Commands 371
Additional measured values
Definition of additional measured values of other measuring com-
mands which are saved as additional column in the list of measuring
points.
General/Hardware
Screenshot and parameters: see DET pH - General/Hardware
Start conditions
Screenshot and parameters: see DET pH - Start conditions
Titration parameters
Parameters defining the run of the titration.
Command name
25 characters, [ MET pH # ]
Name of the command.
Titration rate
Titration rate
Three predefined sets of parameters can be selected for the titration rate:
slow, optimal and fast. For these sets the parameters for the Measured value
acceptance
and Dosing of increments are not displayed. In order to edit these
parameters the value
user must be selected.
slow
For titrations in which the finest details are also to be visible. However, this
can also increase noise and lead to the production of unwanted EPs.
[ optimal ]
Parameter set for all standard titrations; optimized for the most frequent
applications.