User guide
3.1 Attaching the bottle cap
■■■■■■■■■■■■■■■■■■■■■■
4
■■■■■■■■
IC equiment: Dosino Regeneration
3 Installation
Proceed as follows to install the IC equiment: Dosino Regeneration:
■ Attach the bottle cap to the regeneration solution bottle.
■ Assemble the 800 Dosino and fasten it to the holder.
■ Connect the Dosino to the bottle.
■ Connect the Dosino to the inline filter.
■ Connect the inline filter to the suppressor.
3.1 Attaching the bottle cap
Required accessories
■ Bottle with regeneration solution
■ Eluent bottle cap (6.1602.160)
■ Adsorber tube (6.1619.000)
■ Adapter for adsorber tube (6.1624.000)
■ FEP aspiration tubing (6.1829.020)
1
Mounting the aspiration tubing
■ Use the capillary cutter to cut the aspiration tubing to such a
length that it touches the bottom of the bottle.
■ Place the aspiration tubing in the M6 opening of the bottle cap.
■ Place an O-ring in the M6 opening.
2
Inserting the stopper
■ Tighten the M8 stopper to the M8 opening of the bottle cap.