User Manual
506 Method
Temperature
Temperature
-20.0 ... [ 25.0 ] ... 150 °C
(Titrando)
-170.0 ... [ 25.0 ] ... 500 °C
(Titrino)
Manually entered measuring temperature. If a temperature sensor is con-
nected and
temperature measurement on the tab General/Hardware is set to
automatic or continuous, the temperature is measured continuously. The
value is used for the temperature adjustment during the measurement.
CAL LOOP Conc
Loop command for calibrating ion-selective electrodes. The command is used
together with the command
CAL MEAS Conc.
Appearance
The command has the following appearance:
Parameters
The parameters for the command
CAL LOOP Conc are configured in the following
dialog window:
•
CAL LOOP Conc
CAL LOOP Conc - properties
The number of standards and their concentration are defined in the dialog window
CAL LOOP Conc.
Command name
25 characters, [ Calibration loop Conc # ]
Name of the command.
Standards
Number of standards
1 ... [ 2 ] ... 5
Selection of the number of standards for the concentration calibration.
Unit conc.
mol/L, mmol/L, [ ppm ], %, g/L, mg/L, mg/mL, µg/L, µg/mL, mg/mL
Selection of the concentration unit of the standards.
Conc. standard #
0.001 ... 999'999.999
Entry of the corresponding concentration of each standard.