Commands 415
Control parameters
Command name
25 characters, [ KFC # ]
Name of the command.
Endpoint
EP at
-2000.0 ... [ 50.0 ] ... 2000.0 mV
Measured value for the end point.
Control
Dynamics
1.0 ... [ 70 ] ... 2000 mV, off
The dynamics is entered as distance to the end point. Within the dynamics
iodine will be produced with
Min. rate. The bigger the control range the
slower the titration. Outside the dynamics iodine will be produced continu-
ously with
Max. rate.
The standard control parameters are optimal for most applications and
should not be altered. If you nevertheless need to alter the control parame-
ters for special reagents and/or samples take care that the polarization cur-
rent of the indicator electrode (
I(pol)), the end point and the dynamics are
linked to each other.