Instruction Manual
■■■■■■■■■■■■■■■■■■■■■■
28 Parameters
915 KF Ti-Touch
■■■■■■■■
283
[Edit]
Edit the settings of the selected window.
Lower limit
Time for the lower limit.
Input range 0 - 999999 s
Default value 0 s
Upper limit
Time for the upper limit.
Input range 0 - 999999 s
Default value 999999 s
28.4 Calculations
The following calculation commands are available:
CALC Defining calculations of a determination.
CALC LIVE Defining the calculation whose result will be dis-
played in the live display (so-called live result)
during a titration or measurement.
28.4.1 Calculations (CALC)
Calculations are defined with the command CALC. A method may contain
a maximum of nine calculation commands. A maximum of nine calcula-
tions can be defined per command. A series of variables (raw data from
the determination, previously calculated results, etc.) is available for the
calculations.
The command contains the following options, among others:
■ Save result as titer
■ Save result as common variable
■ Save result in result table
■ Define result limits