■■■■■■■■■■■■■■■ 30.7 Karl Fischer titration Problem Cause Remedy A side reaction is taking place. ■ ■ ■ The titration will not be finished. Use special reagents. Adjust the method (increase/decrease the temperature, external extraction). See technical literature. The pH value is not in the optimum range anymore. Add buffer, see technical literature. The titration cell is leaking. ■ ■ Check seals and septum. If necessary, replace. Replace the molecular sieve.
■■■■■■■■■■■■■■■■■■■■■■ Problem 30 Troubleshooting Cause Remedy The electrode has a short circuit. 1. Check the Pt wires. 2. Activate the electrode check. The endpoint is reached too quickly. The dosing rate outside the control range is too high. Define Titration rate = user and reduce the dosing rate (Max. rate) (see Chapter 28.1.2, page 261). The titration times with volumetric titration are constantly longer. With two-component reagents the buffer capacity of the solvent can be exhausted.
■■■■■■■■■■■■■■■■■■■■■■ 31.1 Dosing unit 31 Appendix 31.1 Dosing unit 31.1.1 Maximum dosing and filling rate The maximum dosing rate and maximum filling rate for the dosing unit depend on the cylinder volume: Cylinder volume Maximum rate 2 mL 6.67 mL/min 5 mL 16.67 mL/min 10 mL 33.33 mL/min 20 mL 66.67 mL/min 50 mL 166.00 mL/min Independent of the cylinder volume, values ranging from 0.01 to 166.00 mL/min can always be entered.
■■■■■■■■■■■■■■■■■■■■■■ 31 Appendix 31.2 Exchange unit 31.2.1 Maximum dosing and filling rate The maximum dosing rate and maximum filling rate for the exchange unit depend on the cylinder volume: Cylinder volume Maximum rate 1 mL 3.00 mL/min 5 mL 15.00 mL/min 10 mL 30.00 mL/min 20 mL 60.00 mL/min 50 mL 150.00 mL/min Independent of the cylinder volume, values ranging from 0.01 to 166.00 mL/min can always be entered.
■■■■■■■■■■■■■■■■■■■■■■ 31.4 Balance r/min 2000 1500 1000 500 -15 -10 Figure 16 31.4 -5 0 5 10 15 Rotational speed depending on stirring rate Balance The sample size and the associated unit can be sent from a connected balance. The sample size is transmitted as a number with up to ten characters (including algebraic sign and decimal point). Sample size and unit are sent as a single character string. They are separated by a space character.
■■■■■■■■■■■■■■■■■■■■■■ 31.5 31 Appendix Remote interface The 6.2148.010 Remote Box allows devices to be controlled which cannot be connected directly to the MSB interface of the Ti-Touch. 1 2 3 Figure 17 Connectors of the Remote Box 1 Cable For connecting to the Ti-Touch. 3 Remote connector For connecting instruments with a remote interface. 31.5.1 2 MSB connector Metrohm Serial Bus. For connecting external dosing devices or stirrers.
■■■■■■■■■■■■■■■■■■■■■■ 31.5 Remote interface Outputs Open Collector tp tp > 200 ms active = low, inactive = high IC = 20 mA, VCEO = 40 V +5 V: maximum load = 20 mA The following tables offer information concerning the assignment of the individual pins and their function. Table 8 342 ■■■■■■■■ Inputs and outputs of the remote interface Assignment Pin No.
■■■■■■■■■■■■■■■■■■■■■■ 31 Appendix Assignment Pin No. Output 13 20 0 volts / GND 14 +5 volts 15 0 volts / GND 25 Table 9 Function Explanation of the individual functions Function Explanation Start The current method is started at the time of activation. tpulse > 100 ms Stop The current method is canceled (Stop) at the time of activation. tpulse > 100 ms Quit The current command in the determination run will be canceled at the time of activation.
■■■■■■■■■■■■■■■■■■■■■■ 31.6 Result variables as parameter setting 31.6 Result variables as parameter setting For most of the numerical parameters, the result of a calculation can also be utilized in place of a number. This result must be calculated in a preceding calculation command (command CALC). A typical application is the use of a relative start volume. The following example is used to explain how this method is used to define a sample size-dependent start volume for a titration.
■■■■■■■■■■■■■■■■■■■■■■ 31 Appendix The selection of the result variables is displayed in the number editor. NOTE If the method before the command with the result variables applied contains more than one calculation commands (possibly with result variables bearing the same name), then the result variable of the preceding calculation command will always be used.
