User Manual

Commands 373
Temperature
Temperature
-20.0 ... [ 25.0 ] ... 150.0 °C
(Titrando)
-170.0 ... [ 25.0 ] ... 500.0 °C
(Titrino)
Manually entered titration temperature. If a temperature sensor is con-
nected and the
Temperature measurement in the tab General / Hardware un-
der
Sensor is set to automatic or continuous then the temperature will be
measured continuously. This value is used for temperature correction in pH
measurements.
Stop conditions
Screenshot and parameters: see DET pH - Stop conditions
Potentiometric evaluation
Screenshot and parameters: see DET pH - Potentiometric evaluation
Additional evaluations
Screenshot and parameters: see DET pH - Additional evaluations
Additional measured values
Screenshot and parameters: see DET pH - Additional measured values
MET U
Command for Monotonic Equivalence Point Titrations with potentiometric
voltage measurement.
Devices
This command can be executed with the following devices which have the MET
mode:
Titrino: 702, 716, 721, 736, 751, 785, 794, 798, 799
Titrando: 808, 809, 835, 836, 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command
MET U are configured in the following 7 tabs:
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.
Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.
Titration parameters
Parameters for the run of the titration.