User Manual

464 Method
Lower limit
[ -2'000.0 ] ... 2'000.0 mV
(for Window = Measured value)
[ 0 ] ... 999'999 s
(for Window = Time)
Measured value for the lower limit of the window.
Upper limit
-2'000.0 ... [ 2'000.0 ]
mV (for Window = Measured value)
0 ... [ 999'999] s
(for Window = Time)
Measured value for the upper limit of the window.
Additional measured values
Screenshot and parameters: see DET pH - Additional measured values
MEAS Ipol
Command for voltametric measurements with selectable polarization current.
Devices
This command can be executed with the following devices:
Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798, 799
Titrando: 808, 809, 835, 836, 841
Appearance
The command has the following appearance:
Parameters
The parameters for the command
MEAS Ipol are configured on the following 4
tabs:
General/Hardware
Parameters for device, sensor and stirrer.
Measuring parameters
Parameters for the measurement.
Evaluations
Defining further methods for evaluating measurement curves.
Additional measured values
Defining further measured values which can be saved as additional
columns in the list of measuring points.