Instruction Manual
31.7 Diagnosis
■■■■■■■■■■■■■■■■■■■■■■
348
■■■■■■■■
915 KF Ti-Touch
4
Touch a stylus (a special pen-shaped instrument for operating devices
with touch screens) to the center of the cross hair and maintain this
position until the cross hair moves to a new position on the screen.
5
Repeat this process as long as the cross hair continues to be reposi-
tioned on the screen.
A message in English will appear.
6
Tap at any location on the screen to confirm the adjustment.
7
Press [ ] to finish the adjustment.
31.7.5 Testing the touch screen
System ▶ Diagnosis ▶ Touch screen test
You can use this test to check the function of the touch-sensitive monitor
screen.
A calibration is not possible with this test. If the Touch screen does not
work correctly anymore, it will need to be replaced. For that please con-
tact your Metrohm representative.
NOTE
You can use the [ ] fixed key to stop the test at any time.
Proceed as follows:
1
Tap on [Touch screen test].
A red square is displayed at the upper left.
2
Tap on the square as close to the center as possible.
The actual detected point of contact will be displayed by a cross hair,
and the deviation from the center will also be displayed on the lower
edge of the image.
The next red square will also be displayed.