User Manual
452 Method
Parameters
The parameters for the command
MEAS pH are configured on the following 4 tabs:
•
General/Hardware
Parameters for device, sensor and stirrer.
•
Measuring parameters
Parameters for the measurement.
•
Evaluations
Defining further methods for evaluating measurement curves.
•
Additional measured values
Defining further measured values which can be saved as additional
columns in the list of measuring points.
General/Hardware
The general parameters for the control device, the sensor and the stirrer are de-
fined on this tab.
Command name
25 characters, [ MEAS pH # ]
Name of the command.
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined is selected as Device name, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.