User Manual

Commands 413
General/Hardware
The general parameters for the control device, the electrodes and the stirrer are
defined on this tab.
Command name
25 characters, [ KFC # ]
Name of the command.
Device
Device name
Device name, [ not defined ]
Selection of the device from device table. Only those devices being able to
execute the command are selectable. Additionally,
not defined can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined is selected as Device name, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Indicator electrode
I(pol)
2, 5, [ 10 ], 20, 30 µA
Polarization current at the indicator electrode.
Note
On changing this parameter take care that the polarization current of the indica-
tor electrode, the endpoint and the control range are linked to each other (de-
tails see
Control parameters
).
Electrode test
on, [ off ]
Switching on/off the electrode test for polarized electrodes. The electrode
test for polarized electrodes is carried out during the transition from an in-
active condition to a measuring condition