Owner manual
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
IC-Ausrüstung: MiPT
■■■■■■■■
11
1
■ Tighten the coupling to Port 1 of the Dosino.
2
■ Tighten one end of the transfer capillary to the coupling using a
pressure screw.
3.6 Installing capillaries
Required accessories
■ Transfer capillary (6.1841.000)
■ PEEK capillary, 0.5 mm ID / 3 m (6.1831.180)
■ Capillary cutter (6.2621.080)
1
Connecting the transfer capillary
■ Guide the free end of the transfer capillary through one of the ion
chromatograph's capillary feed-throughs.
■ Tighten the end of the transfer capillary to Port 2 of the injection
valve.
2
■ Tighten the PEEK capillary to Port 1 of the injection valve.
■ Guide the capillary out of the ion chromatograph through one of
the capillary feed-throughs.
■ Shorten the capillary using the capillary cutter in such a way that
it can be easily connected to the Sample Processor needle. Keep
the dead volume to a minimum.
■ Tighten the shortened capillary to the Sample Processor's needle
holder using a pressure screw.