■■■■■■■■■■■■■■■■■■■■■■ 31.7 Diagnosis Touch adjustment Adjust the touch-sensitive screen (see Chapter 31.7.4, page 347). Touch screen test Test the function of the touch-sensitive screen (see Chapter 31.7.5, page 348). Software update Update the software of Ti-Touch and of the control devices (see Chapter 31.7.6, page 349). Service Special functions for the service technician (see Chapter 31.7.7, page 352). ■ ■ ■ ■ 31.7.
■■■■■■■■■■■■■■■■■■■■■■ 31.7.2 31 Appendix Formatting an external storage medium System ▶ Diagnosis ▶ Format storage medium With this function, you have the option of formatting an external storage medium directly on the 915 KF Ti-Touch (quick formatting). The following file systems can be selected: ■ ■ ■ FAT FAT32 ExFAT CAUTION If you format the storage medium with the ExFAT file system, it is possible that it will no longer be recognized by a PC.
■■■■■■■■■■■■■■■■■■■■■■ 31.7 Diagnosis 4 Touch a stylus (a special pen-shaped instrument for operating devices with touch screens) to the center of the cross hair and maintain this position until the cross hair moves to a new position on the screen. 5 Repeat this process as long as the cross hair continues to be repositioned on the screen. A message in English will appear. 6 Tap at any location on the screen to confirm the adjustment. 7 Press [ 31.7.5 ] to finish the adjustment.
■■■■■■■■■■■■■■■■■■■■■■ 31 Appendix 3 Repeat step 2 until all of the areas in the display have been tested. A list with fixed keys will appear. 4 Tap on all of the fixed keys one after the other in any sequence. The result is displayed in each case with a short information text, e.g. [Print] key OK. 5 Finish Touch screen test with [ 31.7.6 ].
■■■■■■■■■■■■■■■■■■■■■■ 31.7 Diagnosis Program files The files are instrument-specific. The file names are structured as follows: ■ ■ Control software for 915 KF Ti-Touch: 5XXXyyyy.BIN 5XXXyyyy.MBIN – XXX = instrument type (i.e. "915" for the 915 KF Ti-Touch) – yyyy = program version Firmware of the control device: 5XXXyyyy.BIN – XXX = instrument type (e.g.
■■■■■■■■■■■■■■■■■■■■■■ 31.7.6.1 31 Appendix Carrying out a software update CAUTION Make sure that the power supply is guaranteed during the entire update process. Otherwise there is the chance that the instrument will no longer be able to be switched on and that it will need to be sent in for repair.
■■■■■■■■■■■■■■■■■■■■■■ 31.7 Diagnosis 3 Selecting the instrument ■ Tap on the Control device list box and select the instrument to be updated. 4 Selecting the file ■ ■ ■ ■ Tap on the Binary file list box. The selection list with the BIN files saved in the "SwUpdates" directory is opened. Select the required file. Tap on [Select]. Update of the Ti-Touch: Continue with Step 5. Update of the Sample Processor: Continue with Step 6.
■■■■■■■■■■■■■■■■■■■■■■ 31.8 31 Appendix Arithmetic algorithms in the Ti-Touch Some of the algorithms and numerical formats used in the device software are explained in this chapter. Numerical format All floating point calculations use the IEEE 754 (1985) numerical format, either in "single precision" (32 Bit) or "double precision" (64 Bit).
■■■■■■■■■■■■■■■■■■■■■■ 31.8 Arithmetic algorithms in the Ti-Touch Explanations The individual values are incorporated in the statistics in the selected Precision (Round, Truncate or Full precision). 7 significant places are yielded when the 32 bit numerical format is applied for the floating point number in decimal presentation. For the 64 bit numerical format, there are 15 significant places.
■■■■■■■■■■■■■■■■■■■■■■ 32 Technical specifications 32 Technical specifications 32.1 Touch screen Display VGA color display (640 pixels x 320 pixels), diagonal ca. 5.7" Touch panel resistive Resistance to chemicals Resistant to the following chemicals (no visible changes after 24 h of duration of action): ■ without protective foil: – Methanol – Toluene – Ethyl acetate – Acetone – Dichloromethane – Hydrochloric acid c(HCl) = 2 mol/L ■ with protective foil (order number 6.2723.
■■■■■■■■■■■■■■■■■■■■■■ 32.2 Measuring inputs 32.2 Measuring inputs The measuring cycle is 100 ms for all measuring modes. 32.2.1 Polarizer A measuring input (Pol.) for polarizable electrodes. Measuring mode Ipol Determination with adjustable polarization current. Polarization current –120 - +120 µA (increment: 1 µA) –125 - –121 µA / +121 - +125 µA: non-guaranteed values, dependent on reference voltage +2.5 V Measuring range –1,200 - +1,200 mV Resolution 0.1 mV Measuring accuracy ±0.
