Instruction Manual

■■■■■■■■■■■■■■■■■■■■■■
7 System settings
915 KF Ti-Touch
■■■■■■■■
35
Blocking commands
Proceed as follows to block unneeded commands:
1
Display the command list
Tap on the button [Command list].
The list of all command groups is displayed.
2
Deactivate command groups
Deactivate those command groups which are not permitted to be
used.
3
Save the settings
Tap on the fixed keys [ ] or [ ].
All deactivated commands appear grayed out in the method editor
and cannot be used for creating methods.
Blocking fixed keys
Proceed as follows to block unneeded fixed keys:
1
Display fixed keys which can be blocked
Tap on the button [Fixed keys].
All fixed keys which can be blocked are displayed.
2
Deactivate fixed keys
Deactivate those fixed keys which are not permitted to be used.
3
Save the settings
Tap on the fixed keys [ ] or [ ].
Deactivated fixed keys cannot be used.
Configuring the routine dialog
A suitable Standard configuration has already been saved for routine
operations.
Methods can only be loaded, but not modified or created.
Determinations cannot be recalculated.
You can readjust this standard configuration by disabling additional func-
tions or re-enabling disabled functions.