6.21 837 IC Degasser
IC Net 2.3
171
Time program
On the Program tab of the 837 IC Degasser window a user-defined
time program can be entered. This program is started automati-
cally as defined in the Start mode window (see 4.4.3) either at the
moment the determination is started (Start with determination) or at
the moment the sample is injected (Start with inject).
First column Time at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.
Second column Program instruction (see below).
Third column Parameter for program instruction (see be-
low).
Enabled Enable program start (a disabled program
is not started).
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
<Verify> Test the time program (error messages are
displayed if program is wrong).