Instruction Manual

■■■■■■■■■■■■■■■■■■■■■■
32 Technical specifications
915 KF Ti-Touch
■■■■■■■■
357
Measuring accu-
racy
Pt1000 ±0.2 °C
(applies for measuring range –20 - +150 °C)
NTC ±0.6 °C
(applies for measuring range +10 - +40 °C)
32.3 Internal stirrer
Rotational speed
Maximum rota-
tional speed
±(1,700 - 1,900) rpm
Setting of the
rotational
speed
±15 steps
Change in rota-
tional speed per
step
±(115 - 125) rpm
Power consump-
tion
4 W
Fuse Electronic overload protection
32.4 Pump
Add
> 600 mL/min (at 25 °C; standard accessories)
Aspirate > 400 mL/min (at 25 °C; standard accessories)
32.5 Interfaces
USB connector
Type A, for connecting USB devices.
MSB connector For connecting dosing devices, stirrers or a Remote Box.
Ethernet connec-
tor
For connection to a data network (LAN).