Instruction Manual

■■■■■■■■■■■■■■■■■■■■■■
27 Manual control
915 KF Ti-Touch
■■■■■■■■
235
27.2 Measuring
Manual control Measure
Manual measurements can be carried out with the function [Measure].
Proceed as follows:
1
Select a sensor
Select the desired sensor from the sensor list.
All of the sensors available in the sensor list are displayed. Sensors
are defined under System Sensors.
Select the measuring input at which the sensor is connected.
2
Select a measuring mode
Select the desired measuring mode.
Only those measuring modes are displayed which are advisable
for the selected sensor.
3
Enter a measuring temperature
Enter the measuring temperature if no temperature sensor is con-
nected. If a temperature sensor is connected, then the tempera-
ture will be measured automatically.
4
Start the measurement
Tap on [Start].
The measured value display is opened: