User Manual
Subwindow Run 131
determination will be terminated, if an error occurs then the Error track
starts.
7 - Post-conditioning (only for methods with conditioning)
In methods that contain commands with switched-on conditioning and in which
the option
Automatic conditioning is switched on in the START command then for all
these commands post-conditioning will be started automatically with the loaded
method immediately after the end of the track that contains the conditioning
command. After the start of post-conditioning the status changes to
COND BUSY.
In this status the conditioning can be stopped with
[Stop] or interrupted with [Hold]
and continued with
[Cont]. If an error occurs during conditioning then neither the
Error track nor the Exit track will be carried out.
• Conditioning between two determinations
Conditioning will continue to run until the new method is loaded for the
next determination.
• Conditioning at the end of the last determination
After the Series end track has been carried out conditioning will con-
tinue to take place until it is terminated with
[Stop].
Device assignment
If at the Start test it is discovered that no device has been assigned for a device-
specific command then the window
Device assignment appears for assigning the
device:
Command type
Shows the command type.
Command name
Shows the command name.
Device type
Shows the device type selected in the command.
Device name
Device name
Selection of a device from the device table. Only those devices being able
to execute the command are selectable.