User Manual
Subwindow Determination overview 239
measured values. Below it are listed the individual measured values and the re-
sults for
Slope, E(0) and Variance.
Calibration curve for CAL LOOP Conc
In the dialog window
Calibration curve a tab marked with the command name and
showing the calibration curve and calibration data will be shown for each
CAL
LOOP Conc
command. The command type is shown above the calibration curve.
The curve display shows the measured values and the curve calculated from
these measured values. Below it are listed the individual measured values and the
results for
Slope, E(0), c(blank) and Variance.
Calibration curve for STDADD
In the dialog window
Calibration curve a tab marked with the command name and
showing the standard addition curve and its evaluation data will be shown for
each STDADD command. The command type is shown above the calibration
curve. The curve display shows the measured values and the curve calculated
from these measured values. Below it are listed the individual measured values
and the results for
Slope, E(0), ion concentration and Variance.
Note
Calibration curves cannot be printed out directly from the dialog window
Calibra-
tion curve
. Instead you should use a report with a Report template in which the
fixed report
Calibration curve
is contained.
Control chart
With the menu item Determinations, Control chart... or the symbol the dialog
window
Control chart opens in which a control chart is shown for the selected de-
terminations containing statistical evaluations (mean value, number of determina-
tions, absolute and relative standard deviation) according to the loaded template.