User Manual
508 Method
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined is selected as Device name, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Sensor
Measuring input
[ 1 ] ... 2
(Titrando)
[ 1 ] (855)
Selection of the measuring input to which the sensor is connected.
Sensor
Sensor name, [ ISE electrode ]
Selection of a ISE sensor from the ISE sensors configured in the sensor ta-
ble.
Temperature measurement
Type of temperature measurement:
continuous
A temperature sensor must be connected. The temperature is then meas-
ured continuously.
[ automatic ]
If a temperature sensor is connected, the temperature will be measured
continuously. Otherwise the temperature manually entered under
Tempera-
ture
(see Measuring parameters) will be used.
off
The temperature will not be measured. The temperature manually entered
under
Temperature (see Measuring parameters) will be used.
Stirrer
Stirrer
[ 1 ] ... 4, off
Selection of the stirrer.
off means that no stirrer will be used.
Stirring rate
-15 ... [ 8 ] ... 15
Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.
Switch off automatically
[ on ], off
The stirrer is automatically switched off at the end of the command if this
option is enabled.
Measuring parameters
Screenshot and parameters: see CAL MEAS pH - Measuring parameters