User Manual
Commands 563
CTRL - Properties
Command name
25 characters, [ Set lines # ]
Name of the command.
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined is selected as Device name, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Remote box
[ 1 ] ... 4
(only Titrando, 814, 815)
Selection of remote box at which the lines are to be set.
Set lines
Output signal
Binary pattern of exactly 14 characters (0, 1,*, p), [ ************** ], signal
template
Input of the binary pattern for the output signal or selecting a predefined
Signal pattern.
The following characters can be entered:
0
= Line inactive
1
= Line active
*
= Any line condition
p
= Set impulse (impulse length 200 ms. If an impulse with another length
should be emitted, a corresponding template must be defined.)
The bits are always numbered from right to left:
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Example:
************1* sets output line 1 to active (= set), for example,
with a connected Titrino this would be a Stop command.
************0*