User Manual

Subwindow Run 129
Select method with assignment ID
If the option
Method selection by assignment ID is switched on the tab
Process then the Sample assignment ID will be used to load the method
defined in the
Sample assignment table.
3 - Carry out start test
During the Start test the following checks and actions are triggered:
Right to carry out the method?
Checks whether the logged-in user has the right to execute the par-
ticular method.
Method executable?
Checks whether the particular method is executable. If the method is
non-executable then a message appears asking for the method test to
be carried out and the method edited accordingly.
Check device (module) and rack
Checks whether the devices (or device types), modules (e.g. measur-
ing inputs, towers, etc.) and racks demanded in the device-specific
commands are present and available or assigned. If the device has
not yet been assigned in a command then the window
Device assign-
ment
appears for assigning the device.
Check solutions
Checks whether the solutions demanded in the device-specific com-
mands are present and available.
Check sensors
Checks whether the sensors demanded in the device-specific com-
mands are present and available.
Check sample data
Checks whether the entered sample data corresponds to that defined
in method variables and is valid.
Note
The Start test for the number of lines defined under
Autostart
can also be trig-
gered manually in the
READY
status with the menu items
Tools, Run test
or
Sam-
ple table, Run test
.
4 - Preconditioning (only for methods with conditioning)
In methods that contain commands with switched-on conditioning and in which
the option
Automatic conditioning is switched on in the START command then pre-
conditioning will start for all these commands. Preconditioning does not belong to
the determination itself and does not increase any counter (
Sample number or Sta-
tistics
).
Conditioning requirement not fulfilled
After the start of preconditioning the status changes to
COND BUSY. In
this status conditioning can be interrupted with
[Hold] and continued
with
[Cont]. If an error occurs during conditioning then neither the Error
track nor the Exit track will be carried out.
Conditioning requirement fulfilled
If the conditioning requirement has been fulfilled the status changes to
COND READY and the method then starts automatically.