User guide

5.2 762 IC Interface
IC Net 2.3
45
Protected Protocol
Data transfer with Protected Protocol be-
tween 830 IC Interface and PC for enhanced
data security.
Data transfer without Protected Protocol be-
tween 830 IC Interface and PC, select this
option if there is a problem with the RS
communication using the Protected Proto-
col.
Baud Rate Baud rate of the device.
Selection: 1200...115200
Default value: 115200
Parity Parity check.
Selection:
None, Even, Odd
Default value: None
Handshake Enable/disable software handshake mode.
Selection: Swchar, none
Default value: SWchar
FIFO size Intermediate memory in byte.
Selection: 1...16
Default value: 16
<ASK> Read current parameters from 762 IC Inter-
face.
<SET> Send current parameters to 762 IC Inter-
face.
Links
The Links tab of the 762 IC Interface is used for COM port selection
and settings.
Alias Name of the interface or device.
COM # If this entry is clicked with the right mouse
button, the following menu appears: