User Manual

474 Method
General/Hardware
The general parameters for the control device, the sensor and the stirrer are de-
fined on this tab.
Command name
25 characters, [ MEAS T # ]
Name of the command.
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined is selected as Device name, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Sensor
Measuring input
[ 1 ] ... 2
(Titrando)
[ 1 ] ... 2, diff.
(Titrino)
Selection of the measuring input to which the sensor is connected.
Sensor
Sensor name, [ Temperature sensor ], pH electrode
Selection of a pH sensor from the pH sensors configured in the sensor ta-
ble.
Stirrer
Stirrer
[ 1 ] ... 4, off
(Titrando only)
Selection of the stirrer.
off means that no stirrer will be used.