Instruction Manual

■■■■■■■■■■■■■■■■■■■■■■
27 Manual control
915 KF Ti-Touch
■■■■■■■■
249
[Set]
Set a defined output signal.
27.6 USB Sample Processor
Manual control Sample Processor
With the function [Sample Processor], you can control a connected
Sample Processor manually. The following functions are available:
Moving a sample rack
Moving the lift
Defining specific lift positions (work position, rinse position, etc.)
Switching pumps on and off
Initializing the rack
Defining external positions outside of the sample rack (only with Swing
Head)
Tower for the manual control.
Selection Tower 1 | Tower 2
Tower 1
View from the front, the right tower.
Tower 2
View from the front, the left tower. This selection is only available for
Sample Processors with two towers.
Lift position
Current lift position.
Rack position
Current rack position. If the rack is not set to a valid position, e.g. after a
rack reset, -- is displayed to indicate this.