Operating instructions

The menu
28
Calibration with external weights (VariCal)
The balance is calibrated with a selectable* external weight.
* With certified versions of the balances, the weight is preallocated and can
not be changed.
Test of the balance with internal weight
In this setting the accuracy test of the balance is performed with the internal
weight.
Test of the balance with external weights
The accuracy of the balance can be checked with any external weight.
You will find information on how to perform the calibration and test function
in Sections 2.6, 5.6 and 5.7.
UAr∫CAL
tESt int
tESt E
4.5 Switching automatic adjustment call-up on or off
In this menu option you can switch the call-up of the automatic adjustment or test on or off.
Note: If you have set «FACT» in the menu option Adjustment (calibration), the automatic adjustment call-up is always
active and will thus be skipped in the menu. It becomes active again as soon as «FACT» is switched off.
The following settings are available:
Automatic adjustment or test call-up switched on
This is the factory setting. The balance uses a flashing «Cal» in the display
to prompt you to adjust (calibrate) or test it with the internal weight or external
weights.
The call-up is initiated by, e.g. ambient temperature changes.
Automatic adjustment or test call-up switched off
The automatic adjustment or test call-up is switched off.
Note
With certified balances, the automatic adjustment or test call-up can not be
switched off.
InFo on
Cal
InFo oFF
Cal