User Manual

Subwindow Run 131
determination will be terminated, if an error occurs then the Error track
starts.
7 - Post-conditioning (only for methods with conditioning)
In methods that contain commands with switched-on conditioning and in which
the option
Automatic conditioning is switched on in the START command then for all
these commands post-conditioning will be started automatically with the loaded
method immediately after the end of the track that contains the conditioning
command. After the start of post-conditioning the status changes to
COND BUSY.
In this status the conditioning can be stopped with
[Stop] or interrupted with [Hold]
and continued with
[Cont]. If an error occurs during conditioning then neither the
Error track nor the Exit track will be carried out.
Conditioning between two determinations
Conditioning will continue to run until the new method is loaded for the
next determination.
Conditioning at the end of the last determination
After the Series end track has been carried out conditioning will con-
tinue to take place until it is terminated with
[Stop].
Device assignment
If at the Start test it is discovered that no device has been assigned for a device-
specific command then the window
Device assignment appears for assigning the
device:
Command type
Shows the command type.
Command name
Shows the command name.
Device type
Shows the device type selected in the command.
Device name
Device name
Selection of a device from the device table. Only those devices being able
to execute the command are selectable.