27.4 Stirring
■■■■■■■■■■■■■■■■■■■■■■
246
■■■■■■■■
915 KF Ti-Touch
27.3.6 Replacing reagent
NOTE
You can easily change the reagent in a dosing unit without no contact
with the chemicals using the functions Prepare and Empty.
Proceed as follows:
1
Execute the function [Empty].
2
Attach the dosing unit to the bottle with a suitable solvent.
3
Execute the function [Prepare].
The last traces of the reagent in the cylinder and tubings are
removed with the solvent.
4
Execute the function [Empty] again.
5
Attach the dosing unit to the bottle with the new reagent.
6
Execute the function [Prepare].
The dosing unit can now be used.
27.4 Stirring
Manual control Stir