User Manual
726 Configuration
Parameters for preparation
Configures the parameters to be used in the commands
PREP and EMPTY.
Dosing port Prep/Empty
[ Dosing port 1 ], Dosing port 2, Fill port, Special port
Dosing port via which the cylinder contents are to be ejected during prepa-
ration and emptying.
Dosing rate Dosing port 1
0.01...166.00 mL/min, [ maximum ]
Speed at which dosing is to be carried out at Dosing port 1. The maximum
dosing rate depends on the cylinder volume of the dosing unit used (see
below). When the function is carried out the speed will automatically be re-
duced to the largest possible value.
Dosing rate Dosing port 2
0.01...166.00 mL/min, [ maximum ]
Speed at which dosing is to be carried out at Dosing port 2. The maximum
dosing rate depends on the cylinder volume of the dosing unit used (see
below). When the function is carried out the speed will automatically be re-
duced to the largest possible value.
Dosing rate Fill port
0.01...166.00 mL/min, [ maximum ]
Speed at which dosing or filling is to be carried out at the Fill port. The
maximum dosing rate depends on the cylinder volume of the dosing unit
used (see below). When the function is carried out the speed will automati-
cally be reduced to the largest possible value.
Dosing rate Special port
0.01...166.00 mL/min, [ maximum ]
Speed at which dosing is to be carried out at the Special port. The maxi-
mum dosing rate depends on the cylinder volume of the dosing unit used
(see below). When the function is carried out the speed will automatically
be reduced to the largest possible value.
Maximum Dosing rate for dosing unit as a function of the cylinder volume:
Cylinder volume Maximum rate
2 mL 6.6 mL/min
5 mL 16.6 mL/min
10 mL 33.3 mL/min
20 mL 66.6 mL/min
50 mL 166.0 mL/min
Note
Enter lower rates for high-viscosity liquids.