Operating instructions

The menu
30
Dynamic weighing with automatic start
Your balance determines an average weighing result over a preset time
interval. This setting is suitable for unstable weighing samples (e.g.
animals). With this setting, the dynamic weighing starts automatically.
Dynamic weighing with manual start
Analogous to dynamic weighing with automatic start, but the weighing cycle
must be started manually.
You will find information on working with the functions in Section 5.
F dYn A
F dYn Ÿ≈
4.7 Setting the vibration adapter
The vibration adapter can be used to match your balance to the ambient conditions (vibrations, drafts at location).
2
3
1
The following settings are available:
Setting for normal ambient conditions
This is the factory setting. The balance operates at moderate speed.
Setting for unstable surroundings
The filter setting of the balance is higher than in the factory setting, but the
balance is less sensitive to external influences.
Setting for virtually disturbance-free, stable surroundings
The filter setting of the balance is lower than in the factory setting, but the
balance is more sensitive to external influences.