Commands 473
Slope
0.0 ... [ 0.9 ] ... 10.0
Minimum difference between the slope before and after the break point.
The smaller the difference, the more break points will be found.
Smoothing factor
2 ... [ 5 ] ... 20
The higher the smoothing factor, the fewer break points will be found.
Window
Measured value, Time, [ off ]
A range (window) can be defined on the measured value axis or on the time
axis. The break point evaluation will only be carried out in the defined win-
dow. Only the first break point in the defined window will be recognized.
Lower limit
[ -200.0 ] ... 200.0 µA
(for Window = Measured value)
[ 0 ] ... 999'999 s
(for Window = Time)
Measured value for the lower limit of the window.
Upper limit
-200.0 ... [ 200.0 ] µA
(for Window = Measured value)
0 ... [ 999'999] s
(for Window = Time)
Measured value for the upper limit of the window.
Additional measured values
Screenshot and parameters: see DET pH - Additional measured values
MEAS T
Command for temperature measurements.
Devices
This command can be executed with the following devices:
Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798, 799
Titrando: 808, 809, 835, 836, 841
Appearance
The command has the following appearance:
Parameters
The parameters for the command
MEAS T are configured on the following 4 tabs:
General/Hardware
Parameters for device, sensor and stirrer.
Measuring parameters
Parameters for the measurement.
Evaluations
Defining further methods for evaluating measurement curves.
Additional measured values
Defining further measured values which can be saved as additional
columns in the list of measuring points.