User Manual

Subwindow Run 119
not yet been assigned in a command then the window
Device assign-
ment
appears for assigning the device.
Check solutions
Checks whether the solutions demanded in the device-specific com-
mands are present and available.
Check sensors
Checks whether the sensors demanded in the device-specific com-
mands are present and available.
Note
The Start test can also be triggered manually in the
READY
status with the menu
item
Tools, Run test
.
4 - Preconditioning (only for methods with conditioning)
In methods that contain commands with switched-on conditioning and in which
the option
Automatic conditioning is switched on in the START command then pre-
conditioning will start for all these commands. Preconditioning does not belong to
the determination itself and does not increase any counter (
Sample number, Statis-
tics
).
Conditioning requirement not fulfilled
After the start of preconditioning the status changes to
COND BUSY. In
this status conditioning can be started with
[Start] even though the
conditioning requirement has not been fulfilled. Conditioning can also
be canceled with
[Stop] or interrupted with [Hold] and then continued
with
[Cont]. If an error occurs during conditioning then neither the Error
track nor the Exit track will be carried out.
Conditioning requirement fulfilled
If the conditioning requirement has been fulfilled then the status
changes to
COND READY. In this status the determination can be
started with
[Start]. Conditioning can also be canceled with [Stop] or
interrupted with
[Hold] and then continued with [Cont]. The condition-
ing requirement will continue to be monitored continuously until the
determination has been started.
5 - Start method
At the start of the method the following checks and actions are triggered:
Check sample data
Checks that the entered Sample data corresponds to that defined in
the Method variables and is valid.
Assign determination ID
An unambiguous determination ID is assigned for the determination.
Reserve device (module)
The devices (or their functional units) used by the device-dependent
commands contained in the method are reserved for the duration of
the determination, i.e. they cannot be used in other workplaces nor
can they be operated manually.
Increment sample number
The
Sample number is increased by +1.
Increment statistics counter
If statistics has been switched on in both the
START command and on
the tab
Single determinations then the statistics counter is increased by
+1. If a new method is loaded or if the statistics counter agrees with