User guide

6.5 761 Compact IC
IC Net 2.3
85
even without connection to the PC.
Entry range: 0.0 ... 25.0 MPa
Full scale, µS/cm
Startup value for full scale range.
Selection: 50, 250, 1000 µS/cm
Remote lines
Startup value for remote line settings 18.
Selection: 0, 1
Initial
The Initial tab of the 761 Compact IC window for settings of system
startup values is only available for a disconnected system.
Flow, mL/min Startup value for flow rate of the high-
pressure pump.
Entry range: 0.20 ... 2.50 mL/min
Pressure max, MPa
Startup value for maximum pressure limit
for high-pressure pump.
Entry range:
0.0 ... 25.0 MPa
Pressure min, MPa
Startup value for minimum pressure limit for
high-pressure pump.
Entry range:
0.0 ... 25.0 MPa
Full scale, µS/cm
Startup value for full scale range.
Selection:
50, 250, 1000 µS/cm
Remote lines
Startup value for remote line settings 18.
Selection: 0, 1
Time program
On the Program tab of the 761 Compact IC window a user-defined
time program for the remote output lines of the 761 Compact IC
can be entered. This program is started automatically as defined
in the Start mode window (see section 4.4.3) either at the moment
the determination is started (
Start with determination) or at the
moment the sample is injected (
Start with inject).