Instruction Manual

■■■■■■■■■■■■■■■■■■■■■■
28 Parameters
915 KF Ti-Touch
■■■■■■■■
321
Rotate rack
The rack is moved to the next vessel on a sample position. Exception:
If the sample variable is newly defined in the method before this com-
mand (command SAMPLE), then the option Display message is
automatically valid.
Display message
A message is displayed. You can select whether you want to continue
with the determination or cancel the run.
Options
Shift rate
Rate at which the sample rack is moved.
Input range 3 - 20 °/s
Default value 20 °/s
Shift direction
Direction in which the rack is moved.
Selection auto | + | -
Default value auto
auto
A shift direction, with which the shorter way has to be passed, is auto-
matically selected.
+
Counterclockwise rotation.
-
Clockwise rotation.
Swing rate
Rate at which the robotic arm is swung.
Input range 10 - 55 °/s
Default value 55 °/s
28.8.2 Moving the lift (LIFT)
The lift of a Sample Processor is moved with the command LIFT. The
movement can however only be carried out if the Sample Processor is
located at a valid rack position. This will not be the case, for example,
after a Rack Reset.
Tower
Selection of the tower, with which the command is carried out. Both
Tower 1 and Tower 2 can always be selected, even if your Sample Pro-
cessor has only one tower.