Instruction Manual
28.3 Evaluations (EVAL)
■■■■■■■■■■■■■■■■■■■■■■
278
■■■■■■■■
915 KF Ti-Touch
Fixed quantity
Selection of the quantity to which the associated value is interpolated
from the measuring point list.
Selection Measured value | Time | Volume
Default value Measured value
Fixed EP1 at
The fixed endpoint must lie between the first and the final entry in the
measuring point list.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 355.
Measured value (measuring mode Ipol):
Input range –2,000.0 - 2,000.0 mV
Selection off
Default value off
Measured value (measuring mode Upol):
Input range
–200.00 - 200.00 µA
Selection off
Default value off
Measured value (measuring mode T):
Input range
–20.0 - 150.0 °C
Selection off
Default value off
Time:
Input range
0.0 - 999,999.9 s
Selection off
Default value off
Volume:
Input range
0.00000 - 9,999.99 mL
Selection off
Default value off
Fixed EP2 at…Fixed EP9 at
See Fixed EP1 at.