User guide

3.1 Equipping the eluent bottle
■■■■■■■■■■■■■■■■■■■■■■
4
■■■■■■■■
IC equipment: Additional eluent for 941
3 Installation
Proceed as follows to install the IC equipment: Additional eluent for 941:
Equip the eluent bottle.
Equip the concentrate bottle.
Equip the ultrapure water container.
Assemble the 800 Dosino and connect all tubings.
Plug level sensor cable into the 941 Eluent Production Module.
3.1 Equipping the eluent bottle
For the automated eluent production, the eluent bottle has to be
equipped with a bottle cap for level sensor and a sensor. The sensor
choice depends on which fill level is to be monitored: bottle full (FULL) or
bottle empty (EMPTY).
Inserting the measuring sensor
Required accessories
A bottle cap for level sensor (6.1626.000)
FULL sensor (6.2769.000) for anion eluents or EMPTY sensor
(6.2769.1X0) for cation eluents
Figure 2 Bottle cap for level sensor – Openings
1
M6 opening
2
M6 opening
3
M8 opening
4
Opening for level sensor cable
5
Opening for level sensor cable
1
Remove the stoppers from the bottle cap.
2
Place the bottle cap on a surface with the opening facing up.