Instruction Manual

27.6 USB Sample Processor
■■■■■■■■■■■■■■■■■■■■■■
254
■■■■■■■■
915 KF Ti-Touch
Input range 1 - Number of positions on the rack
attached.
Default value 1
Selection Special beaker 1…16
Special beaker 1…16
The selection depends on how many rack positions are defined as spe-
cial beakers (see "Editing special beakers", page 98).
Shift rate
Rate at which the sample rack is moved.
Input range 3 - 20 °/s
Default value 20 °/s
Shift direction
Direction in which the rack is moved.
Selection auto | + | -
Default value auto
auto
A shift direction, with which the shorter way has to be passed, is auto-
matically selected.
+
Counterclockwise rotation.
-
Clockwise rotation.
Swing rate
Rate at which the robotic arm is swung in the manual control.
Input range 10 - 55 °/s
Default value 55 °/s
27.6.3 External positions
Manual control: Sample Processor External position