User guide
3.1 Equipping the eluent bottle
■■■■■■■■■■■■■■■■■■■■■■
6
■■■■■■■■
IC equipment: Additional eluent for 941
1
Mounting the adsorber tube
Fill the adsorber tube with adsorber material and cotton and insert it
into the M6 opening (2-1).
2
Connecting the FEP tubing
Fasten the tubing connection to the Dosino to the M6 opening (2-2).
■ Insert the FEP aspiration tubing into the M6 opening.
■ Screw the FEP tubing into the M6 opening.
3
Connecting the eluent aspiration tubing
Fasten the capillary connection to the ion chromatograph to the M8
opening (2-3).
■ Start by pushing the M8 tubing nipple over the end of the aspira-
tion tubing, followed by the O-ring.
■ Guide the eluent aspiration tubing through the M8 opening of
the bottle cap and temporarily secure it in place with the tubing
nipple.
■ Attach the tubing weighting and the aspiration filter to the end of
the eluent aspiration tubing, see the manual of the ion chro-
matograph.
4
Attaching the bottle cap
■ Screw the bottle cap onto the eluent bottle.
■ Loosen the tubing nipple a little.
■ Push the eluent aspiration tubing so far into the eluent bottle that
the aspiration filter is at the bottom of the bottle.
■ Tighten the tubing nipple again.
5
Connecting the level sensor cable
Connect the two plugs of the level sensor cable to the two sockets
(2-4) and (2-5).