Instruction Manual
■■■■■■■■■■■■■■■■■■■■■■
11 Device manager
915 KF Ti-Touch
■■■■■■■■
101
2
Carry out the adjustment
■ Tap on the button [Start] (NOT on the fixed key [ ]).
The rack is being initialized. The rack then moves to position 1
and the lift is lowered to the work position.
■ If necessary, the lift position can be corrected with the arrow keys
[⇧] and [⇩].
■ Now use the arrow keys [⇦] and [⇨] to rotate the rack so that the
robotic arm or the center of the titration head is positioned
exactly above the center of rack position 1.
■ Finish the adjustment with [OK].
The lift is moved to the uppermost stop.
11.7 Printer
Instrument list: Printer ▶ Edit