User Manual

496 Method
Unit
mol/L, mmol/L, µmol/L, [ ppm ], g/L, mg/L, µg/L, mg/mL, µg/mL, %, mEq/L
Concentration unit of addition solution.
Dosing rate
0.01 ... 166.00 mL/min, [ maximum ]
Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the used exchange unit or dosing
unit.
Measuring parameters
Screenshot and parameters: see STDADD man - Measuring parameters
STDADD auto
Command for Standard addition with automatic addition of the standard addi-
tion solution from a dosing device in such a way that a constant potential differ-
ence results.
Devices
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command
STDADD auto are configured on the following 3
tabs:
General/Hardware
Parameters for device, dosing device, sensor and stirrer..
Standard addition
Parameters for standard addition.
Measuring parameters
Parameters for the measurement.
Standard addition curve
The standard addition curve with its evaluation data for a determination is shown
in the program part Database in the dialog window Calibration curve which is
opened with the menu item Determinations, Show calibration curve... or the sym-
bol
.
General/Hardware
Screenshot and parameters: see STDADD dos - General/Hardware