User Manual

370 Method
Commands
Depending on the measured value one of the following four
MET commands can
be chosen:
MET pH
Potentiometric pH measurement with pH electrodes (measured quan-
tity pH).
MET U
Potentiometric voltage measurement with metal electrodes (measured
quantity voltage U).
MET Ipol
Voltametric measurement with selectable polarization current (meas-
ured quantity voltage U).
MET Upol
Amperometric measurement with selectable polarization voltage
(measured quantity current I).
MET pH
Command for Monotonic Equivalence Point Titrations with potentiometric pH
measurement.
Devices
This command can be executed with the following devices which have the MET
mode:
Titrino: 702, 716, 721, 736, 751, 785, 794, 798, 799
Titrando: 808, 809, 835, 836, 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command
MET pH are configured in the following 7 tabs:
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.
Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.
Titration parameters
Parameters for the run of the titration.
Stop conditions
Definitions of conditions which cause the stop of the titration.
Potentiometric evaluation
Parameters for the potentiometric evaluation of titration curves.
Additional evaluations
Definition of additional methods for evaluation of the titrations curves.