■■■■■■■■■■■■■■■■■■■■■■ 32 Technical specifications Measuring accuracy Pt1000 ±0.2 °C (applies for measuring range –20 - +150 °C) NTC ±0.6 °C (applies for measuring range +10 - +40 °C) 32.3 Internal stirrer Rotational speed Maximum rotational speed ±(1,700 - 1,900) rpm Setting of the rotational speed ±15 steps Change in rotational speed per step ±(115 - 125) rpm Power consumption 4W Fuse Electronic overload protection 32.
■■■■■■■■■■■■■■■■■■■■■■ 32.6 Power supply 32.6 Power supply External power supply unit Input voltage range Power consumption 32.
■■■■■■■■■■■■■■■■■■■■■■ 32.9 32 Technical specifications Ambient temperature Nominal function range +5 - +45 °C (at a maximum of 85% humidity) Storage –20 - +60 °C Transport –40 - +60 °C 32.10 Reference conditions Ambient temperature +25 °C (± 3 °C) Relative humidity ≤ 60% Operating temperature status Instrument in operation at least 30 min Validity of the data After adjustment 32.
■■■■■■■■■■■■■■■■■■■■■■ 33 Warranty (guarantee) Metrohm guarantees that the deliveries and services it provides are free of defects in materials, design or manufacturing. The general warranty period is 36 months (exclusions below) from the date of delivery, or 18 months in the event of continuous operation. The warranty remains valid on the condition that the servicing is provided by a service organization authorized by Metrohm at defined intervals and with a defined scope.
■■■■■■■■■■■■■■■■■■■■■■ 33 Warranty (guarantee) The same warranty periods that are specified for a corresponding new part apply to parts that are replaced or repaired within the above-mentioned warranty periods. However, replacement or repair of a part does not extend the warranty period of the entire system. Deficiencies arising from circumstances that are not the responsibility of Metrohm, such as improper storage or improper use, etc., are expressly excluded from the warranty.
■■■■■■■■■■■■■■■■■■■■■■ 34 Accessories Up-to-date information on the scope of delivery and optional accessories for your instrument can be found on the Internet. NOTE When you receive your new instrument, we recommend downloading the accessories list from the Internet, printing it out and keeping it together with the manual for reference purposes.
■■■■■■■■■■■■■■■■■■■■■■ 34 Accessories The Partslists webpage will be displayed. 7 Select the desired output language. 8 With the article number entered, click on the command Generate PDF. The PDF file with the accessories data will be created in the language selected. Direct access for all instruments If you are unable to find your instrument using the search as described above, this may be due to the instrument not being sold anymore.
■■■■■■■■■■■■■■■■■■■■■■ Index Index Numbers/Symbols 800 Dosino .............................. 13 801 Stirrer ................................ 14 803 Ti Stand ............................. 14 804 Ti Stand ............................. 14 805 Dosimat ............................ 13 A Acoustic signals ........................ 45 ADD ....................................... 306 Automatic system test ............ 122 Automation ............................ 319 Axial distance ...........................
■■■■■■■■■■■■■■■■■■■■■■ Dosing ................................... 237 Continuous ....................... 239 Fixed volume .................... 241 Dosing curve .......................... 210 Dosing device Parameters ....... 270, 307, 313 Dosing devices Connect ............................. 13 Dosing parameters ADD ................................. 306 Dosing port .............................. 52 Dosing rate Maximum ................. 338, 339 Dosing unit Data ...................................
■■■■■■■■■■■■■■■■■■■■■■ Index Message Determination .................. 202 Send e-mail ........................ 77 Method .................................. 154 Command list ................... 157 Create .............................. 154 Edit .................................. 157 Insert command ............... 158 Loading ............................ 156 Note ................................. 167 Options ............................ 159 Properties ......................... 167 Sample data ...............
■■■■■■■■■■■■■■■■■■■■■■ Save ................................. 224 Result variable ........................ 288 As parameter setting ........ 344 Robotic arm ............................. 91 Max. swing angle ............... 91 Rotation offset ................... 92 Swing direction .................. 92 Swing increment ................ 90 Swing offset ....................... 91 Swing radius ....................... 91 Swing rate ........ 254, 256, 321 Rotation increment ................... 88 Rotation offset .
■■■■■■■■■■■■■■■■■■■■■■ Index Valve disc Shift direction ..................... 56 W WAIT ...................................... 328 368 ■■■■■■■■ Warning limit Titer ................................... 65 Warranty ................................ 360 Water determination according to Karl Fischer ............................... 23 Working life Sensor ................................ 72 Titrant ................................ 50 Write protection External memory ..............