User manual
50 
SCALE OPERATION  
SECURITY SEAL INSTALLATION 
The certification of the weighing mechanism of the scale version of this scanner is subject to federal, state and local 
Weights and Measures statutes and regulations and can only be performed by authorized government agencies 
and/or their duly registered agents. Each time the scale or weighing mechanism is calibrated, it should be properly 
sealed with a paper seal or a wire seal prior to being placed into service in commerce. 
It is the responsibility of the owner of the scale to confirm compliance with the relevant Weights and Measures statutes 
and regulations applicable in your area by checking with the appropriate government agency before placing a newly 
calibrated unit into service or removing any official seals. 
Type of seal to be used will depend on the guidelines specified by the local Weights and Measures authorities. 
The security seal must only be installed if there were no
 errors during the scale calibration verification tests. 
Pressure Sensitive Security Seal  
1.  Temporarily remove the platter and place it in a safe location. 
2.  Apply the adhesive calibration security seal over the 
calibration button access hole (see Figure 55). 
3.  With very light pressure, push down on the adhesive 
calibration security seal label to ensure a firm adhesion is 
created (see Figure 56). 
4.  Reinstall the platter. 
\ 
* These procedures are for MS2420 Scanner/Diva Scale models only. 
Figure 55. 
Figure 56. 
Figure 57. Platter Installation 










