Owner manual

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
IC-Ausrüstung: MiPT
■■■■■■■■
9
3.3 Equipping the supply bottle
Required accessories Bottle (6.1608.070) filled with ultrapure water
Eluent bottle cap (6.1602.160)
Adsorber tube (6.1619.000), included in the accessories for the dosing
unit (6.3032.120)
Adapter for adsorber tube (6.1624.000)
FEP aspiration tubing (6.1819.110)
M8 stopper (6.1446.080), included in the accessories for the eluent
bottle cap (6.1602.160)
1
Mounting the aspiration tubing
Insert the aspiration tubing into the M6 opening of the eluent
bottle cap.
Use the capillary cutter to cut the aspiration tubing to such a
length that it touches the bottom of the bottle.
2
Inserting the stopper
Tighten the M8 stopper in the M8 opening of the eluent bottle
cap.
3
Mounting the adsorber tube
Fill the adsorber tube with some cotton and adsorber material.
Place the adsorber tube on the adapter.
Insert the adapter into the SGJ opening of the eluent bottle cap.