User Manual

Subwindow Sensors 741
Warning limits for slope
on, [ off ]
If this option is switched on then the values for the slope will be shown in
orange if the limits are infringed.
Lower limits
-2.147E+9 ... 2.147E+9
pH electrode:
97.5 %
ISE (+1):
57.2 mV ISE (-1): -61.2 mV
ISE (+2):
27.7 mV ISE (-2): -31.7 mV
ISE (+3):
17.7 mV ISE (-3): -21.7 mV
ISE (+4):
12.8 mV ISE (-4): -16.8 mV
Lower warning limits for the slope.
Upper limits
-2.147E+9 ... 2.147E+9
pH electrode:
102.5 %
ISE (+1):
61.2 mV ISE (-1): -57.2 mV
ISE (+2):
31.7 mV ISE (-2): -27.7 mV
ISE (+3):
21.7 mV ISE (-3): -17.7 mV
ISE (+4):
16.8 mV ISE (-4): -12.8 mV
Upper warning limits for the slope.
Intervention limits for the slope
on, [ off ]
If this option is switched on then the values for the slope will be shown in
red if the limits are infringed.
Lower limits
-2.147E+9 ... 2.147E+9
pH:
95.0 %
ISE (+1):
55.0 mV ISE (-1): -62.4 mV
ISE (+2):
26.7 mV ISE (-2): -32.7 mV
ISE (+3):
16.7 mV ISE (-3): -22.7 mV
ISE (+4):
11.8 mV ISE (-4): -17.8 mV
Lower intervention limits for the slope.
Upper limits
-2.147E+9 ... 2.147E+9
pH:
105.0 %
ISE (+1):
62.4 mV ISE (-1): -55.0 mV
ISE (+2):
32.7 mV ISE (-2): -26.7 mV
ISE (+3):
22.7 mV ISE (-3): -16.7 mV
ISE (+4):
17.8 mV ISE (-4): -11.8 mV
Upper intervention limits for the slope.
Warning limits for pH(0) and E(0)
on, [ off ]
If this option is switched on then the values for pH(0) or E(0) will be shown
in orange if the limits are infringed.