Owner manual
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
IC-Ausrüstung: MiPT
■■■■■■■■
9
3.3 Equipping the supply bottle
Required accessories ■ Bottle (6.1608.070) filled with ultrapure water
■ Eluent bottle cap (6.1602.160)
■ Adsorber tube (6.1619.000), included in the accessories for the dosing
unit (6.3032.120)
■ Adapter for adsorber tube (6.1624.000)
■ FEP aspiration tubing (6.1819.110)
■ M8 stopper (6.1446.080), included in the accessories for the eluent
bottle cap (6.1602.160)
1
Mounting the aspiration tubing
■ Insert the aspiration tubing into the M6 opening of the eluent
bottle cap.
■ Use the capillary cutter to cut the aspiration tubing to such a
length that it touches the bottom of the bottle.
2
Inserting the stopper
■ Tighten the M8 stopper in the M8 opening of the eluent bottle
cap.
3
Mounting the adsorber tube
■ Fill the adsorber tube with some cotton and adsorber material.
■ Place the adsorber tube on the adapter.
■ Insert the adapter into the SGJ opening of the eluent bottle cap.